1 From the Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, Ill. Received October 10, 2002; revision requested December 19; final revision received June 5, 2003; accepted June 13. Address correspondence to D.S.F., Cedars-Sinai Medical Center, S. Mark Taper Imaging Center, Room 1258, 8700 Beverly Blvd, Los Angeles, CA 90048 (e-mail: david.fieno@cshs.org).
PURPOSE: To assess the accuracy of cine magnetic resonance (MR) imaging with a segmented true fast imaging with steady-state precession (FISP) technique for right ventricular (RV) mass quantification.
MATERIALS AND METHODS: Fourteen dogs were imaged with a 1.5-T clinical MR imaging unit by using an electrocardiographically gated true FISP sequence. Contiguous segmented k-space cine images were acquired from the base of the RV to the apex during suspended respiration (repetition time msec/echo time msec, 3.2/1.6; section thickness, 5 mm; in-plane resolution, 1.0 x 1.3 mm2). After imaging, each dog was sacrificed, and the RV free wall was isolated and weighed. Each MR imaging data set was analyzed twice by each of two independent observers who were blinded to the results of RV mass measurement at autopsy, and the mass measurements at MR imaging were compared with the autopsy results by using linear regression and Bland-Altman analysis.
RESULTS: RV mass measurements calculated by using the true FISP cine MR images were nearly identical to those at autopsy (R = 0.82, standard error of the estimate = 1.7 g, P > .05), with a mean difference between the autopsy and MR imaging measurements of 0.3 g ± 1.7 (1.9% ± 8.2) (P > .05). Inter- and intraobserver variations were small, with a mean interobserver variability of -0.1 g ± 2.3 and a mean intraobserver variability of 0.2 g ± 1.6 at every-section analysis.
CONCLUSION: In this animal model, true FISP cine MR imaging enabled accurate quantification of RV mass.
© RSNA, 2003
Index terms: Animals Heart, MR, 523.121416 Heart, ventricles, 523.121416 Magnetic resonance (MR), experimental studies Magnetic resonance (MR), pulse sequences
This article has been cited by other articles:
![]() |
L. Williams and M. Frenneaux Assessment of right ventricular function Heart, April 1, 2008; 94(4): 404 - 405. [Full Text] [PDF] |
||||
![]() |
M. Carlsson, M. Ugander, E. Heiberg, and H. Arheden The quantitative relationship between longitudinal and radial function in left, right, and total heart pumping in humans Am J Physiol Heart Circ Physiol, July 1, 2007; 293(1): H636 - H644. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Finn, K. Nael, V. Deshpande, O. Ratib, and G. Laub Cardiac MR Imaging: State of the Technology. Radiology, November 1, 2006; 241(2): 338 - 354. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. I. Norton, C. Tong, R. B. J. Glass, and J. C. Nielsen Cardiac MR Imaging Assessment Following Tetralogy of Fallot Repair RadioGraphics, January 1, 2006; 26(1): 197 - 211. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ector, S. De Buck, J. Adams, S. Dymarkowski, J. Bogaert, F. Maes, and H. Heidbuchel Cardiac Three-Dimensional Magnetic Resonance Imaging and Fluoroscopy Merging: A New Approach for Electroanatomic Mapping to Assist Catheter Ablation Circulation, December 13, 2005; 112(24): 3769 - 3776. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Berg, V. L. Sorrell, K. B. Kern, R. W. Hilwig, M. I. Altbach, M. M. Hayes, K. A. Bates, and G. A. Ewy Magnetic Resonance Imaging During Untreated Ventricular Fibrillation Reveals Prompt Right Ventricular Overdistention Without Left Ventricular Volume Loss Circulation, March 8, 2005; 111(9): 1136 - 1140. [Abstract] [Full Text] [PDF] |
||||