Radiology, Vol 186, 395-398, Copyright © 1993 by Radiological Society of North America
ARTICLES |
HG Chotas, CE Floyd Jr, JT Dobbins 3d and CE Ravin
Department of Radiology, Duke University Medical Center, Durham, NC 27710.
A photostimulable storage phosphor (PSP) digital radiography system was evaluated regarding the signal-to-noise ratio (S/N) on chest images acquired at differing peak kilovoltage settings but with matched risk from radiation exposure. Images of two chest phantoms were acquired by using bedside (portable) imaging equipment at tube voltages ranging from 60 to 120 kV. Exposure factors were set so that the effective dose equivalent, a risk estimator weighted for various organs, was approximately equal in all exposures. The S/N in the lung-equivalent regions was found to be slightly higher (maximum, 15%) in the low- energy exposures, while the S/N values in the mediastinum- and subdiaphragm-equivalent regions were approximately equal at all kilovoltage settings. The absence of a high sensitivity of S/N to kilovoltage in risk-matched PSP images should enable institutions to select x-ray beam quality on the basis of other imaging criteria.
This article has been cited by other articles:
![]() |
E Alpoz, E Sogur, and B. Baksi Akdeniz Perceptibility curve test for digital radiographs before and after application of various image processing algorithms Dentomaxillofac. Radiol., December 1, 2007; 36(8): 490 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
C S Moore, G P Liney, A W Beavis, and J R Saunderson A method to optimize the processing algorithm of a computed radiography system for chest radiography Br. J. Radiol., September 1, 2007; 80(957): 724 - 730. [Abstract] [Full Text] [PDF] |
||||
![]() |
H Warren-Forward, L Arthur, L Hobson, R Skinner, A Watts, K Clapham, D Lou, and A Cook An assessment of exposure indices in computed radiography for the posterior-anterior chest and the lateral lumbar spine Br. J. Radiol., January 1, 2007; 80(949): 26 - 31. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. W. Hamer, C. B. Sirlin, M. Strotzer, I. Borisch, N. Zorger, S. Feuerbach, and M. Volk Chest Radiography with a Flat-Panel Detector: Image Quality with Dose Reduction after Copper Filtration Radiology, November 1, 2005; 237(2): 691 - 700. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Tingberg and D. Sjostrom Optimisation of image plate radiography with respect to tube voltage Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 286 - 293. [Abstract] [Full Text] [PDF] |
||||
![]() |
I D Honey, A Mackenzie, and D S Evans Investigation of optimum energies for chest imaging using film-screen and computed radiography Br. J. Radiol., May 1, 2005; 78(929): 422 - 427. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Uffmann, U. Neitzel, M. Prokop, N. Kabalan, M. Weber, C. J. Herold, and C. Schaefer-Prokop Flat-Panel-Detector Chest Radiography: Effect of Tube Voltage on Image Quality Radiology, May 1, 2005; 235(2): 642 - 650. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Bernhardt, U. Rapp-Bernhardt, H. Lenzen, F. W. Roehl, S. Diederich, K. Papke, K. Ludwig, and W. Heindel Low-Voltage Digital Selenium Radiography: Detection of Simulated Interstitial Lung Disease, Nodules, and Catheters--A Phantom Study Radiology, September 1, 2004; 232(3): 693 - 700. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. T. Dobbins III, E. Samei, H. G. Chotas, R. J. Warp, A. H. Baydush, C. E. Floyd Jr, and C. E. Ravin Chest Radiography: Optimization of X-ray Spectrum for Cesium Iodide-Amorphous Silicon Flat-Panel Detector Radiology, January 1, 2003; 226(1): 221 - 230. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |