Experimental Studies |
1 From the Departments of Surgery A (F.B., A.G., A.N., R.S., R.L.) and Pathology (T.C.), Hospital Clínico Universitario Lozano Blesa, Zaragoza, Spain; and Departments of Electric Engineering and Communications (J.M.B.) and Animal Pathology and Surgery, Veterinary Faculty (I.C., O.B.), University of Zaragoza, Spain. Received August 5, 2002; revision requested September 24; final revision received February 26, 2003; accepted March 11. Supported by Spanish Government grant FIS 00/255 for medical research. Address correspondence to F.B., Tapis 155, 17600 Figueres (Girona), Spain (e-mail: fburdio@comll.es).
PURPOSE: To evaluate whether a bipolar saline-enhanced radiofrequency (RF) ablation system embedded in one needle is able to consistently produce homogeneous and predictable areas of coagulation necrosis with or without the Pringle maneuver of vascular inflow occlusion.
MATERIALS AND METHODS: RF ablation (480 kHz) of the liver was performed in 24 healthy pigs by means of laparotomy: group A (n = 5), 4-cm distance between electrodes 1 and 2; group B (n = 7), 4-cm distance and the Pringle maneuver; group C (n = 5), 2-cm distance; and group D (n = 7), 2-cm distance with the Pringle maneuver. Twenty percent NaCl solution was infused continuously at a rate of 100 mL/h via each electrode during the procedure. The pigs were followed up, and they were euthanized on the 7th day. Livers were removed for histologic assessment. Time, impedance, current, power output, specific voltage of the contacts, energy output, temperatures in the liver, volume of the lesion, and energy delivered per lesion volume were determined and compared among groups. Predictability of lesion volume was evaluated with the coefficient of variability. Mean values of the variables were compared among the groups by means of one-way analysis of variance or Kruskall-Wallis test.
RESULTS: Impedance at the end of the RF ablation procedure was almost twofold lower than the corresponding initial value in all groups. In Pringle groups B and D, regular ellipsoids of coagulation necrosis were created (mean lesion volume, 149.50 cm3 ± 34.26 and 69.43 cm3 ± 15.48, respectively). In non-Pringle groups A and C, the shape of coagulation necrosis was influenced by the vessels encountered, and mean lesion size was lower than that in the Pringle groups (P < .01). The coefficient of variability of lesion size was lower in the Pringle groups (23% and 22%, respectively) than that in the non-Pringle groups (75% and 30%, respectively).
CONCLUSION: The bipolar saline-enhanced RF ablation method produces homogeneous and predictable areas of coagulation necrosis between two electrodes, regardless of the distance between them, preferably with vascular inflow occlusion.
© RSNA, 2003
Index terms: Animals Experimental study Liver, interventional procedures, 761.1269 Liver neoplasms, 761.1269 Radiofrequency (RF) ablation, 761.1269
This article has been cited by other articles:
![]() |
S. Mulier, Y. Ni, L. Frich, F. Burdio, A. L. Denys, J.-F. De Wispelaere, B. Dupas, N. Habib, M. Hoey, M. C. Jansen, et al. Experimental and Clinical Radiofrequency Ablation: Proposal for Standardized Description of Coagulation Size and Geometry Ann. Surg. Oncol., April 1, 2007; 14(4): 1381 - 1396. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Clasen, D. Schmidt, A. Boss, K. Dietz, S. M. Krober, C. D. Claussen, and P. L. Pereira Multipolar Radiofrequency Ablation with Internally Cooled Electrodes: Experimental Study in ex Vivo Bovine Liver with Mathematic Modeling Radiology, March 1, 2006; 238(3): 881 - 890. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Lee, J. K. Han, S. H. Kim, S. H. Choi, S. K. An, C. J. Han, and B. I. Choi Bipolar Radiofrequency Ablation Using Wet-Cooled Electrodes: An In Vitro Experimental Study in Bovine Liver Am. J. Roentgenol., February 1, 2005; 184(2): 391 - 397. [Abstract] [Full Text] [PDF] |
||||
![]() |
J M Lee, J K Han, S H Kim, J Y Lee, S H Choi, and B I Choi Hepatic bipolar radiofrequency ablation using perfused-cooled electrodes: a comparative study in the ex vivo bovine liver Br. J. Radiol., November 1, 2004; 77(923): 944 - 949. [Abstract] [Full Text] [PDF] |
||||