Breast Imaging |
1 From the Department of Radiology, Breast Imaging Center, Ullevaal University Hospital, Kirkeveien 166, N-0407 Oslo, Norway. From the 2001 RSNA scientific assembly. Received September 12, 2002; revision requested November 22; final revision received June 6, 2003; accepted July 15. Address correspondence to P.S. (e-mail: per.skaane@ulleval.no).
PURPOSE: To compare screen-film and full-field digital mammography with soft-copy reading in a population-based screening program.
MATERIALS AND METHODS: Full-field digital and screen-film mammography were performed in 3,683 women aged 5069 years. Two standard views of each breast were acquired with each modality. Images underwent independent double reading with use of a five-point rating scale for probability of cancer. Recall rates and positive predictive values were calculated. Cancer detection rates determined with both modalities were compared by using the McNemar test for paired proportions. Retrospective side-by-side analysis for conspicuity of cancers was performed by an external independent radiologist group with experience in both modalities.
RESULTS: In 3,683 cases, 31 cancers were detected. Screen-film mammography depicted 28 (0.76%) malignancies, and full-field digital mammography depicted 23 (0.62%) malignancies. The difference between cancer detection rates was not significant (P = .23). The recall rate for full-field digital mammography (4.6%; 168 of 3,683 cases) was slightly higher than that for screen-film mammography (3.5%; 128 of 3,683 cases). The positive predictive value based on needle biopsy results was 46% for screen-film mammography and 39% for full-field digital mammography. Side-by-side image comparison for cancer conspicuity led to classification of 19 cancers as equal for probability of malignancy, six cancers as slightly better demonstrated at screen-film mammography, and six cancers as slightly better demonstrated at full-field digital mammography.
CONCLUSION: There was no statistically significant difference in cancer detection rate between screen-film and full-field digital mammography. Cancer conspicuity was equal with both modalities. Full-field digital mammography with soft-copy reading is comparable to screen-film mammography in population-based screening.
© RSNA, 2003
Index terms: Breast neoplasms, radiography, 00.112, 00.1215 Breast radiography, comparative studies, 00.112, 00.1215 Cancer screening Radiography, digital, 00.1215
This article has been cited by other articles:
![]() |
M. Sala, M. Comas, F. Macia, J. Martinez, M. Casamitjana, and X. Castells Implementation of Digital Mammography in a Population-based Breast Cancer Screening Program: Effect of Screening Round on Recall Rate and Cancer Detection Radiology, July 1, 2009; 252(1): 31 - 39. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Vinnicombe, S. M. Pinto Pereira, V. A. McCormack, S. Shiel, N. Perry, and I. M. dos Santos Silva Full-Field Digital versus Screen-Film Mammography: Comparison within the UK Breast Screening Program and Systematic Review of Published Data Radiology, May 1, 2009; 251(2): 347 - 358. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Haygood, J. Wang, E. N. Atkinson, D. Lane, T. W. Stephens, P. Patel, and G. J. Whitman Timed Efficiency of Interpretation of Digital and Film-Screen Screening Mammograms Am. J. Roentgenol., January 1, 2009; 192(1): 216 - 220. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. B. Hruska, S. W. Phillips, D. H. Whaley, D. J. Rhodes, and M. K. O'Connor Molecular Breast Imaging: Use of a Dual-Head Dedicated Gamma Camera to Detect Small Breast Tumors Am. J. Roentgenol., December 1, 2008; 191(6): 1805 - 1815. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Van Ongeval, A. Van Steen, and H. Bosmans Teaching syllabus for radiological aspects of breast cancer screening with digital mammography Radiat Prot Dosimetry, March 1, 2008; 129(1-3): 191 - 194. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N.A. Tosteson, N. K. Stout, D. G. Fryback, S. Acharyya, B. A. Herman, L. G. Hannah, E. D. Pisano, and for the DMIST Investigators Cost-Effectiveness of Digital Mammography Breast Cancer Screening Ann Intern Med, January 1, 2008; 148(1): 1 - 10. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. D. Pisano, R. E. Hendrick, M. Yaffe, E. F. Conant, and C. Gatsonis Should Breast Imaging Practices Convert to Digital Mammography? A Response from Members of the DMIST Executive Committee Radiology, October 1, 2007; 245(1): 12 - 13. [Full Text] [PDF] |
||||
![]() |
