1 From the Institute of Clinical Radiology (C.H., C.R.B., A.H., U.J.S., R.B., M.F.R.) and Division of Cardiology (A.K.), Klinikum Grosshadern, University of Munich, Marchioninistrasse 15, 81377 Munich, Germany; Siemens Medical Systems, Forchheim, Germany (B.O.); and the Department of Radiology, Tongji Medical University, Tongji, China (C.H.). From the 2000 RSNA scientific assembly. Received November 30, 2000; revision requested January 11, 2001; revision received March 12; accepted March 22. Address correspondence to C.R.B. (e-mail: christoph.becker@ikra.med.uni-muenchen.de).
PURPOSE: To evaluate the effectiveness of electrocardiographically (ECG)-gated retrospective image reconstruction for multidetector row computed tomographic (CT) coronary angiography in reducing cardiac motion artifacts and to evaluate the influence of heart rate on cardiac image quality.
MATERIALS AND METHODS: Sixty-five patients with different heart rates underwent coronary CT angiography. Raw helical CT data and ECG tracings were combined to retrospectively reconstruct at the defined consecutive z position with a temporal resolution of 250 msec per section. The starting points of the reconstruction were chosen between 30% and 80% of the R-R intervals. The relationships between heart rate, trigger delay, and image quality were analyzed.
RESULTS: Optimal image quality was achieved with a 50% trigger delay for the right coronary artery and 60% for the left circumflex coronary artery. Optimal image quality for the left anterior descending coronary artery was equally obtained at 50% and 60% triggering. A significant negative correlation was observed between heart rate and image quality (P < .05). The best image quality was achieved when the heart rate was less than 74.5 beats per minute.
CONCLUSION: To achieve high image quality, the heart rate should be sufficiently slow. Selection of appropriate trigger delays and a decreasing heart rate are effective to reduce cardiac motion artifacts.
Index terms: Coronary vessels, CT, 54.12111, 54.12115, 54.12116 Heart, CT, 54.12111, 54.12115, 54.12116
This article has been cited by other articles:
![]() |
Y. Okumura, B. D. Henz, S. B. Johnson, T. J. Bunch, C. J. O'Brien, D. O. Hodge, A. Altman, A. Govari, and D. L. Packer Three-Dimensional Ultrasound for Image-Guided Mapping and Intervention: Methods, Quantitative Validation, and Clinical Feasibility of a Novel Multimodality Image Mapping System Circ Arrhythmia Electrophysiol, June 1, 2008; 1(2): 110 - 119. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Earls, E. L. Berman, B. A. Urban, C. A. Curry, J. L. Lane, R. S. Jennings, C. C. McCulloch, J. Hsieh, and J. H. Londt Prospectively Gated Transverse Coronary CT Angiography versus Retrospectively Gated Helical Technique: Improved Image Quality and Reduced Radiation Dose Radiology, March 1, 2008; 246(3): 742 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Matt, H. Scheffel, S. Leschka, T. G. Flohr, B. Marincek, P. A. Kaufmann, and H. Alkadhi Dual-Source CT Coronary Angiography: Image Quality, Mean Heart Rate, and Heart Rate Variability Am. J. Roentgenol., September 1, 2007; 189(3): 567 - 573. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. M. Kroft, A. de Roos, and J. Geleijns Artifacts in ECG-Synchronized MDCT Coronary Angiography Am. J. Roentgenol., September 1, 2007; 189(3): 581 - 591. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. J. Schoepf, P. L. Zwerner, G. Savino, C. Herzog, J. M. Kerl, and P. Costello Coronary CT Angiography Radiology, July 1, 2007; 244(1): 48 - 63. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Herzog, S. A. Nguyen, G. Savino, P. L. Zwerner, J. Doll, C. D. Nielsen, T. G. Flohr, T. J. Vogl, P. Costello, and U. J. Schoepf Does Two-Segment Image Reconstruction at 64-Section CT Coronary Angiography Improve Image Quality and Diagnostic Accuracy? Radiology, July 1, 2007; 244(1): 121 - 129. