Figure 7. DQE plotted for the measurements with residual fixed-pattern noise removed. Data were measured with 70 kVp, 0.5 mm of added copper filtration, 0.3-mR (8 x 10-8 C/kg) incident exposure, with the grid removed. The DQE of the FPD system is more than twice that of a storage phosphor CR system (1).