M. R. Del Turco, P. Mantellini, S. Ciatto, R. Bonardi, F. Martinelli, B. Lazzari, and N. Houssami Full-Field Digital Versus Screen-Film Mammography: Comparative Accuracy in Concurrent Screening Cohorts Am. J. Roentgenol., October 1, 2007; 189(4): 860 - 866. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Skaane, S. Hofvind, and A. Skjennald Randomized Trial of Screen-Film versus Full-Field Digital Mammography with Soft-Copy Reading in Population-based Screening Program: Follow-up and Final Results of Oslo II Study Radiology, September 1, 2007; 244(3): 708 - 717. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Yang, W. K. Moon, N. Cho, J. S. Park, J. H. Cha, S. M. Kim, S. J. Kim, and J.-G. Im Screening Mammography-detected Cancers: Sensitivity of a Computer-aided Detection System Applied to Full-Field Digital Mammograms Radiology, July 1, 2007; 244(1): 104 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Jiang, D. L. Miglioretti, C. E. Metz, and R. A. Schmidt Breast Cancer Detection Rate: Designing Imaging Trials to Demonstrate Improvements Radiology, May 1, 2007; 243(2): 360 - 367. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Skaane, A. Kshirsagar, S. Stapleton, K. Young, and R. A. Castellino Effect of Computer-Aided Detection on Independent Double Reading of Paired Screen-Film and Full-Field Digital Screening Mammograms Am. J. Roentgenol., February 1, 2007; 188(2): 377 - 384. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. A. J. Roelofs, N. Karssemeijer, N. Wedekind, C. Beck, S. van Woudenberg, P. R. Snoeren, J. H. C. L. Hendriks, M. Rosselli del Turco, N. Bjurstam, H. Junkermann, et al. Importance of Comparison of Current and Prior Mammograms in Breast Cancer Screening Radiology, January 1, 2007; 242(1): 70 - 77. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. A. Berns, R. E. Hendrick, M. Solari, L. Barke, D. Reddy, J. Wolfman, L. Segal, P. DeLeon, S. Benjamin, and L. Willis Digital and screen-film mammography: comparison of image acquisition and interpretation times. Am. J. Roentgenol., July 1, 2006; 187(1): 38 - 41. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Crystal, S. Strano, J. D. Keen, M. H. Ebell, C. Gatsonis, E. D. Pisano, and E. Hendrick Digital and film mammography. N. Engl. J. Med., February 16, 2006; 354(7): 765 - 767. [Full Text] [PDF] |
||||
![]() |
H. Bosmans, A.-K. Carton, F. Rogge, F. Zanca, J. Jacobs, C. Van Ongeval, K. Nijs, A. Van Steen, and G. Marchal Image quality measurements and metrics in full field digital mammography: an overview Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 120 - 130. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. V. Ongeval, H. Bosmans, and A. Van Steen Current challenges of full field digital mammography Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 148 - 153. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. D. Dershaw Film or digital mammographic screening? N. Engl. J. Med., October 27, 2005; 353(17): 1846 - 1847. [Full Text] [PDF] |
||||
![]() |
E. D. Pisano, C. Gatsonis, E. Hendrick, M. Yaffe, J. K. Baum, S. Acharyya, E. F. Conant, L. L. Fajardo, L. Bassett, C. D'Orsi, et al. Diagnostic Performance of Digital versus Film Mammography for Breast-Cancer Screening N. Engl. J. Med., October 27, 2005; 353(17): 1773 - 1783. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. K. Nishino, X. Wu, and R. F. Johnson Jr. Thickness of Molybdenum Filter and Squared Contrast-to-Noise Ratio per Dose for Digital Mammography Am. J. Roentgenol., October 1, 2005; 185(4): 960 - 963. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Skaane, C. Balleyguier, F. Diekmann, S. Diekmann, J.-C. Piguet, K. Young, and L. T. Niklason Breast Lesion Detection and Classification: Comparison of Screen-Film Mammography and Full-Field Digital Mammography with Soft-copy Reading--Observer Performance Study Radiology, October 1, 2005; 237(1): 37 - 44. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. D. Pisano, C. A. Gatsonis, M. J. Yaffe, R. E. Hendrick, A. N. A. Tosteson, D. G. Fryback, L. W. Bassett, J. K. Baum, E. F. Conant, R. A. Jong, et al. American College of Radiology Imaging Network Digital Mammographic Imaging Screening Trial: Objectives and Methodology Radiology, August 1, 2005; 236(2): 404 - 412. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Elmore, K. Armstrong, C. D. Lehman, and S. W. Fletcher Screening for Breast Cancer JAMA, March 9, 2005; 293(10): 1245 - 1256. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Gur Technology and Practice Assessment: In Search of a "Desirable" Statement Radiology, March 1, 2005; 234(3): 659 - 660. [Full Text] [PDF] |
||||
![]() |
E. D. Pisano and M. J. Yaffe Digital Mammography Radiology, February 1, 2005; 234(2): 353 - 362. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Taplin, L. Ichikawa, M. U. Yood, M. M. Manos, A. M. Geiger, S. Weinmann, J. Gilbert, J. Mouchawar, W. A. Leyden, R. Altaras, et al. Reason for Late-Stage Breast Cancer: Absence of Screening or Detection, or Breakdown in Follow-up? J Natl Cancer Inst, October 20, 2004; 96(20): 1518 - 1527. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Skaane and A. Skjennald Screen-Film Mammography versus Full-Field Digital Mammography with Soft-Copy Reading: Randomized Trial in a Population-based Screening Program--The Oslo II Study Radiology, July 1, 2004; 232(1): 197 - 204. [Abstract] [Full Text] [PDF] |
||||