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Rispler, Z. Keidar, E. Ghersin, A. Roguin, A. Soil, R. Dragu, D. Litmanovich, A. Frenkel, D. Aronson, A. Engel, et al. Integrated Single-Photon Emission Computed Tomography and Computed Tomography Coronary Angiography for the Assessment of Hemodynamically Significant Coronary Artery Lesions J. Am. Coll. Cardiol., March 13, 2007; 49(10): 1059 - 1067. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Jones, T. Athanasiou, N. Dunne, J. Kirby, O. Aziz, A. Haq, C. Rao, V. Constantinides, S. Purkayastha, and A. Darzi Multi-Detector Computed Tomography in Coronary Artery Bypass Graft Assessment: A Meta-Analysis Ann. Thorac. Surg., January 1, 2007; 83(1): 341 - 348. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. R. C. Johnson, K. Nikolaou, B. J. Wintersperger, A. Knez, P. Boekstegers, M. F. Reiser, and C. R. Becker ECG-Gated 64-MDCT Angiography in the Differential Diagnosis of Acute Chest Pain Am. J. Roentgenol., January 1, 2007; 188(1): 76 - 82. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, S. Wildermuth, T. Boehm, L. Desbiolles, L. Husmann, A. Plass, P. Koepfli, T. Schepis, B. Marincek, P. A. Kaufmann, et al. Noninvasive Coronary Angiography with 64-Section CT: Effect of Average Heart Rate and Heart Rate Variability on Image Quality Radiology, November 1, 2006; 241(2): 378 - 385. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Jones, T. Athanasiou, N. Dunne, J. Kirby, S. Attaran, A. Chow, S. Purkayastha, and A. Darzi Multi-slice computed tomography in coronary artery disease. Eur. J. Cardiothorac. Surg., September 1, 2006; 30(3): 443 - 450. [Abstract] [Full Text] [PDF] |
||||
![]() |
A S Lowe and C L Kay Recent developments in CT: a review of the clinical applications and advantages of multidetector computed tomography Imaging, June 1, 2006; 18(2): 62 - 67. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. D. Raptopoulos, P. B. Boiselle, N. Michailidis, J. Handwerker, A. Sabir, J. A. Edlow, I. Pedrosa, and J. B. Kruskal MDCT Angiography of Acute Chest Pain: Evaluation of ECG-Gated and Nongated Techniques Am. J. Roentgenol., June 1, 2006; 186(6_Supplement_2): S346 - S356. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Herzog, M. Arning-Erb, S. Zangos, K. Eichler, R. Hammerstingl, S. Dogan, H. Ackermann, and T. J. Vogl Multi-Detector Row CT Coronary Angiography: Influence of Reconstruction Technique and Heart Rate on Image Quality Radiology, January 1, 2006; 238(1): 75 - 86. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hacker, T. Jakobs, F. Matthiesen, C. Vollmar, K. Nikolaou, C. Becker, A. Knez, T. Pfluger, M. Reiser, K. Hahn, et al. Comparison of Spiral Multidetector CT Angiography and Myocardial Perfusion Imaging in the Noninvasive Detection of Functionally Relevant Coronary Artery Lesions: First Clinical Experiences J. Nucl. Med., August 1, 2005; 46(8): 1294 - 1300. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, H. Yamamoto, Y. Akiyama, N. Hirai, K. Marukawa, H. Fukuda, and K. Ito Variability of Repeated Coronary Artery Calcium Measurements by 16-MDCT with Retrospective Reconstruction Am. J. Roentgenol., June 1, 2005; 184(6): 1917 - 1923. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. G. Flohr, S. Schaller, K. Stierstorfer, H. Bruder, B. M. Ohnesorge, and U. J. Schoepf Multi-Detector Row CT Systems and Image-Reconstruction Techniques Radiology, June 1, 2005; 235(3): 756 - 773. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. P. Lawler, H. K. Pannu, and E. K. Fishman MDCT Evaluation of the Coronary Arteries, 2004: How We Do It--Data Acquisition, Postprocessing, Display, and Interpretation Am. J. Roentgenol., May 1, 2005; 184(5): 1402 - 1412. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-F. Ko, M.-J. Hsieh, M.-C. Chen, S.-H. Ng, F.-M. Fang, C.-C. Huang, Y.-L. Wan, and T.-Y. Lee Effects of Heart Rate on Motion Artifacts of the Aorta on Non-ECG-Assisted 0.5-Sec Thoracic MDCT Am. J. Roentgenol., April 1, 2005; 184(4): 1225 - 1230. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Shim, Y. Kim, and S. M. Lim Improvement of Image Quality with {beta}-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography Am. J. Roentgenol., February 1, 2005; 184(2): 649 - 654. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yamamuro, E. Tadamura, S. Kubo, H. Toyoda, T. Nishina, M. Ohba, R. Hosokawa, T. Kimura, N. Tamaki, M. Komeda, et al. Cardiac Functional Analysis with Multi-Detector Row CT and Segmental Reconstruction Algorithm: Comparison with Echocardiography, SPECT, and MR Imaging Radiology, February 1, 2005; 234(2): 381 - 390. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. K. Hoffmann, H. Shi, R. Manzke, F. T. Schmid, L. De Vries, M. Grass, H.-J. Brambs, and A. J. Aschoff Noninvasive Coronary Angiography with 16-Detector Row CT: Effect of Heart Rate Radiology, January 1, 2005; 234(1): 86 - 97. [Abstract] [Full Text] [PDF] |
||||
![]() |
A F Kopp, A Kuttner, T Trabold, M Heuschmid, S Schroder, and C D Claussen Multislice CT in cardiac and coronary angiography Br. J. Radiol., December 1, 2004; 77(suppl_1): S87 - S97. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. Hofmann, K. H. Zou, P. Costello, and U. J. Schoepf Electrocardiographically Gated 16-Section CT of the Thorax: Cardiac Motion Suppression Radiology, December 1, 2004; 233(3): 927 - 933. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Lu, N. Zhuang, S.-S. Mao, J. Child, S. Carson, and M. J. Budoff Baseline Heart Rate-adjusted Electrocardiographic Triggering for Coronary Artery Electron-Beam CT Angiography Radiology, November 1, 2004; 233(2): 590 - 595. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hong, G. S. Chrysant, P. K. Woodard, and K. T. Bae Coronary Artery Stent Patency Assessed with In-Stent Contrast Enhancement Measured at Multi-Detector Row CT Angiography: Initial Experience Radiology, October 1, 2004; 233(1): 286 - 291. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kuettner, T. Trabold, S. Schroeder, A. Feyer, T. Beck, A. Brueckner, M. Heuschmid, C. Burgstahler, A. F. Kopp, and C. D. Claussen Noninvasive detection of coronary lesions using 16-detector multislice spiral computed tomography technology: Initial clinical results J. Am. Coll. Cardiol., September 15, 2004; 44(6): 1230 - 1237. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. K. Willmann, D. Weishaupt, R. Kobza, F. R. Verdun, B. Seifert, B. Marincek, and T. Boehm Coronary Artery Bypass Grafts: ECG-gated Multi-Detector Row CT Angiography--Influence of Image Reconstruction Interval on Graft Visibility Radiology, August 1, 2004; 232(2): 568 - 577. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Schoenhagen, S. S. Halliburton, A. E. Stillman, S. A. Kuzmiak, S. E. Nissen, E. M. Tuzcu, and R. D. White Noninvasive Imaging of Coronary Arteries: Current and Future Role of Multi-Detector Row CT Radiology, July 1, 2004; 232(1): 7 - 17. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. J. Schoepf, C. R. Becker, B. M. Ohnesorge, and E. K. Yucel CT of Coronary Artery Disease Radiology, July 1, 2004; 232(1): 18 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, H. Yamamoto, Y. Akiyama, K. Marukawa, N. Hirai, and K. Ito Coronary Artery Calcium Scoring Using 16-MDCT and a Retrospective ECG-Gating Reconstruction Algorithm Am. J. Roentgenol., July 1, 2004; 183(1): 103 - 108. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. S. Choi, B. W. Choi, K. O. Choe, D. Choi, K.-J. Yoo, M.-I. Kim, and J. Kim Pitfalls, Artifacts, and Remedies in Multi- Detector Row CT Coronary Angiography RadioGraphics, May 1, 2004; 24(3): 787 - 800. