Radiology, Vol 183, 179-187, Copyright © 1992 by Radiological Society of North America
ARTICLES |
JT Dobbins 3d, JJ Rice, CA Beam and CE Ravin
Department of Radiology, Duke University Medical Center, Durham, NC 27710.
Images of a phantom obtained with computed radiography and standard screen-film imaging were compared to evaluate observer threshold perception performance with a modified contrast-detail technique. Optimum exposure necessary for performance with the imaging plate technique to match that with screen-film techniques was determined, as was comparative performance with variation in kilovoltages, plate type, spatial enhancement, and hard-copy interpolation method. It was found that computed radiography necessitates about 75%-100% more exposure than screen-film radiography to optimally match performance with Ortho- C film with Lanex regular or medium screens (Eastman Kodak, Rochester, NY) for detection of objects 0.05-2.0 cm in diameter. However, only minimal loss of detection performance (approximately 10% overall) was experienced if standard screen-film exposures were used with computed radiography. Little change in observer performance was found with variation in plate type, spatial enhancement, or method of hard-copy interpolation. However, perception performance with computed radiographic images was better at lower kilovoltages.
This article has been cited by other articles:
![]() |
C S MOORE, A W BEAVIS, and J R SAUNDERSON Investigation of optimum X-ray beam tube voltage and filtration for chest radiography with a computed radiography system Br. J. Radiol., October 1, 2008; 81(970): 771 - 777. [Abstract] [Full Text] [PDF] |
||||
![]() |
C S Moore, G P Liney, A W Beavis, and J R Saunderson A method to optimize the processing algorithm of a computed radiography system for chest radiography Br. J. Radiol., September 1, 2007; 80(957): 724 - 730. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Mazzocchi, G. Belli, S. Busoni, C. Gori, I. Menchi, P. Salucci, A. Taddeucci, and G. Zatelli AEC set-up optimisation with computed radiography imaging Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 169 - 173. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. M. Kroft, W. J. H. Veldkamp, B. J. A. Mertens, M. V. Boot, and J. Geleijns Comparison of Eight Different Digital Chest Radiography Systems: Variation in Detection of Simulated Chest Disease Am. J. Roentgenol., August 1, 2005; 185(2): 339 - 346. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Tingberg and D. Sjostrom Optimisation of image plate radiography with respect to tube voltage Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 286 - 293. [Abstract] [Full Text] [PDF] |
||||
![]() |
I D Honey, A Mackenzie, and D S Evans Investigation of optimum energies for chest imaging using film-screen and computed radiography Br. J. Radiol., May 1, 2005; 78(929): 422 - 427. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Eisenhuber, A. Stadler, M. Prokop, M. Fuchsjager, M. Weber, and C. Schaefer-Prokop Detection of Monitoring Materials on Bedside Chest Radiographs with the Most Recent Generation of Storage Phosphor Plates: Dose Increase Does Not Improve Detection Performance Radiology, April 1, 2003; 227(1): 216 - 221. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. P. Frush, L. F. Donnelly, and H. G. Chotas Contemporary Pediatric Thoracic Imaging Am. J. Roentgenol., September 1, 2000; 175(3): 841 - 851. [Full Text] [PDF] |
||||
![]() |
R. V. Kaczmarek, B. J. Conway, R. O. Slayton, and O. H. Suleiman Results of a Nationwide Survey of Chest Radiography: Comparison with Results of a Previous Study Radiology, June 1, 2000; 215(3): 891 - 896. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |