1 From the Department of Radiology, School of Medicine, Gazi University, Besevler 06510, Ankara, Turkey (E.T.I., I.O., O.K., S.I.); the Departments of Physics (N.M.) and Engineering Physics (D.B.), Faculty of Science, the University of Ankara, Turkey. Received November 3, 1999; revision requested December 15; revision received March 16, 2000; accepted March 20. Address correspondence to E.T.I. (e-mail: erhanti@med.gazi.edu.tr).
PURPOSE: To evaluate the absorbed radiation dose to the lens of the eye, which is the critical organ in the primary beam during fluoroscopically guided transluminal balloon dilation of the lacrimal drainage system (balloon dacryocystoplasty) for obstructive epiphora and to evaluate the possibility of deterministic radiation effect on the lens.
MATERIALS AND METHODS: The radiation dose to the lens of the eye during balloon dacryocystoplasty (which includes pre- and postintervention dacryocystography) was measured in 10 consecutive patients by using thermoluminescent dosimeters on the lids of both eyes as close as possible to the lenses. A C-arm angiographic unit coupled with a digital imaging system was used, with similar exposure and geometric parameters in all cases.
RESULTS: The mean radiation dose to the lens of the treated eye was 4.6 mGy ± 2.2 (dose range, 1.99.1 mGy) and to that of the untreated eye was 38.5 mGy ± 17.5 (dose range, 14.767.8 mGy).
CONCLUSION: The lens of the untreated eye receives a higher dose than that of the treated eye because of its closer proximity to the x-ray tube in a lateral projection. In the lens, even the highest measured radiation dose (67.8 mGy) still was well below the deterministic threshold for lens opacity and cataract formation.
Index terms: Lacrimal gland and duct, 223.1295, 223.1296 Lacrimal gland and duct, interventional procedures, 223.1295, 223.1296 Radiation, exposure to patients and personnel, 223.1295, 223.1296
This article has been cited by other articles:
![]() |
K. E. Wilhelm, E. T. Ilgit, D. Bor, and N. Meric Radiation Dose in Fluoroscopically Guided Dacryocystoplasty Dr Ilgit and colleagues respond: Radiology, May 1, 2001; 219(2): 577 - 578. [Full Text] |
||||
![]() |
J. M. Boone, T. R. Nelson, K. K. Lindfors, and J. A. Seibert Dedicated Breast CT: Radiation Dose and Image Quality Evaluation Radiology, December 1, 2001; 221(3): 657 - 667. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |