MC Light, SY Molloi, DR Yandow and FN Ranallo
Knee arthrography, as performed at the authors' institution, was simulated and scattered radiation exposure to a radiologist's gonads, thyroid, and eye lens was measured with a sensitive ionization chamber. Results show that radiologists who regularly conduct knee arthrography examinations can incur doses to the gonads that are less than 6% of the U.S. limits, and to the thyroid and eye that are approximately 10% of the U.S. limits. Since the scatter radiation from overhead imaging of stress views constituted most (greater than or equal to 60%) of the dose to the lens of the eye and the thyroid, spot imaging was evaluated as a substitute for overhead imaging in the assessment of the anterior cruciate ligament. This substitution resulted in no loss of clinical information and has now completely replaced overhead imaging of stress views at this institution.