1 From the Department of Radiology, Brigham and Women's Hospital, 75 Francis St, Boston, MA 02115. From the 1998 RSNA scientific assembly. Received July 22; revision requested September 2; revision received October 21; accepted November 22. Address correspondence to R.D.N. (e-mail: nawfel@bwh.harvard.edu).
PURPOSE: To estimate patient dose and personnel exposure from phantom measurements during computed tomographic (CT) fluoroscopy, to use the estimates to provide users with dose information, and to recommend methods to reduce exposure.
MATERIALS AND METHODS: Surface dose was estimated on a CT dosimetric phantom by using thermoluminescent dosimetric (TLD) and CT pencil chamber measurements. Scatter exposure was estimated from scattered radiation measured at distances of 10 cm to 1 m from the phantom. Scatter exposures measured with and without placement of a lead drape on the phantom surface adjacent to the scanning plane were compared.
RESULTS: Phantom surface dose rates ranged from 2.3 to 10.4 mGy/sec. Scattered exposure rates for a commonly used CT fluoroscopic technique (120 kVp, 50 mA, 10-mm section thickness) were 27 and 1.2 µGy/sec at 10 cm and 1 m, respectively, from the phantom. Lead drapes reduced the scattered exposure by approximately 71% and 14% at distances of 10 and 60 cm from the scanning plane, respectively.
CONCLUSION: High exposures to patients and personnel may occur during CT fluoroscopyguided interventions. Radiation exposure to patients and personnel may be reduced by modifying CT scanning techniques and by limiting fluoroscopic time. In addition, scatter exposure to personnel may be substantially reduced by placing a lead drape adjacent to the scanning plane.
Index terms: Computed tomography (CT), guidance, 70.12119 Computed tomography (CT), radiation exposure, 70.12119 Fluoroscopy Phantoms Radiations, exposure to patients and personnel, 70.12119 Radiations, measurement, 70.12119 Radiations, protective and therapeutic agents and devices, 70.12119
Related Article
Radiology 2000 216: 9-10.
This article has been cited by other articles:
![]() |
J. Fritz, C. Thomas, S. Clasen, C. D. Claussen, J. S. Lewin, and P. L. Pereira Freehand Real-Time MRI-Guided Lumbar Spinal Injection Procedures at 1.5 T: Feasibility, Accuracy, and Safety Am. J. Roentgenol., April 1, 2009; 192(4): W161 - W167. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. A. Tsalafoutas, V. Tsapaki, C. Triantopoulou, C. Pouli, V. Kouridou, I. Fagadaki, and J. Papailiou Comparison of Measured and Calculated Skin Doses in CT-Guided Interventional Procedures Am. J. Roentgenol., November 1, 2008; 191(5): 1601 - 1607. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hohl, C. Suess, J. E. Wildberger, D. Honnef, M. Das, G. Muhlenbruch, A. Schaller, R. W. Gunther, and A. H. Mahnken Dose Reduction during CT Fluoroscopy: Phantom Study of Angular Beam Modulation Radiology, February 1, 2008; 246(2): 519 - 525. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Das, F. Sauer, U. J. Schoepf, A. Khamene, S. K. Vogt, S. Schaller, R. Kikinis, E. vanSonnenberg, and S. G. Silverman Augmented Reality Visualization for CT-guided Interventions: System Description, Feasibility, and Initial Evaluation in an Abdominal Phantom Radiology, July 1, 2006; 240(1): 230 - 235. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Carlson, J. P. Felmlee, C. E. Bender, R. L. Ehman, K. L. Classic, T. L. Hoskin, W. S. Harmsen, and H. H. Hu CT Fluoroscopy-guided Biopsy of the Lung or Upper Abdomen with a Breath-hold Monitoring and Feedback System: A Prospective Randomized Controlled Clinical Trial Radiology, November 1, 2005; 237(2): 701 - 708. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Aviles Lucas, D. R. Dance, I. A. Castellano, and E. Vano Estimation of the peak entrance surface air kerma for patients undergoing computed tomography-guided procedures Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 317 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Meleka, A. Patra, E. Minkoff, and K. Murphy Value of CT Fluoroscopy for Lumbar Facet Blocks AJNR Am. J. Neuroradiol., May 1, 2005; 26(5): 1001 - 1003. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Wagner Selective Lumbar Nerve Root Blocks with CT Fluoroscopic Guidance: Technique, Results, Procedure Time, and Radiation Dose AJNR Am. J. Neuroradiol., October 1, 2004; 25(9): 1592 - 1594. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Nawfel, P. F. Judy, A. R. Schleipman, and S. G. Silverman Patient Radiation Dose at CT Urography and Conventional Urography Radiology, July 1, 2004; 232(1): 126 - 132. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Maher, D. A. Gervais, M. K. Kalra, B. Lucey, D. V. Sahani, R. Arellano, P. F. Hahn, and P. R. Mueller The Inaccessible or Undrainable Abscess: How to Drain It RadioGraphics, May 1, 2004; 24(3): 717 - 735. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Carlson, J. P. Felmlee, C. E. Bender, R. L. Ehman, K. L. Classic, H. H. Hu, and T. L. Hoskin Intermittent-Mode CT Fluoroscopy-guided Biopsy of the Lung or Upper Abdomen with Breath-hold Monitoring and Feedback: System Development and Feasibility Radiology, December 1, 2003; 229(3): 906 - 912. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Binkert, F. R. Verdun, M. Zanetti, C. W. Pfirrmann, and J. Hodler CT Arthrography of the Glenohumeral Joint: CT Fluoroscopy Versus Conventional CT and Fluoroscopy--Comparison of Image-Guidance Techniques Radiology, October 1, 2003; 229(1): 153 - 158. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Sakarya, O. Unal, B. Ozbay, K. Uzun, I. Kati, S. Ozen, and O. Etlik MR Fluoroscopy-guided Transthoracic Fine-Needle Aspiration Biopsy: Feasibility Radiology, August 1, 2003; 228(2): 589 - 592. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. B. Solomon, A. Patriciu, M. E. Bohlman, L. R. Kavoussi, and D. Stoianovici Robotically Driven Interventions: A Method of Using CT Fluoroscopy without Radiation Exposure to the Physician Radiology, October 1, 2002; 225(1): 277 - 282. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. K. Paulson, D. H. Sheafor, D. S. Enterline, H. P. McAdams, and T. T. Yoshizumi CT Fluoroscopy-guided Interventional Procedures: Techniques and Radiation Dose to Radiologists Radiology, July 1, 2001; 220(1): 161 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. Wagner CT Fluoroscopy: Another Advancement with Additional Challenges in Radiation Management Radiology, July 1, 2000; 216(1): 9 - 10. [Full Text] |
||||