Radiology, Vol 169, 35-39, Copyright © 1988 by Radiological Society of North America
ARTICLES |
LJ Kool, DL Busscher, H Vlasbloem, J Hermans, PC v.d. Merwe, PR Algra and W Herstel
Department of Diagnostic Radiology, University Hospital Leiden, The Netherlands.
To evaluate the efficacy of AMBER, a multiple-beam equalization system for chest radiography, the authors performed a nodule detection study using an anthropomorphic chest phantom. AMBER and conventional images were compared. The images were read by four observers, and analysis was done by means of modified receiver-operating characteristic (ROC) curves (free ROC curves [FROC]). The results of the FROC analysis show a significant increase in the detectability of nodules (P less than .001) projected over the mediastinum with the use of AMBER. No significant difference between AMBER and conventional images was noted in detectability of nodules projected over the lung.
This article has been cited by other articles:
![]() |
References J. ICRU, April 1, 2008; 8(1): 57 - 62. [PDF] |
||||
![]() |
H. P. McAdams, E. Samei, J. Dobbins III, G. D. Tourassi, and C. E. Ravin Recent Advances in Chest Radiography Radiology, December 1, 2006; 241(3): 663 - 683. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. M. Kroft, J. Geleijns, B. J. A. Mertens, W. J. H. Veldkamp, H. M. Zonderland, and A. de Roos Digital Slot-Scan Charge-coupled Device Radiography versus AMBER and Bucky Screen-Film Radiography for Detection of Simulated Nodules and Interstitial Disease in a Chest Phantom Radiology, April 1, 2004; 231(1): 156 - 163. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |