1 From the Department of Radiology, CT Unit, Hôpital Marie-Lannelongue, 133 Avenue de la Résistance, 92350 Le Plessis-Robinson, France (A.B.S.C., R.H., J.F.P.); Department of Interventional Radiology, Hôpital René Dubos, Cergy-Pontoise, France (H.T.A.); and Department of CT Physics and Applications, Siemens Medical Solutions, Forchheim, Germany (X.C.). From the 2002 RSNA Scientific Assembly. Received February 17, 2003; revision requested May 7; revision received June 25; accepted August 18. Address correspondence to J.F.P. (e-mail: pauljf@ccml.com).
PURPOSE: To assess the feasibility of low-kilovoltage (ie, 80-kV) chest computed tomography (CT) protocols for adults and the effect of such protocols on image quality and iodine dose.
MATERIALS AND METHODS: Preliminarily, 90 patients (30 women, 60 men; mean age, 59 years) requiring contrast materialenhanced chest CT were randomly assigned to one of three protocol groups: protocol A, with use of 80 kV and 135 mAs; protocol B, with use of 80 kV and 180 mAs; or the standard protocol, with use of 120 kV and 90 mAs. Contrast material injection protocols were standardized in all groups. Image noise was calculated and plotted against patient weight. Subsequently, another 52 consecutive patients (11 women, 41 men; mean age, 57 years) were assigned to one of the protocols according to their weight: Patients weighing less than 60 kg were assigned to protocol A; patients weighing 6075 kg, to protocol B; and patients weighing more than 75 kg, to the standard protocol. Two readers evaluated the CT images qualitatively by using a five-point scale. Statistical analyses were performed by using analysis of variance,
, and Fisher exact tests.
RESULTS: In the preliminary study, the mean image noise values with protocols A (24 HU) and B (20 HU) were significantly higher (P < .001) than that with the standard protocol (12 HU). With protocols A and B, in the patients weighing more than 60 kg and more than 75 kg, respectively, the noise increased exponentially with patient weight. In the subsequent study, qualitative analysis revealed no significant difference between the low-kilovoltage examinations and the standard examination. Compared with use of the standard protocol, use of protocols A and B resulted in the iodine-based contrast material dose being reduced by 54% and 39%, respectively.
CONCLUSION: Weight-adapted low-kilovoltage contrast-enhanced chest CT examinations can be routinely performed with 80 kV. Use of these protocols results in good diagnostic image quality and makes it possible to reduce contrast material use by more than 50%.
© RSNA, 2004
Index terms: Computed tomography (CT), multidetector row, 60.12112, 60.12114 Computed tomography (CT), radiation exposure Dosimetry Thorax, CT, 60.12112, 60.12114
This article has been cited by other articles:
![]() |
H Scheffel, H Alkadhi, S Leschka, A Plass, L Desbiolles, I Guber, T Krauss, J Gruenenfelder, M Genoni, T F Luescher, et al. Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance Heart, September 1, 2008; 94(9): 1132 - 1137. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Herzog, D. M. Mulvihill, S. A. Nguyen, G. Savino, B. Schmidt, P. Costello, T. J. Vogl, and U. J. Schoepf Pediatric Cardiovascular CT Angiography: Radiation Dose Reduction Using Automatic Anatomic Tube Current Modulation Am. J. Roentgenol., May 1, 2008; 190(5): 1232 - 1240. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kubo, P.-J. P. Lin, W. Stiller, M. Takahashi, H.-U. Kauczor, Y. Ohno, and H. Hatabu Radiation Dose Reduction in Chest CT: A Review Am. J. Roentgenol., February 1, 2008; 190(2): 335 - 343. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Heyer, P. S. Mohr, S. P. Lemburg, S. A. Peters, and V. Nicolas Image Quality and Radiation Exposure at Pulmonary CT Angiography with 100- or 120-kVp Protocol: Prospective Randomized Study Radiology, November 1, 2007; 245(2): 577 - 583. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. M. Kanal, B. K. Stewart, O. Kolokythas, and W. P. Shuman Impact of Operator-Selected Image Noise Index and Reconstruction Slice Thickness on Patient Radiation Dose in 64-MDCT Am. J. Roentgenol., July 1, 2007; 189(1): 219 - 225. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, E. Oechslin, L. Husmann, L. Desbiolles, B. Marincek, M. Genoni, R. Pretre, R. Jenni, S. Wildermuth, and H. Alkadhi Pre- and Postoperative Evaluation of Congenital Heart Disease in Children and Adults with 64-Section CT RadioGraphics, May 1, 2007; 27(3): 829 - 846. [Abstract] [Full Text] [PDF] |
||||
![]() |
A J Van Der Molen, W J H Veldkamp, and J Geleijns 16-slice CT: achievable effective doses of common protocols in comparison with recent CT dose surveys Br. J. Radiol., April 1, 2007; 80(952): 248 - 255. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Waaijer, M. Prokop, B. K. Velthuis, C. J. G. Bakker, G. A. P. de Kort, and M. S. van Leeuwen Circle of Willis at CT Angiography: Dose Reduction and Image Quality--Reducing Tube Voltage and Increasing Tube Current Settings Radiology, March 1, 2007; 242(3): 832 - 839. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. V. Sahani, S. P. Kalva, P. F. Hahn, and S. Saini 16-MDCT Angiography in Living Kidney Donors at Various Tube Potentials: Impact on Image Quality and Radiation Dose Am. J. Roentgenol., January 1, 2007; 188(1): 115 - 120. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Schueller-Weidekamm, C. M. Schaefer-Prokop, M. Weber, C. J. Herold, and M. Prokop CT Angiography of Pulmonary Arteries to Detect Pulmonary Embolism: Improvement of Vascular Enhancement with Low Kilovoltage Settings Radiology, December 1, 2006; 241(3): 899 - 907. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Nakayama, K. Awai, Y. Funama, D. Liu, T. Nakaura, Y. Tamura, and Y. Yamashita Lower tube voltage reduces contrast material and radiation doses on 16-MDCT aortography. Am. J. Roentgenol., November 1, 2006; 187(5): W490 - W497. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. T. Abada, C. Larchez, B. Daoud, A. Sigal-Cinqualbre, and J.-F. Paul MDCT of the Coronary Arteries: Feasibility of Low-Dose CT with ECG-Pulsed Tube Current Modulation to Reduce Radiation Dose Am. J. Roentgenol., June 1, 2006; 186(6_Supplement_2): S387 - S390. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hausleiter, T. Meyer, M. Hadamitzky, E. Huber, M. Zankl, S. Martinoff, A. Kastrati, and A. Schomig Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice: Impact of Different Scanning Protocols on Effective Dose Estimates Circulation, March 14, 2006; 113(10): 1305 - 1310. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Nakayama, K. Awai, Y. Funama, M. Hatemura, M. Imuta, T. Nakaura, D. Ryu, S. Morishita, S. Sultana, N. Sato, et al. Abdominal CT with Low Tube Voltage: Preliminary Observations about Radiation Dose, Contrast Enhancement, Image Quality, and Noise Radiology, December 1, 2005; 237(3): 945 - 951. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Funama, K. Awai, Y. Nakayama, K. Kakei, N. Nagasue, M. Shimamura, N. Sato, S. Sultana, S. Morishita, and Y. Yamashita Radiation Dose Reduction without Degradation of Low-Contrast Detectability at Abdominal Multisection CT with a Low-Tube Voltage Technique: Phantom Study Radiology, December 1, 2005; 237(3): 905 - 910. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Mulkens, P. Bellinck, M. Baeyaert, D. Ghysen, X. Van Dijck, E. Mussen, C. Venstermans, and J.-L. Termote Use of an Automatic Exposure Control Mechanism for Dose Optimization in Multi-Detector Row CT Examinations: Clinical Evaluation Radiology, October 1, 2005; 237(1): 213 - 223. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-F. Paul, M. Wartski, C. Caussin, A. Sigal-Cinqualbre, B. Lancelin, C. Angel, and G. Dambrin Late Defect on Delayed Contrast-enhanced Multi-Detector Row CT Scans in the Prediction of SPECT Infarct Size after Reperfused Acute Myocardial Infarction: Initial Experience Radiology, August 1, 2005; 236(2): 485 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. R. Rizzo, M. K. Kalra, B. Schmidt, J.-F. Paul, A. Sigal-Cinqualbre, and H. Abada Automatic Exposure Control Techniques for Individual Dose Adaptation * Dr Paul and colleagues respond: Radiology, April 1, 2005; 235(1): 335 - 336. [Full Text] [PDF] |
||||
![]() |
J.-F. Paul, H. T. Abada, and A. B. Sigal-Cinqualbre Automatic Dose Reduction Should Not Mask Needs for Individual Dose Reduction Radiology, October 1, 2004; 233(1): 297 - 297. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |