Radiology, Vol 188, 215-218, Copyright © 1993 by Radiological Society of North America
ARTICLES |
LK Jordan 3d, CE Floyd Jr, JY Lo and CE Ravin
Department of Radiology, Duke University Medical Center, Durham, NC 27710.
Scatter fractions (SFs) measured in patients undergoing erect posteroanterior (PA) and lateral chest radiography with a 12:1 antiscatter grid are reported. Modifications to the posterior beam-stop (PBS) technique allowed measurement of scatter in these patients, without altering the diagnostic image and without additional radiation exposure. The SF measurements are reported by anatomic location on 42 clinical chest images. Average SF values ranged from 0.27 to 0.90 on lateral radiographs and from 0.27 to 0.68 on PA radiographs. Scatter measurements with the 12:1 grid were found to be greater than estimates from previous PA chest phantom experiments. To the authors' knowledge, they were the first to measure radiation scatter with the PBS technique in patients undergoing PA and lateral chest radiography with the antiscatter grid.
This article has been cited by other articles:
![]() |
E. Samei, J. Y. Lo, T. T. Yoshizumi, J. L. Jesneck, J. T. Dobbins II, C. E. Floyd Jr, H. P. McAdams, and C. E. Ravin Comparative Scatter and Dose Performance of Slot-Scan and Full-Field Digital Chest Radiography Systems Radiology, June 1, 2005; 235(3): 940 - 949. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Ullman, M. Sandborg, D. R. Dance, R. Hunt, and G. Alm Carlsson Distributions of scatter-to-primary and signal-to-noise ratios per pixel in digital chest imaging Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 355 - 358. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |