Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Birnbaum, B. A.
Right arrow Articles by Glick, S. N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Birnbaum, B. A.
Right arrow Articles by Glick, S. N.
(Radiology. 1999;212:431-438.)
© RSNA, 1999


Gastrointestinal Imaging

Extravasation Detection Accessory: Clinical Evaluation in 500 Patients1

Bernard A. Birnbaum, MD, Rendon C. Nelson, MD, Judith L. Chezmar, MD and Seth N. Glick, MD

1 From the Departments of Radiology, Hospital of the University of Pennsylvania, 3400 Spruce St, Philadelphia, PA 19104 (B.A.B.); Duke University Medical Center, Durham, NC (R.C.N.); Emory University Hospital, Atlanta, Ga (J.L.C.); and Hahnemann University Medical Center, Philadelphia, Pa (S.N.G.). Received July 30, 1998; revision requested September 4; revision received October 29; accepted February 15, 1999. Address reprint requests to B.A.B. (e-mail: birnbaum@oasis.rad.upenn.edu).

PURPOSE: To assess the ability of an extravasation detection accessory (EDA) to detect clinically important (>=10 mL) extravascular injection of iodinated contrast material delivered with an automated power injector.

MATERIALS AND METHODS: Five hundred patients referred for contrast material–enhanced body computed tomography (CT) participated in a prospective, multiinstitutional, observational study in which the EDA was used to identify and interrupt any injection associated with clinically important extravasation. The presence or absence of extravasation was definitively established with helical CT at the injection site (injection rate, from 1.0 to 5.0 mL/sec; mean, 2.9 mL/sec; median, 3.0 mL/sec).

RESULTS: There were four true-positive (extravasation volumes, 13–18 mL), 484 true-negative, 12 false-positive, and no false-negative cases. The prevalence of overall and clinically important (>=10 mL) extravasation was 3.6% (18 of 500 cases) and 0.8% (four of 500 cases), respectively. The EDA had a sensitivity of 100% (four of four cases; 95% CI: 51%, 100%) and a specificity of 98% (484 of 496 cases; 95% CI: 96%, 99%) in the detection of clinically important extravasation.

CONCLUSION: The EDA is easy to use, safe, and accurate in the monitoring of intravenous injections for extravasation, which could prove especially useful in high-flow-rate CT applications.

Index terms: Angiography, complications, **.442 • Computed tomography (CT), contrast media • Contrast media, complications