Vascular and Interventional Radiology |
1 From the Department of Radiology, University of Regensburg, Klinikum der Universität, Franz-Josef-Strauss-Allee 11, D-93042 Regensburg, Germany. Received March 31, 2000; revision requested May 23; revision received September 12; accepted September 19. Address correspondence to C.M. (e-mail: chmanke@web.de).
PURPOSE: To assess the feasibility of magnetic resonance (MR) imagingguided stent placement in iliac arterial stenoses.
MATERIALS AND METHODS: Thirteen patients with 14 iliac arterial stenoses were examined prospectively. Angioplasty was performed through a femoral sheath by using a conventional 1.5-T MR imaging system. Stents and catheters were visualized on the basis of their artifacts. Nitinol stents were placed with gradient-echo MR imaging guidance. Angioplasty balloons were inflated with gadolinium-based contrast material. Results were evaluated clinically and with both digital subtraction angiography (DSA) and contrast materialenhanced MR angiography.
RESULTS: Ten of 13 patients were treated with technical success by using MR imagingguided intervention alone. Three patients were treated with additional fluoroscopic guidance, because complications (ie, panic attack, subintimal recanalization, and stent misplacement) occurred with MR guidance. The quality of the postinterventional contrast-enhanced MR angiograms of three of 12 lesions with stents was limited owing to stent-induced signal loss of the lumen. The mean stenosis degree after stent placement was significantly higher at contrast-enhanced MR angiography than at DSA (24.6% vs 6.2%). The mean MR imagingguided procedure time was 74 minutes.
CONCLUSION: MR imagingguided stent placement in iliac arteries is feasible in select patients. The presented technique has limitationsthat is, long procedure times, lack of real-time monitoring, and stent artifactsthat necessitate further modifications before it can be recommended for clinical use.
Index terms: Angiography, comparative studies, 984.122, 984.12942 Arteries, grafts and prostheses, 984.1268, 984.1286 Arteries, stenosis or obstruction, 984.72 Arteries, transluminal angioplasty, 984.1282, 984.1286 Magnetic resonance (MR), vascular studies, 984.129412, 984.129416, 984.12942, 984.12943
This article has been cited by other articles:
![]() |
F. Poschenrieder, O. W. Hamer, T. Herold, T. Schleicher, I. Borisch, S. Feuerbach, and N. Zorger Diagnostic Accuracy of Intraarterial and IV MR Angiography for the Detection of Stenoses of the Infrainguinal Arteries Am. J. Roentgenol., January 1, 2009; 192(1): 117 - 121. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Blum, M. Schmook, R. Schernthaner, G. Edelhauser, S. Puchner, J. Lammer, and M. A. Funovics Quantification and Detectability of In-Stent Stenosis with CT Angiography and MR Angiography in Arterial Stents In Vitro Am. J. Roentgenol., November 1, 2007; 189(5): 1238 - 1242. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Paetzel, N Zorger, O W Hamer, J Seitz, T Schleicher, S Feuerbach, W R Nitz, M Lenhart, and T Herold Intra-arterial MR angiography in the iliac system: initial clinical experience with 25 patients. Br. J. Radiol., April 1, 2006; 79(940): 298 - 302. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Eggebrecht, H. Kuhl, G. M. Kaiser, S. Aker, M. O. Zenge, F. Stock, F. Breuckmann, F. Grabellus, M. E. Ladd, R. H. Mehta, et al. Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection Eur. Heart J., March 1, 2006; 27(5): 613 - 620. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Raval, P. V. Karmarkar, M. A. Guttman, C. Ozturk, S. Sampath, R. DeSilva, R. J. Aviles, M. Xu, V. J. Wright, W. H. Schenke, et al. Real-Time Magnetic Resonance Imaging-Guided Endovascular Recanalization of Chronic Total Arterial Occlusion in a Swine Model Circulation, February 28, 2006; 113(8): 1101 - 1107. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Lederman Cardiovascular Interventional Magnetic Resonance Imaging Circulation, November 8, 2005; 112(19): 3009 - 3017. [Full Text] [PDF] |
||||
![]() |
A. N. Raval, J. D. Telep, M. A. Guttman, C. Ozturk, M. Jones, R. B. Thompson, V. J. Wright, W. H. Schenke, R. DeSilva, R. J. Aviles, et al. Real-Time Magnetic Resonance Imaging-Guided Stenting of Aortic Coarctation With Commercially Available Catheter Devices in Swine Circulation, August 2, 2005; 112(5): 699 - 706. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kuehne, S. Weiss, F. Brinkert, J. Weil, S. Yilmaz, B. Schmitt, P. Ewert, P. Lange, and M. Gutberlet Catheter Visualization with Resonant Markers at MR Imaging-guided Deployment of Endovascular Stents in Swine Radiology, December 1, 2004; 233(3): 774 - 780. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. G. Shellock and J. V. Crues MR Procedures: Biologic Effects, Safety, and Patient Care Radiology, September 1, 2004; 232(3): 635 - 652. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. K. Wacker, D. Elgort, C. M. Hillenbrand, J. L. Duerk, and J. S. Lewin The Catheter-Driven MRI Scanner: A New Approach to Intravascular Catheter Tracking and Imaging-Parameter Adjustment for Interventional MRI Am. J. Roentgenol., August 1, 2004; 183(2): 391 - 395. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Olin, J. A. Kaufman, D. A. Bluemke, R. O. Bonow, M. D. Gerhard, M. R. Jaff, G. D. Rubin, and W. Hall Atherosclerotic Vascular Disease Conference: Writing Group IV: Imaging Circulation, June 1, 2004; 109(21): 2626 - 2633. [Full Text] [PDF] |
||||
![]() |
F. K. Wacker, K. Reither, W. Ebert, M. Wendt, J. S. Lewin, and K.-J. Wolf MR Image-guided Endovascular Procedures with the Ultrasmall Superparamagnetic Iron Oxide SH U 555 C as an Intravascular Contrast Agent: Study in Pigs Radiology, February 1, 2003; 226(2): 459 - 464. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kuehne, M. Saeed, C. B. Higgins, K. Gleason, G. A. Krombach, O. M. Weber, A. J. Martin, D. Turner, D. Teitel, and P. Moore Endovascular Stents in Pulmonary Valve and Artery in Swine: Feasibility Study of MR Imaging-guided Deployment and Postinterventional Assessment Radiology, February 1, 2003; 226(2): 475 - 481. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Tello, K. R. Thomson, and C. Manke Intraarterial MR Angioplasty in Iliac Arterial Stenosis * Dr Manke responds: Radiology, March 1, 2002; 222(3): 859 - 859. [Full Text] [PDF] |
||||
![]() |
E. Spuentrup, A. Ruebben, T. Schaeffter, W. J. Manning, R. W. Gunther, and A. Buecker Magnetic Resonance-Guided Coronary Artery Stent Placement in a Swine Model Circulation, February 19, 2002; 105(7): 874 - 879. [Abstract] [Full Text] [PDF] |
||||