1 From the Departments of Radiology, University of Pittsburgh and Magee-Womens Hospital, Imaging Research, Suite 4200, 300 Halket St, Pittsburgh, PA 15213. Received April 29, 2002; revision requested June 21; final revision received September 23; accepted October 23. Supported in part by grants CA77850, CA85241, and CA80836 from the National Cancer Institute of the National Institutes of Health and by the U.S. Army Medical Research Acquisition Center, Fort Detrick, Md, under contract DAMD17-00-1-0410. Address correspondence to B.Z. (e-mail: zhengb@msx.upmc.edu).
PURPOSE: To examine the performance and reproducibility of a commercially available computer-aided detection (CAD) system with a set of mammograms obtained in 100 patients who had undergone biopsy after positive findings at mammography.
MATERIALS AND METHODS: One hundred positive mammographic examinations (four views each), depicting 96 masses and 50 microcalcification clusters, were scanned and analyzed three times by the CAD system. Reproducibility of detection sensitivity and the individual CAD-generated cues in the three images were examined. Both abnormality- and region-based detection sensitivities were compared.
RESULTS: Forty-eight (96.0%) of 50 microcalcification clusters were marked on all three images in the abnormality-based analysis. Of the remaining two clusters, one was marked in two images and one was marked in only one. The abnormality-based sensitivity for mass detection ranged from 66.7% (64 of 96) to 70.8% (68 of 96). The system generated identical patterns (including images with and those without cues) for all three images in 53.3% (213 of 400) of images. For true-positive cluster regions, 88.9% (80 of 90) were marked at the same location in all images. For true-positive mass regions, 69.5% (82 of 118) were marked at the same locations in all images. In false-positive detections, only 44.0% (81 of 184) of false-positive mass regions and 31.9% (38 of 119) of false-positive cluster regions were marked at the same locations on all three images.
CONCLUSION: Reproducibility of marked regions generated by the CAD system is improved from that reported previously, largely as a result of the substantial reduction in the false-positive detection rates. Reproducibility of true-positive identification of masses remains an important issue that may have methodologic and clinical practice implications.
Index terms: Breast neoplasms, diagnosis, 00.119 Breast radiography, technology, 00.119 Computers, diagnostic aid, 00.119
This article has been cited by other articles:
![]() |
S. J. Kim, W. K. Moon, N. Cho, J. H. Cha, S. M. Kim, and J.-G. Im Computer-aided Detection in Full-Field Digital Mammography: Sensitivity and Reproducibility in Serial Examinations Radiology, December 1, 2007; 246(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Kim, W. K. Moon, N. Cho, J. H. Cha, S. M. Kim, and J.-G. Im Computer-aided Detection in Digital Mammography: Comparison of Craniocaudal, Mediolateral Oblique, and Mediolateral Views Radiology, December 1, 2006; 241(3): 695 - 701. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Ko, M. J. Nicholas, J. B. Mendel, and P. J. Slanetz Prospective assessment of computer-aided detection in interpretation of screening mammography. Am. J. Roentgenol., December 1, 2006; 187(6): 1483 - 1491. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. R. Pai, N. E. Gregory, A. E. Swinford, and M. Rebner Ductal Carcinoma in Situ: Computer-aided Detection in Screening Mammography Radiology, December 1, 2006; 241(3): 689 - 694. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. V. Ongeval, H. Bosmans, and A. Van Steen Current challenges of full field digital mammography Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 148 - 153. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Zheng, G. S. Maitz, M. A. Ganott, G. Abrams, J. K. Leader, and D. Gur Performance and Reproducibility of a Computerized Mass Detection Scheme for Digitized Mammography Using Rotated and Resampled Images: An Assessment Am. J. Roentgenol., July 1, 2005; 185(1): 194 - 198. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Gur, J. S. Stalder, L. A. Hardesty, B. Zheng, J. H. Sumkin, D. M. Chough, B. E. Shindel, and H. E. Rockette Computer-aided Detection Performance in Mammographic Examination of Masses: Assessment Radiology, November 1, 2004; 233(2): 418 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Baker, J. Y. Lo, D. M. Delong, and C. E. Floyd Computer-aided Detection in Screening Mammography: Variability in Cues Radiology, November 1, 2004; 233(2): 411 - 417. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Guenin and B. Zheng Long-term Retention of Mammographic Computer-assisted Diagnosis Information Is Neither Necessary Nor Desirable [letter] * Dr Zheng responds: Radiology, February 1, 2004; 230(2): 595 - 597. [Full Text] [PDF] |
||||