Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rueter, F. G.
Right arrow Articles by Suleiman, O. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rueter, F. G.
Right arrow Articles by Suleiman, O. H.

Radiology, Vol 185, 43-46, Copyright © 1992 by Radiological Society of North America


ARTICLES

Radiography of the lumbosacral spine: characteristics of examinations performed in hospitals and other facilities

FG Rueter, BJ Conway, JL McCrohan, RJ Slayton and OH Suleiman
Center for Devices and Radiological Health, Food and Drug Administration, Rockville, MD 20857.

A survey of 523 facilities performing radiographic lumbosacral spine examinations was conducted as a part of the Nationwide Evaluation of X- Ray Trends program in 1987 and 1989. Average patient radiation exposure was measured by using a standard phantom developed by the Center for Devices and Radiological Health of the Food and Drug Administration, U.S. Public Health Service. Data pertaining to radiographic equipment, film processing, and radiographic techniques were obtained. The overall average entrance skin air kerma was 3.65 mGy (420 mR). Ninety-eight percent (496 of 506) of observed facilities used a grid, 89% (466 of 523) had tube potentials between 70 and 89 kVp, and 67% (280 of 418) used screen-film systems with a speed of 400. The most important finding is that underprocessing of film remains a major concern. Thirty- three percent (78 of 234) of the hospitals, 25% of the radiologists in private practice (four of 16), 33% of the nonradiologist private practitioners (27 of 82), and 48% (69 of 143) of the chiropractors underprocessed their film.


This article has been cited by other articles:


Home page
RadiologyHome page
D. C. Spelic, R. V. Kaczmarek, and O. H. Suleiman
Nationwide Evaluation of X-ray Trends Survey of Abdomen and Lumbosacral Spine Radiography
Radiology, July 1, 2004; 232(1): 115 - 125.
[Abstract] [Full Text] [PDF]