Radiology, Vol 173, 65-69, Copyright © 1989 by Radiological Society of North America
ARTICLES |
L Tabar and AG Haus
Falun Central Hospital, Sweden.
Sensitometrically exposed film strips and clinical mammograms from single-emulsion Kodak Ortho M SO-177 and double-emulsion Kodak T-Mat M II films were processed in separate film processors set up for standard and extended-cycle processing. For the extended-cycle method, it is necessary to use a process that is dedicated to mammographic films only. Radiation dose reductions of approximately 30% for Ortho M film and 13% for T-Mat M II film were achieved with the extended-cycle process. In the mammogram comparisons, higher-contrast images were obtained with Ortho M film in the extended-cycle process, which allowed for improved demonstration of marginal structural characteristics of soft-tissue masses and better differentiation of benign and malignant tumors. No significant differences in contrast were observed in the T- Mat M II mammograms obtained with the extended-cycle process.
This article has been cited by other articles:
![]() |
A D Dimakopoulou, I A Tsalafoutas, E K Georgiou, and E N Yakoumakis Image quality and breast dose of 24 screen-film combinations for mammography Br. J. Radiol., February 1, 2006; 79(938): 123 - 129. [Abstract] [Full Text] [PDF] |
||||
![]() |
I A Tsalafoutas, A D Dimakopoulou, E D Koulentianos, A N Serefoglou, and E N Yakoumakis Variation of the sensitometric characteristics of seven mammographic films with processing conditions Br. J. Radiol., August 1, 2004; 77(920): 666 - 671. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |