Breast Imaging |
1 From the Division of Imaging Research, Department of Radiology (B.Z., D.G.), the Departments of Radiology (C.A.B., M.A.G., C.M.H., L.A.H., T.S.C.) and Biostatistics (H.E.R.), University of Pittsburgh, 300 Halket St, Suite 4200, Pittsburgh, PA 15213; and the Magee Womens Hospital, University of Pittsburgh Medical Center Health System, Pa (M.A.G., C.M.H., L.A.H.). Received January 12, 2001; revision requested March 5; revision received March 29; accepted May 1. Supported in part by the U.S. Army Medical Research Acquisition Activity under contracts DAMD17-98-1-8018 and DAMD17-00-1-0410 and by grant CA77850 from the National Cancer Institute, National Institutes of Health. Address correspondence to B.Z. (e-mail: bzheng@radserv.arad.upmc.edu).
PURPOSE: To assess the performance of radiologists in the detection of masses and microcalcification clusters on digitized mammograms by using different computer-assisted detection (CAD) cuing environments.
MATERIALS AND METHODS: Two hundred nine digitized mammograms depicting 57 verified masses and 38 microcalcification clusters in 85 positive and 35 negative cases were interpreted independently by seven radiologists using five display modes. Except for the first mode, for which no CAD results were provided, suspicious regions identified with a CAD scheme were cued in all the other modes by using a combination of two cuing sensitivities (90% and 50%) and two false-positive rates (0.5 and 2.0 per image). A receiver operating characteristic study was performed by using soft-copy images.
RESULTS: CAD cuing at 90% sensitivity and a rate of 0.5 false-positive region per image improved observer performance levels significantly (P < .01). As accuracy of CAD cuing decreased so did observer performances (P < .01). Cuing specificity affected mass detection more significantly, while cuing sensitivity affected detection of microcalcification clusters more significantly (P < .01). Reduction of cuing sensitivity and specificity significantly increased false-negative rates in noncued areas (P < .05). Trends were consistent for all observers.
CONCLUSION: CAD systems have the potential to significantly improve diagnostic performance in mammography. However, poorly performing schemes could adversely affect observer performance in both cued and noncued areas.
Index terms: Breast neoplasms, diagnosis, 00.30, 00.81 Cancer screening, 00.11 Computers, diagnostic aid Diagnostic radiology, observer performance
This article has been cited by other articles:
![]() |
S. K. Yang, W. K. Moon, N. Cho, J. S. Park, J. H. Cha, S. M. Kim, S. J. Kim, and J.-G. Im Screening Mammography-detected Cancers: Sensitivity of a Computer-aided Detection System Applied to Full-Field Digital Mammograms Radiology, July 1, 2007; 244(1): 104 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Taplin, C. M. Rutter, and C. D. Lehman Testing the effect of computer-assisted detection on interpretive performance in screening mammography. Am. J. Roentgenol., December 1, 2006; 187(6): 1475 - 1482. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. R. Pai, N. E. Gregory, A. E. Swinford, and M. Rebner Ductal Carcinoma in Situ: Computer-aided Detection in Screening Mammography Radiology, December 1, 2006; 241(3): 689 - 694. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Zheng, G. S. Maitz, M. A. Ganott, G. Abrams, J. K. Leader, and D. Gur Performance and Reproducibility of a Computerized Mass Detection Scheme for Digitized Mammography Using Rotated and Resampled Images: An Assessment Am. J. Roentgenol., July 1, 2005; 185(1): 194 - 198. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. T. Bae, J.-S. Kim, Y.-H. Na, K. G. Kim, and J.-H. Kim Pulmonary Nodules: Automated Detection on CT Images with Morphologic Matching Algorithm--Preliminary Results Radiology, July 1, 2005; 236(1): 286 - 293. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. F. Brem, J. W. Hoffmeister, G. Zisman, M. P. DeSimio, and S. K. Rogers A Computer-Aided Detection System for the Evaluation of Breast Cancer by Mammographic Appearance and Lesion Size Am. J. Roentgenol., March 1, 2005; 184(3): 893 - 896. [Abstract] [Full Text] [PDF] |
||||
![]() |
S M Astley Evaluation of computer-aided detection (CAD) prompting techniques for mammography Br. J. Radiol., January 1, 2005; 78(suppl_1): S20 - S25. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. D. Rubin, J. K. Lyo, D. S. Paik, A. J. Sherbondy, L. C. Chow, A. N. Leung, R. Mindelzun, P. K. Schraedley-Desmond, S. E. Zinck, D. P. Naidich, et al. Pulmonary Nodules on Multi-Detector Row CT Scans: Performance Comparison of Radiologists and Computer-aided Detection Radiology, January 1, 2005; 234(1): 274 - 283. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Gur Imaging Technology and Practice Assessments: Diagnostic Performance, Clinical Relevance, and Generalizability in a Changing Environment Radiology, November 1, 2004; 233(2): 309 - 312. [Full Text] [PDF] |
||||
![]() |
D. Gur, J. S. Stalder, L. A. Hardesty, B. Zheng, J. H. Sumkin, D. M. Chough, B. E. Shindel, and H. E. Rockette Computer-aided Detection Performance in Mammographic Examination of Masses: Assessment Radiology, November 1, 2004; 233(2): 418 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. V. Destounis, P. DiNitto, W. Logan-Young, E. Bonaccio, M. L. Zuley, and K. M. Willison Can Computer-aided Detection with Double Reading of Screening Mammograms Help Decrease the False-Negative Rate? Initial Experience Radiology, August 1, 2004; 232(2): 578 - 584. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Zheng, J. K. Leader, G. Abrams, B. Shindel, V. Catullo, W. F. Good, and D. Gur Computer-Aided Detection Schemes: The Effect of Limiting the Number of Cued Regions in Each Case Am. J. Roentgenol., March 1, 2004; 182(3): 579 - 583. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Zheng, L. A. Hardesty, W. R. Poller, J. H. Sumkin, and S. Golla Mammography with Computer-aided Detection: Reproducibility Assessment—Initial Experience Radiology, July 1, 2003; 228(1): 58 - 62. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Smith, D. Saslow, K. Andrews Sawyer, W. Burke, M. E. Costanza, W. P. Evans III, R. S. Foster Jr., E. Hendrick, H. J. Eyre, and S. Sener American Cancer Society Guidelines for Breast Cancer Screening: Update 2003 CA Cancer J Clin, May 1, 2003; 53(3): 141 - 169. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Harvey, B. Geller, R. G. Oppenheimer, M. Pinet, L. Riddell, and B. Garra Increase in Cancer Detection and Recall Rates with Independent Double Interpretation of Screening Mammography Am. J. Roentgenol., May 1, 2003; 180(5): 1461 - 1467. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. D'Orsi Computer-aided Detection: There Is No Free Lunch Radiology, December 1, 2001; 221(3): 585 - 586. [Full Text] [PDF] |
||||