Radiology, Vol 138, 477-481, Copyright © 1981 by Radiological Society of North America
ARTICLES |
PB Hoffer, HJ Berger, J Steidley, AF Brendel, A Gottschalk and BL Zaret
The authors describe a miniature cadmium telluride (CdTe) detector module for continuous monitoring of ventricular function using an equilibrium radionuclide blood-pool label. The detector and collimator are small, light, and suitable for direct attachment to the chest wall. Clinical studies in 18 patients using a prototype system demonstrated reasonably good correlation with left-ventricular ejection fractions (LVEF) determined by first-pass studies performed with a multicrystal scintillation camera (r = 0.74) and gated equilibrium studies performed with a computerized sodium iodide (Nal) probe (r = 0.76). The CdTe device may prove to be useful in patients in intensive and coronary care units as well as in ambulatory patients.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |