Thoracic Imaging |
1 From the Departments of Radiology (S.I., S.A., T. Kodaira, T. Isomura, K.M., T. Ishigaki) and Medical Information and Medical Records (M.I.), Nagoya University School of Medicine, 65 Turumai-cho Showa-ku, 466-0065 Nagoya, Japan; the Department of Biofunctional Research, National Institute for Longevity Science, Oubu, Japan (T. Kato); and the Department of Radiology, Tosei General Hospital, Seto, Japan (K.Y.). From the 1998 RSNA scientific assembly. Received November 10, 1998; revision requested January 21, 1999; final revision received September 1; accepted September 9. Address reprint requests to S.I.
PURPOSE: To investigate the minimum tube current required for helical computed tomography (CT) for lung cancer screening.
MATERIALS AND METHODS: Thirty helical scans of the lung were obtained at effective tube currents of 50, 30, 20, 18, 12, 10, and 6 mAs in seven healthy volunteers. Computer-generated nodules 6 mm in diameter that showed ground-glass opacity were superimposed on the images. The image quality and detectability of nodules were evaluated subjectively by six observers. The SDs of measured CT numbers were calculated. The results were analyzed according to location in the lung.
RESULTS: Compared with the subjective quality of images obtained at 50 mAs, the subjective quality of images obtained at 20 mAs was not significantly different. The detectability of nodules was not significantly degraded by reducing the tube current to 20 mAs in the upper zone of the lung, to 12 mAs in the middle zone, or to 18 mAs in the lower zone. The SDs at the apex and base of the lung were larger than those at other levels, and the difference became greater as the dose was reduced.
CONCLUSION: The minimum tube current required for screening helical CT differs for different locations in the lung. An ideal CT protocol for the lung should permit the tube current to be changed during helical scanning.
Index terms: Cancer screening, 60.12115, 60.32 Computed tomography (CT), helical, 60.12115 Computed tomography (CT), image quality, 60.12115 Lung neoplasms, CT, 60.12115, 60.32 Lung neoplasms, diagnosis, 60.12115, 60.32
This article has been cited by other articles:
![]() |
T. Kubo, P.-J. P. Lin, W. Stiller, M. Takahashi, H.-U. Kauczor, Y. Ohno, and H. Hatabu Radiation Dose Reduction in Chest CT: A Review Am. J. Roentgenol., February 1, 2008; 190(2): 335 - 343. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A. Gietema, A. M. Schilham, B. van Ginneken, R. J. van Klaveren, J. W. J. Lammers, and M. Prokop Monitoring of Smoking-induced Emphysema with CT in a Lung Cancer Screening Setting: Detection of Real Increase in Extent of Emphysema Radiology, September 1, 2007; 244(3): 890 - 897. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Barentsz, S. Takahashi, W. Oyen, R. Mus, P. De Mulder, R. Reznek, M. Oudkerk, and W. Mali Commonly Used Imaging Techniques for Diagnosis and Staging J. Clin. Oncol., July 10, 2006; 24(20): 3234 - 3244. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Mulkens, P. Bellinck, M. Baeyaert, D. Ghysen, X. Van Dijck, E. Mussen, C. Venstermans, and J.-L. Termote Use of an Automatic Exposure Control Mechanism for Dose Optimization in Multi-Detector Row CT Examinations: Clinical Evaluation Radiology, October 1, 2005; 237(1): 213 - 223. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Remy-Jardin, A. Sobaszek, A. Duhamel, I. Mastora, C. Zanetti, and J. Remy Asbestos-related Pleuropulmonary Diseases: Evaluation with Low-Dose Four-Detector Row Spiral CT Radiology, October 1, 2004; 233(1): 182 - 190. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Mastora, M. Remy-Jardin, V. Delannoy, A. Duhamel, C. Scherf, C. Suess, and J. Remy Multi-Detector Row Spiral CT Angiography of the Thoracic Outlet: Dose Reduction with Anatomically Adapted Online Tube Current Modulation and Preset Dose Savings Radiology, January 1, 2004; 230(1): 116 - 124. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Ko, H. Rusinek, E. L. Jacobs, J. S. Babb, M. Betke, G. McGuinness, and D. P. Naidich Small Pulmonary Nodules: Volume Measurement at Chest CT--Phantom Study Radiology, September 1, 2003; 228(3): 864 - 870. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Tack, V. De Maertelaer, and P. A. Gevenois Dose Reduction in Multidetector CT Using Attenuation-Based Online Tube Current Modulation Am. J. Roentgenol., August 1, 2003; 181(2): 331 - 334. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Mayo, J. Aldrich, and N. L. Muller Radiation Exposure at Chest CT: A Statement of the Fleischner Society Radiology, July 1, 2003; 228(1): 15 - 21. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Itoh, M. Ikeda, Y. Mori, K. Suzuki, A. Sawaki, S. Iwano, H. Satake, S. Arahata, T. Isomura, M. Ozaki, et al. Lung: Feasibility of a Method for Changing Tube Current during Low-Dose Helical CT Radiology, September 1, 2002; 224(3): 905 - 912. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Ravenel, E. M. Scalzetti, W. Huda, and W. Garrisi Radiation Exposure and Image Quality in Chest CT Examinations Am. J. Roentgenol., August 1, 2001; 177(2): 279 - 284. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Haaga Radiation Dose Management: Weighing Risk Versus Benefit Am. J. Roentgenol., August 1, 2001; 177(2): 289 - 291. [Full Text] [PDF] |
||||