Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tucker, D. M.
Right arrow Articles by Barnes, G. T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tucker, D. M.
Right arrow Articles by Barnes, G. T.

Radiology, Vol 188, 271-274, Copyright © 1993 by Radiological Society of North America


ARTICLES

Scatter in computed radiography

DM Tucker, M Souto and GT Barnes
Department of Radiology, University of Alabama Hospitals and Clinics, Birmingham 35233.

The effects of scatter radiation on images obtained with storage phosphor computed radiography (CR) were evaluated. The scatter degradation factor was extended to account separately for forward scatter and backscatter. Values for the ratios of forward scatter to primary radiation (sf/p) and backscatter to primary radiation (sb/p) were measured for various x-ray tube potentials, phantom thicknesses, and field sizes for CR imaging plates (IPs). Similar measurements were made for a common screen-film system. The sb/p values for the IPs were 0.05-3.80. Standard CR cassettes do not have lead foil behind the IP to control backscatter; the results of this study indicate that such control is needed. The sf/p values were 40%-70% higher in IPs (50%-120% higher when backscatter was included) than those in the conventional screen-film system. This difference is attributed to the differences in the K-absorption edges of barium and gadolinium and the fact that scatter has lower effective energy than the primary beam as they exit the patient.


This article has been cited by other articles:


Home page
Br. J. Radiol.Home page
H Warren-Forward, L Arthur, L Hobson, R Skinner, A Watts, K Clapham, D Lou, and A Cook
An assessment of exposure indices in computed radiography for the posterior-anterior chest and the lateral lumbar spine
Br. J. Radiol., January 1, 2007; 80(949): 26 - 31.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
RADIOLOGY RADIOGRAPHICS RSNA JOURNALS ONLINE
Copyright © 1993 by the Radiological Society of North America.