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. W. Leber, A. Knez, A. Becker, C. Becker, F. von Ziegler, K. Nikolaou, C. Rist, M. Reiser, C. White, G. Steinbeck, et al. Accuracy of multidetector spiral computed tomography in identifying and differentiating the composition of coronary atherosclerotic plaques: A comparative study with intracoronary ultrasound J. Am. Coll. Cardiol., April 7, 2004; 43(7): 1241 - 1247. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Herzog, S. Dogan, T. Diebold, M. F. Khan, H. Ackermann, S. Schaller, T. G. Flohr, G. Wimmer-Greinecker, A. Moritz, and T. J. Vogl Multi-Detector Row CT versus Coronary Angiography: Preoperative Evaluation before Totally Endoscopic Coronary Artery Bypass Grafting Radiology, October 1, 2003; 229(1): 200 - 208. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. R. Van Hoe, K. G. De Meerleer, P. Ph. Leyman, and P. K. Vanhoenacker Coronary Artery Calcium Scoring Using ECG-Gated Multidetector CT: Effect of Individually Optimized Image-Reconstruction Windows on Image Quality and Measurement Reproducibility Am. J. Roentgenol., October 1, 2003; 181(4): 1093 - 1100. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. K. Pannu, T. G. Flohr, F. M. Corl, and E. K. Fishman Current Concepts in Multi-Detector Row CT Evaluation of the Coronary Arteries: Principles, Techniques, and Anatomy RadioGraphics, October 1, 2003; 23(90001): S111 - 125. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. Morgan-Hughes, P. E. Owens, A. J. Marshall, and C. A. Roobottom Thoracic Aorta at Multi-Detector Row CT: Motion Artifact with Various Reconstruction Windows Radiology, August 1, 2003; 228(2): 583 - 588. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hong, K. T. Bae, T. K. Pilgram, and F. Zhu Coronary Artery Calcium Quantification at Multi-Detector Row CT: Influence of Heart Rate and Measurement Methods on Interacquisition Variability—Initial Experience Radiology, July 1, 2003; 228(1): 95 - 100. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Saito, M. Saito, S. Komatu, and K. Ohtomo Real-Time Four-dimensional Imaging of the Heart with Multi-Detector Row CT RadioGraphics, January 1, 2003; 23(1): e8 - e8. [Abstract] [Full Text] |
||||
![]() |
K Nieman, B J Rensing, R-J M van Geuns, J Vos, P M T Pattynama, G P Krestin, P W Serruys, and P J de Feyter Non-invasive coronary angiography with multislice spiral computed tomography: impact of heart rate Heart, December 1, 2002; 88(5): 470 - 474. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. A. Fayad, V. Fuster, K. Nikolaou, and C. Becker Computed Tomography and Magnetic Resonance Imaging for Noninvasive Coronary Angiography and Plaque Imaging: Current and Potential Future Concepts Circulation, October 8, 2002; 106(15): 2026 - 2034. [Full Text] [PDF] |
||||
![]() |
T. Giesler, U. Baum, D. Ropers, S. Ulzheimer, E. Wenkel, M. Mennicke, W. Bautz, W. A. Kalender, W. G. Daniel, and S. Achenbach Noninvasive Visualization of Coronary Arteries Using Contrast-Enhanced Multidetector CT: Influence of Heart Rate on Image Quality and Stenosis Detection Am. J. Roentgenol., October 1, 2002; 179(4): 911 - 916. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. K. Willmann, D. Weishaupt, M. Lachat, R. Kobza, J. E. Roos, B. Seifert, T. F. Luscher, B. Marincek, and P. R. Hilfiker Electrocardiographically Gated Multi-Detector Row CT for Assessment of Valvular Morphology and Calcification in Aortic Stenosis Radiology, October 1, 2002; 225(1): 120 - 128. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Lu, N. Zhuang, S.-S. Mao, J. Child, S. Carson, H. Bakhsheshi, and M. J. Budoff EKG-triggered CT Data Acquisition to Reduce Variability in Coronary Arterial Calcium Score Radiology, September 1, 2002; 224(3): 838 - 844. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |