Radiology, Vol 209, 397-403, Copyright © 1998 by Radiological Society of North America
ARTICLES |
G Starck, L Lonn, A Cederblad, M Alpsten, L Sjostrom and S Ekholm
Division of Medical Physics and Medical Engineering, Sahlgrenska University Hospital, MR Centre, Gothenburg, Sweden.
PURPOSE: To study the potential for radiation dose reduction in computed tomographic (CT) determination of adipose and muscle tissue area and volume, which is an application in which lower spatial resolution or higher image noise can be accepted. MATERIALS AND METHODS: Measurements in vivo (in a healthy man aged 52 years; weight, 80 kg) and in phantoms were performed with two standard clinical CT scanners. Tissue areas were determined, and measurement uncertainties were analyzed. Comparisons were made with simulated and theoretic data. RESULTS: A 25-fold reduction in radiation dose was achieved and resulted in an effective dose of 13 microSv per section; an accuracy of greater than 5% in area measurements was maintained. The dose savings achieved in this study were close to those that are theoretically possible. CONCLUSION: Further dose reduction to about 1 microSv may be possible for this application. Large dose reductions are possible in CT determinations of tissue area and volume, which makes this application available also in large randomized studies in which high-performance clinical standard scanners are used. This work demonstrates that modern CT technology may have a large potential for dose reduction in specific applications.
This article has been cited by other articles:
![]() |
J BRANDBERG, L LONN, E BERGELIN, L SJOSTROM, E FORSSELL-ARONSSON, and G STARCK Accurate tissue area measurements with considerably reduced radiation dose achieved by patient-specific CT scan parameters Br. J. Radiol., October 1, 2008; 81(970): 801 - 808. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Golding Multi-slice computed tomography (MSCT): the dose challenge of the new revolution Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 303 - 307. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Itoh, M. Ikeda, Y. Mori, K. Suzuki, A. Sawaki, S. Iwano, H. Satake, S. Arahata, T. Isomura, M. Ozaki, et al. Lung: Feasibility of a Method for Changing Tube Current during Low-Dose Helical CT Radiology, September 1, 2002; 224(3): 905 - 912. [Abstract] [Full Text] [PDF] |
||||
![]() |
G Starck, L Lonn, A Cederblad, E Forssell-Aronsson, L Sjostrom, and M Alpsten A method to obtain the same levels of CT image noise for patients of various sizes, to minimize radiation dose Br. J. Radiol., February 1, 2002; 75(890): 140 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
S J Golding and P C Shrimpton Radiation dose in CT: are we meeting the challenge? Br. J. Radiol., January 1, 2002; 75(889): 1 - 4. [Full Text] [PDF] |
||||
![]() |
W. Huda, E. M. Scalzetti, and G. Levin Technique Factors and Image Quality as Functions of Patient Weight at Abdominal CT Radiology, November 1, 2000; 217(2): 430 - 435. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |