Radiology, Vol 206, 803-810, Copyright © 1998 by Radiological Society of North America
ARTICLES |
P Steiner, R Botnar, B Dubno, GG Zimmermann, GS Gazelle and JF Debatin
Department of Radiology, University Hospital, Zurich, Switzerland.
PURPOSE: To evaluate the feasibility and accuracy of monitoring radio- frequency (RF) ablation with an open-configuration, 0.5-T magnetic resonance (MR) imager. MATERIALS AND METHODS: Thirty-six in vivo RF ablation experiments were performed in porcine paravertebral muscle (n = 24) and liver (n = 12). A 90 degrees C tip temperature was applied for 3-9 minutes. MR images were acquired after continuous or during intermittent RF application. Temperature changes were monitored as signal intensity and proton-frequency-shift (PFS) alterations in two T1- weighted gradient-echo sequences. An update image was obtained every 2.5 seconds (20/10 [repetition time msec/echo time [TE] msec]) or every 5.0 seconds (40/20). A color-coded subtraction technique enhanced the signal intensity and PFS changes. Macroscopic coagulation size was compared with MR image lesion size. RESULTS: The RF application mode had no significant effect on coagulation size in muscle or liver (P > .05). Twenty-two of 24 coagulative lesions in muscle and nine of 12 in liver were demonstrated with the PFS technique. Accuracy of lesion size determination depended on TE (TE = 20 msec, r = .95; TE = 10 msec, r = .78 [P < .01]). The T1-weighted technique depicted only six of 24 muscle and three of 12 liver lesion ablations. In the lesions depicted, macroscopic size was significantly underestimated (P < .001). CONCLUSION: PFS MR monitoring of RF effects in liver and muscle is feasible and accurate. The PFS technique outperformed the T1-weighted technique.
This article has been cited by other articles:
![]() |
S. Mulier, Y. Ni, L. Frich, F. Burdio, A. L. Denys, J.-F. De Wispelaere, B. Dupas, N. Habib, M. Hoey, M. C. Jansen, et al. Experimental and Clinical Radiofrequency Ablation: Proposal for Standardized Description of Coagulation Size and Geometry Ann. Surg. Oncol., April 1, 2007; 14(4): 1381 - 1396. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Kim, H. K. Lim, Y. H. Kim, W. J. Lee, S. J. Lee, S. H. Kim, J. H. Lim, and S. A. Kim Hepatocellular Carcinoma Treated with Radio-frequency Ablation: Spectrum of Imaging Findings RadioGraphics, January 1, 2003; 23(1): 107 - 121. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Aschoff, E. M. Merkle, S. N. Emancipator, C. A. Petersilge, J. L. Duerk, and J. S. Lewin Femur: MR Imaging-guided Radio-frequency Ablation in a Porcine Model— Feasibility Study Radiology, November 1, 2002; 225(2): 471 - 478. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Gazelle, S. N. Goldberg, L. Solbiati, and T. Livraghi Tumor Ablation with Radio-frequency Energy Radiology, December 1, 2000; 217(3): 633 - 646. [Abstract] [Full Text] |
||||
![]() |
E. M. Merkle, J. R. Haaga, J. L. Duerk, G. H. Jacobs, H.-J. Brambs, and J. S. Lewin MR Imaging-guided Radio-frequency Thermal Ablation in the Pancreas in a Porcine Model with a Modified Clinical C-Arm System Radiology, November 1, 1999; 213(2): 461 - 467. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |