Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Krutchen, A. E.
Right arrow Articles by Hunter, D. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Krutchen, A. E.
Right arrow Articles by Hunter, D. W.

Radiology, Vol 200, 159-163, Copyright © 1996 by Radiological Society of North America


ARTICLES

The mechanisms of positional dysfunction of subclavian venous catheters

AE Krutchen, H Bjarnason, DJ Stackhouse, GK Nazarian, JE Magney and DW Hunter
Department of Radiology, University of Minnesota Hospital and Clinic, UMHC, Minneapolis 55455, USA.

PURPOSE: To determine the origin of subclavian vein catheter and lead dysfunction. MATERIALS AND METHODS: Cineradiography was performed on 10 patients with subclavian venous catheter dysfunction and three patients with pacemaker or defibrillator lead dysfunction. The leads and catheters were removed and replaced with use of a fluoroscopically guided technique; the needle entered the vein lateral to the first rib. Repeat cine examinations were performed following placement of new catheters. RESULTS: The cause of the dysfunction of all 10 catheters was shown to be pinch by the subclavicular musculotendinous tissues as the catheter passed below the clavicle toward its entry into the vein. All three leads were entrapped in the subclavicular tissues and stretched during abduction. The abnormal motion and clinical problems were eliminated after replacement. CONCLUSION: Subclavian catheter and lead malfunction is not due to compression between the first rib and the clavicle. It is due to entrapment in the subclavius muscle- costoclavicular ligament complex, which binds or compresses the device during movements. These problems can be avoided by employing fluoroscopically guided puncture techniques that enter the vein lateral to the first rib.


This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
B. Bessoud, T. de Baere, V. Kuoch, E. Desruennes, M.-F. Cosset, N. Lassau, and A. Roche
Experience at a Single Institution with Endovascular Treatment of Mechanical Complications Caused by Implanted Central Venous Access Devices in Pediatric and Adult Patients
Am. J. Roentgenol., February 1, 2003; 180(2): 527 - 532.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
B. Funaki
Central Venous Access: A Primer for the Diagnostic Radiologist
Am. J. Roentgenol., August 1, 2002; 179(2): 309 - 318.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
RADIOLOGY RADIOGRAPHICS RSNA JOURNALS ONLINE
Copyright © 1996 by the Radiological Society of North America.