Radiology, Vol 189, 769-774, Copyright © 1993 by Radiological Society of North America
ARTICLES |
M Simon, DJ Rabkin, S Kleshinski, D Kim and BJ Ransil
Department of Radiology, Harvard Thorndike Laboratory, Harvard Medical School, Beth Israel Hospital, Boston, MA 02215.
PURPOSE: A physiologic in vitro model of the human inferior vena cava (IVC) was developed to evaluate the clot-trapping efficiency of various IVC filters. MATERIALS AND METHODS: The flow model closely simulates the physical parameters of the human IVC in fluid viscosity, specific gravity, temperature, and pulsatility; blood flow velocity, volume, and inherent turbulence; and vein orientation, wall compliance, and clot composition. Five filters--Greenfield (G), Bird's Nest (BN), Vena Tech (VT), Simon nitinol (SN), and titanium Greenfield (TG)--were compared by using two vena cava and blood clot sizes and horizontal and vertical orientation of the model. RESULTS: Each filter varied significantly in its clot-capturing efficiency, depending on vena cava size and orientation and emboli size. Overall, decreasing rank order in filter clot-trapping efficiency was SN, BN, VT, TG, and G. CONCLUSION: The in vitro model of the human IVC is an alternative to multicenter clinical trials in evaluating the relative clot-trapping efficiency of various IVC filters.
This article has been cited by other articles:
![]() |
J. M. P. de Godoy and D. M. Braile In-Vitro Evaluation of a New Inferior Vena Cava Filter--The Stent-Filter Vascular and Endovascular Surgery, May 1, 2004; 38(3): 225 - 228. [Abstract] [PDF] |
||||
![]() |
M. B. Streiff Vena caval filters: a comprehensive review Blood, June 15, 2000; 95(12): 3669 - 3677. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |