Radiology, Vol 203, 471-476, Copyright © 1997 by Radiological Society of North America
ARTICLES |
OH Suleiman, BJ Conway, P Quinn, RG Antonsen, FG Rueter, RJ Slayton and DC Spelic
Center for Devices and Radiological Health, U.S. Food and Drug Administration, Rockville, Md. 20850, USA.
PURPOSE: To determine the average abdominal entrance air kerma, low- contrast sensitivity, and spatial resolution in upper gastrointestinal tract fluoroscopy in the United States. MATERIALS AND METHODS: A random sample of fluoroscopic facilities was selected to be surveyed for the Nationwide Evaluation of X-ray Trends program. Measurements were performed by using a newly developed fluoroscopic phantom. The surveys were conducted by state radiation control personnel. RESULTS: Average air kerma rates 1 cm above the tabletop, free in air, were 43 mGy/min (n = 340). The rate increased to 64 mGy/min when a 1.6-mm-thick copper filter, which simulated the use of barium contrast medium, was added to increase attenuation. The average entrance air kerma, free in air, for radiographs was 3.4 mGy, and an average of 12 radiographs were obtained per examination. Of 352 facilities surveyed, 306 (87%) were able to resolve wire mesh with 20 or more lines per inch. Of 339 facilities for which percentage contrast could be calculated, 192 (57%) had minimum percentage contrast values of 4% or more. CONCLUSION: Spatial resolution for fluoroscopy is adequate for most of the facilities surveyed, but a substantial proportion of facilities could not visualize low-contrast test objects, which strongly suggests image quality problems.
This article has been cited by other articles:
![]() |
D. C. Spelic, R. V. Kaczmarek, and O. H. Suleiman Nationwide Evaluation of X-ray Trends Survey of Abdomen and Lumbosacral Spine Radiography Radiology, July 1, 2004; 232(1): 115 - 125. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Taubel, B. A. Schueler, T. J. Vrieze, and J. E. Gray Quantification of Fluoroscopic Imaging System Contrast by Using Video Waveform Monitoring RadioGraphics, March 1, 2001; 21(2): 507 - 518. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Peter, W. Pavlicek, and J. M. Owen Soft-Copy Quality Control of Digital Spot Images Obtained by Using X-ray Image Intensifiers Radiology, September 1, 2000; 216(3): 810 - 819. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |