Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by DeBruhl, N. D.
Right arrow Articles by Bassett, L. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by DeBruhl, N. D.
Right arrow Articles by Bassett, L. W.

Radiology, Vol 189, 95-98, Copyright © 1993 by Radiological Society of North America


ARTICLES

Silicone breast implants: US evaluation

ND DeBruhl, DP Gorczyca, CY Ahn, WW Shaw and LW Bassett
Iris Cantor Center for Breast Imaging, Department of Radiological Sciences, UCLA Medical Center 90024-6952.

PURPOSE: To determine the value of breast ultrasonography (US) in the assessment of silicone breast implants for rupture. MATERIALS AND METHODS: Seventy-four women with local or systemic symptoms related to silicone implants underwent breast US. Of these, 28 underwent surgical removal of the implants. RESULTS: Of 57 implants removed, 37 were intact. The most reliable sign of an intact implant was an anechoic interior, although reverberation artifact and radical folds could be seen. Of 20 ruptured implants, 16 were intracapsular and four were extracapsular ruptures. The most reliable US sign of rupture was echogenic, horizontal ("stepladder") lines (14 of 20 ruptures). Two of the four extracapsular ruptures were accurately identified as echogenic nodules outside the implant; two were false-negative findings. Three intracapsular ruptures identified at US were false-positive; six were false-negative. Overall sensitivity for rupture was 70%, specificity was 92%, positive predictive value was 82%, and negative predictive value was 85%. CONCLUSION: Breast US is capable of depicting intracapsular and extracapsular rupture of breast implants.


This article has been cited by other articles:


Home page
Journal of Diagnostic Medical SonographyHome page
S. Lu, G. A. Rouse, and M. De Lange
Sonoographic Detection of Silicone Breast Implant Rupture
Journal of Diagnostic Medical Sonography, January 1, 1995; 11(1): 3 - 8.