Cardiac Imaging |
1 From the Department of Clinical Radiology, Klinikum Grosshadern, University of Munich, Marchioninistrasse 15, 81377 Munich, Germany (C.H., C.R.B., R.B., M.F.R.); Department of Radiology, Huazhong University of Science and Technology, Wuhan, Peoples Republic of China (C.H.); and Siemens Medical Systems, Forchheim, Germany (B.O.). Received May 14, 2001; revision requested June 13; revision received August 27; accepted September 20. Address correspondence to C.R.B. (e-mail: christoph.becker@ikra.med.uni-muenchen.de).
PURPOSE: (a) To determine the accuracy of multidetector row computed tomography (CT) in the measurement of the calcium concentration in a cardiac CT calibration phantom and (b) to assess the correlation of a traditional 3-mm section width CT coronary screening protocol and a 1.25-mm section width CT angiography imaging protocol in the quantification of the absolute mass of coronary calcium in patients who underwent both coronary screening and CT angiography with a multidetector row CT scanner.
MATERIALS AND METHODS: A heart phantom containing calcified cylinders was scanned to determine calibration factors and absolute calcium mass. In 50 patients, the variability (value 1 - value 2/mean value 1 - value 2), limit of agreement (±2SD value 1 - value 2), and systematic error (mean value 1 - value 2) of the total amount of coronary calcium calculated at traditional 3-mm section width CT and at 1.25-mm section width CT angiography were determined.
RESULTS: The correlation coefficient between the 3-mm section width, nonenhanced protocol and the 1.25-mm section width CT angiography protocol was very high (r = 0.977) and the mean variability was low (19.7%) for the absolute mass. There was a systematic error of -6.7 mg and a limit of agreement between 45.0 mg and -58.5 mg.
CONCLUSION: Use of the mass quantification algorithm in combination with a calibration phantom allows accurate quantification of coronary calcium. Measurements of calcium mass obtained at 1.25-mm section width CT angiography have the best agreement with those obtained at the traditional 3-mm section width imaging protocol.
© RSNA, 2002
Index terms: Coronary vessels, calcification, 54.812 Coronary vessels, CT, 54.12111, 54.12116 Phantoms
This article has been cited by other articles:
![]() |
R. Raman, B. Raman, S. Napel, and G. D. Rubin Semiautomated Quantification of the Mass and Distribution of Vascular Calcification with Multidetector CT: Method and Evaluation Radiology, April 1, 2008; 247(1): 241 - 250. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. R. Nandalur, A. D. Hardie, P. Raghavan, M. J. Schipper, E. Baskurt, and C. M. Kramer Composition of the Stable Carotid Plaque: Insights From a Multidetector Computed Tomography Study of Plaque Volume Stroke, March 1, 2007; 38(3): 935 - 940. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, R. S. Blumenthal, J. J. Carr, J. G. Goldin, P. Greenland, A. D. Guerci, J. A.C. Lima, D. J. Rader, G. D. Rubin, et al. Assessment of Coronary Artery Disease by Cardiac Computed Tomography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology Circulation, October 17, 2006; 114(16): 1761 - 1791. [Full Text] [PDF] |
||||
![]() |
E S Houslay, S J Cowell, R J Prescott, J Reid, J Burton, D B Northridge, N A Boon, D E Newby, and on behalf of the Scottish Aortic Stenosis and Lipi Progressive coronary calcification despite intensive lipid-lowering treatment: a randomised controlled trial Heart, September 1, 2006; 92(9): 1207 - 1212. [Abstract] [Full Text] [PDF] |
||||
![]() |
F Cademartiri Is calcium the key for the assessment of progression/regression of coronary artery disease? Heart, September 1, 2006; 92(9): 1187 - 1188. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. R. Nandalur, E. Baskurt, K. D. Hagspiel, M. Finch, C. D. Phillips, S. R. Bollampally, and C. M. Kramer Carotid Artery Calcification on CT May Independently Predict Stroke Risk Am. J. Roentgenol., February 1, 2006; 186(2): 547 - 552. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Fuster, Z. A. Fayad, P. R. Moreno, M. Poon, R. Corti, and J. J. Badimon Atherothrombosis and High-Risk Plaque: Part II: Approaches by Noninvasive Computed Tomographic/Magnetic Resonance Imaging J. Am. Coll. Cardiol., October 4, 2005; 46(7): 1209 - 1218. