Experimental Studies |
1 From the Mallinckrodt Institute of Technology, St Louis Children's Hospital, Washington University School of Medicine, 510 S Kingshighway Blvd, St Louis, MO 63110. Received October 7, 1998; revision requested November 11; revision received January 25, 1999; accepted March 2. S.D. supported in part by a grant from the Society for Pediatric Radiology Research and Education Foundation. Address reprint requests to S.D. (e-mail: don@mirlink.wustl.edu).
PURPOSE: To determine if computed radiography is equivalent to screen-film radiography in depicting pulmonary edema and to determine if radiation exposure can be reduced with computed radiography while maintaining equivalent diagnostic accuracy for pulmonary edema.
MATERIALS AND METHODS: Oleic acid was intravenously injected into three rabbits at each of four doses: 0, 0.02, 0.04, and 0.06 mL/kg. Two hours later, chest computed radiographs and screen-film radiographs were obtained at 60 kVp and 1.1 mAs. Additional computed radiographs were obtained after reducing milliampere seconds or by reducing milliampere seconds and increasing the kilovolt peak, which reduced bone marrow exposure by up to 20%. The presence of pulmonary opacities, "truth," was established by the wet-dry weight ratio and by chest computed tomography (CT). The radiographs were masked and randomized. Four observers rated the images for the presence of parenchymal opacities with a dichotomous score and judged the quality of the radiographs on a scale from 1 (worst) to 6 (best). Cochran Q tests and McNemar tests were used to analyze the differences in paired comparisons. Image quality was evaluated with logistic regression analysis.
RESULTS: There was no significant difference between truth and observer ability to detect opacity for either modality or for any exposure (P > .05). There was no significant difference between computed radiography and screen-film radiography for image quality (P > .05).
CONCLUSION: Computed radiography is equivalent to screen-film radiography in the detection of pulmonary edema. Radiation exposure reduction of 20% can be achieved without affecting pulmonary edema detection or image quality.
Index terms: Computed tomography (CT), comparative studies, 60.1211 Lung, fluid, 60.711 Radiography, comparative studies, 60.1211, 60.1215 Radiography, in infants and children, 60.1215 Radiography, storage phosphor, 60.1215
This article has been cited by other articles:
![]() |
S. Don, B. R. Whiting, J. S. Ellinwood, D. H. Foos, K. A. Kronemer, and R. A. Kraus Neonatal Chest Computed Radiography: Image Processing and Optimal Image Display Am. J. Roentgenol., April 1, 2007; 188(4): 1138 - 1144. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. S. Kim, J.-G. Im, J. M. Goo, K. H. Lee, Y.-J. Lee, S. H. Kim, and S. Kim Detection of Pulmonary Edema in Pigs: Storage Phosphor versus Amorphous Selenium-based Flat-Panel-Detector Radiography Radiology, June 1, 2002; 223(3): 695 - 701. [Abstract] [Full Text] [PDF] |
||||
![]() |
J V Cook Radiation protection and quality assurance in paediatric radiology Imaging, December 1, 2001; 13(4): 229 - 238. [Abstract] [Full Text] [PDF] |
||||