Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Cardenosa, G.
Right arrow Articles by Eklund, G. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Cardenosa, G.
Right arrow Articles by Eklund, G. W.

Radiology, Vol 182, 894-895, Copyright © 1992 by Radiological Society of North America


ARTICLES

Effects of a defective filter on magnification image quality in mammography

G Cardenosa and GW Eklund
Susan G. Komen Breast Center, St Francis Medical Center, Peoria, IL 61614.

A defective molybdenum filter in a dedicated mammographic unit was found to have a detrimental effect on image quality. The authors present a simple method for testing of molybdenum filters and suggest that it be incorporated into the acceptance testing routine for new mammographic units and that it be used in annual quality control checks.