In 11 volunteers, magnetic resonance (MR) imaging of the cervical spine was performed with a magnetization transfer preparatory pulse (prepulsed), three-dimensional Fourier transform, turbo field echo sequence. The effects of flip angle, number of shots, phase-encoding profile order, and magnetization transfer prepulse offset frequency on cerebrospinal fluid-to-cord contrast were evaluated. The contrast was improved by lowering the flip angle, increasing the number of shots, and implementing a magnetization transfer prepulse and linear phase- encoding profile order. Maximum myelographic effect was achieved with the magnetization prepulse (500-Hz frequency offset), 3 degrees flip angle, six shots, and linear phase-encoding profile order.
This article has been cited by other articles:
![]() |
V. Baskaran, F. S. Pereles, E. J. Russell, S. A. Georganos, A. Shaibani, K. A. Spero, E. A. Krupinski, A. Zhang, and J. P. Finn Myelographic MR Imaging of the Cervical Spine with a 3D True Fast Imaging with Steady-State Precession Technique: Initial Experience Radiology, May 1, 2003; 227(2): 585 - 592. [Abstract] [Full Text] [PDF] |
||||