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Halliburton, A. E. Stillman, M. Lieber, J. M. Kasper, S. A. Kuzmiak, and R. D. White Potential Clinical Impact of Variability in the Measurement of Coronary Artery Calcification with Sequential MDCT Am. J. Roentgenol., February 1, 2005; 184(2): 643 - 648. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yamamuro, E. Tadamura, S. Kubo, H. Toyoda, T. Nishina, M. Ohba, R. Hosokawa, T. Kimura, N. Tamaki, M. Komeda, et al. Cardiac Functional Analysis with Multi-Detector Row CT and Segmental Reconstruction Algorithm: Comparison with Echocardiography, SPECT, and MR Imaging Radiology, February 1, 2005; 234(2): 381 - 390. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. R. Nandalur, E. Baskurt, K. D. Hagspiel, C. D. Phillips, and C. M. Kramer Calcified Carotid Atherosclerotic Plaque Is Associated Less with Ischemic Symptoms Than Is Noncalcified Plaque on MDCT Am. J. Roentgenol., January 1, 2005; 184(1): 295 - 298. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Koos, A. H. Mahnken, A. M. Sinha, J. E. Wildberger, R. Hoffmann, and H. P. Kuhl Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT Am. J. Roentgenol., December 1, 2004; 183(6): 1813 - 1818. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hong, T. K. Pilgram, F. Zhu, and K. T. Bae Coronary Artery Calcification: Effect of Size of Field of View on Multi-Detector Row CT Measurements Radiology, October 1, 2004; 233(1): 281 - 285. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. J. Schoepf, C. R. Becker, B. M. Ohnesorge, and E. K. Yucel CT of Coronary Artery Disease Radiology, July 1, 2004; 232(1): 18 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Hokanson, T. MacKenzie, G. Kinney, J. K. Snell-Bergeon, D. Dabelea, J. Ehrlich, R. H. Eckel, and M. Rewers Evaluating Changes in Coronary Artery Calcium: An Analytic Method That Accounts for Interscan Variability Am. J. Roentgenol., May 1, 2004; 182(5): 1327 - 1332. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Vliegenthart, B. Song, A. Hofman, J. C. M. Witteman, and M. Oudkerk Coronary Calcification at Electron-Beam CT: Effect of Section Thickness on Calcium Scoring in Vitro and in Vivo Radiology, November 1, 2003; 229(2): 520 - 525. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Hoffmann, D. C. Kwait, J. Handwerker, R. Chan, G. Lamuraglia, and T. J. Brady Vascular Calcification in ex Vivo Carotid Specimens: Precision and Accuracy of Measurements with Multi-Detector Row CT Radiology, November 1, 2003; 229(2): 375 - 381. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. R. Van Hoe, K. G. De Meerleer, P. Ph. Leyman, and P. K. Vanhoenacker Coronary Artery Calcium Scoring Using ECG-Gated Multidetector CT: Effect of Individually Optimized Image-Reconstruction Windows on Image Quality and Measurement Reproducibility Am. J. Roentgenol., October 1, 2003; 181(4): 1093 - 1100. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Rumberger and L. Kaufman A Rosetta Stone for Coronary Calcium Risk Stratification: Agatston, Volume, and Mass Scores in 11,490 Individuals Am. J. Roentgenol., September 1, 2003; 181(3): 743 - 748. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hong, K. T. Bae, T. K. Pilgram, and F. Zhu Coronary Artery Calcium Quantification at Multi-Detector Row CT: Influence of Heart Rate and Measurement Methods on Interacquisition Variability—Initial Experience Radiology, July 1, 2003; 228(1): 95 - 100. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. B. Mark, L. J. Shaw, M. S. Lauer, P. G. O'Malley, and P. Heidenreich Task force #5--is atherosclerosis imaging cost effective? J. Am. Coll. Cardiol., June 4, 2003; 41(11): 1906 - 1917. [Full Text] [PDF] |
||||
![]() |
C. Hong, K. T. Bae, and T. K. Pilgram Coronary Artery Calcium: Accuracy and Reproducibility of Measurements with Multi-Detector Row CT—Assessment of Effects of Different Thresholds and Quantification Methods Radiology, June 1, 2003; 227(3): 795 - 801. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hong, K. T. Bae, T. K. Pilgram, J. Suh, and D. Bradley Coronary Artery Calcium Measurement with Multi-Detector Row CT: In Vitro Assessment of Effect of Radiation Dose Radiology, December 1, 2002; 225(3): 901 - 906. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |