Experimental Studies |
1 From the Department of Diagnostic Radiology (P.L.P., J.T., I.S., C.T.R., D.S., C.D.S.), Section of Experimental Radiology (J.B.), and Departments of General Surgery (M.S.), Forensic Medicine (J.S.), and Pathology (S.K.), Eberhard-Karls-University, Hoppe-Seyler-Strasse 3, 72076 Tübingen, Germany. From the 2002 RSNA scientific assembly. Received February 13, 2003; revision requested May 1; final revision received October 31; accepted January 2, 2004. Address correspondence to P.L.P. (e-mail: philippe.pereira@med.uni-tuebingen.de).
PURPOSE: To compare in vivo coagulation necrosis obtained with four radiofrequency (RF) ablation devices, to determine shape and reproducibility of induced coagulation by means of three-dimensional measurements of the ablation zone, and to achieve representations of the coagulated areas in three-dimensional spaces.
MATERIALS AND METHODS: Four commercially available RF devices (perfusion, internally cooled cluster, and nine- and 12-tine expandable electrodes) that represent the most widely used systems on the market were tested. Sixteen in vivo ablation procedures were performed in porcine livers (four ablations for each RF system). After macroscopic and histopathologic analyses of 3-mm-thick liver sections, morphometric and volumetric findings in the central zone of white coagulation necrosis were assessed. Coagulation volume, diameter, length, and shape were determined digitally. After analysis of variance, measurements with each system were tested with the Tukey post hoc test.
RESULTS: Mean coagulation volumes were 31.5 cm3 ± 15.8 (SD) for the perfusion electrode, 20.5 cm3 ± 2.6 for the cluster electrode, 16.2 cm3 ± 7.3 for the 12-tine electrode, and 9.8 cm3 ± 3.2 for the nine-tine electrode (P < .05, perfusion vs nine-tine electrode). No significant differences were observed regarding the mean short axis perpendicular to the needle shaft: 2.30 cm ± 0.94, 3.04 cm ± 0.26, 3.44 cm ± 0.21, and 2.70 cm ± 0.76, respectively. Variation coefficients were 0.50, 0.13, 0.45, and 0.33, respectively.
CONCLUSION: Larger coagulation volumes were obtained with the perfusion and internally cooled cluster devices. More spherical volumes of ablation were achieved with the 12-tine and cluster electrodes. The former proved superior with regard to the short axis perpendicular to the needle shaft. The cluster and nine-tine electrode produced better reproducibility, which is suggestive of improved predictability of the extent of coagulation with these systems.
© RSNA, 2004
Index terms: Animals Experimental study, 761.1269 Liver, interventional procedures, 761.1269 Radiofrequency (RF) ablation, 761.1269
This article has been cited by other articles:
![]() |
O. Seror, G. N'Kontchou, M. Ibraheem, Y. Ajavon, C. Barrucand, N. Ganne, E. Coderc, J. Claude Trinchet, M. Beaugrand, and N. Sellier Large (>=5.0-cm) HCCs: Multipolar RF Ablation with Three Internally Cooled Bipolar Electrodes--Initial Experience in 26 Patients Radiology, May 15, 2008; (2008) 2481071101. [Abstract] [Full Text] |
||||
![]() |
H. C. Fernando Radiofrequency Ablation to Treat Non-Small Cell Lung Cancer and Pulmonary Metastases Ann. Thorac. Surg., February 1, 2008; 85(2): S780 - S784. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. F. Laeseke, T. M. Frey, C. L. Brace, L. A. Sampson, T. C. Winter III, J. R. Ketzler, and F. T. Lee Jr. Multiple-Electrode Radiofrequency Ablation of Hepatic Malignancies: Initial Clinical Experience Am. J. Roentgenol., June 1, 2007; 188(6): 1485 - 1494. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Mulier, Y. Ni, L. Frich, F. Burdio, A. L. Denys, J.-F. De Wispelaere, B. Dupas, N. Habib, M. Hoey, M. C. Jansen, et al. Experimental and Clinical Radiofrequency Ablation: Proposal for Standardized Description of Coagulation Size and Geometry Ann. Surg. Oncol., April 1, 2007; 14(4): 1381 - 1396. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. F. Laeseke, L. A. Sampson, D. Haemmerich, C. L. Brace, J. P. Fine, T. M. Frey, T. C. Winter III, and F. T. Lee Jr Multiple-Electrode Radiofrequency Ablation Creates Confluent Areas of Necrosis: In Vivo Porcine Liver Results Radiology, October 1, 2006; 241(1): 116 - 124. [Abstract] [Full Text] [PDF] |
||||
![]() |
J M Lee, J K Han, J M Chang, S Y Chung, S H Kim, J Y Lee, and B I Choi Radiofrequency ablation in pig lungs: in vivo comparison of internally cooled, perfusion and multitined expandable electrodes. Br. J. Radiol., July 1, 2006; 79(943): 562 - 571. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. U. Hines-Peralta, N. Pirani, P. Clegg, N. Cronin, T. P. Ryan, Z. Liu, and S. N. Goldberg Microwave Ablation: Results with a 2.45-GHz Applicator in ex Vivo Bovine and in Vivo Porcine Liver Radiology, April 1, 2006; 239(1): 94 - 102. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Clasen, D. Schmidt, A. Boss, K. Dietz, S. M. Krober, C. D. Claussen, and P. L. Pereira Multipolar Radiofrequency Ablation with Internally Cooled Electrodes: Experimental Study in ex Vivo Bovine Liver with Mathematic Modeling Radiology, March 1, 2006; 238(3): 881 - 890. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Shibata, T. Shibata, Y. Maetani, H. Isoda, and M. Hiraoka Radiofrequency Ablation for Small Hepatocellular Carcinoma: Prospective Comparison of Internally Cooled Electrode and Expandable Electrode Radiology, January 1, 2006; 238(1): 346 - 353. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Boss, S. Clasen, M. Kuczyk, A. Anastasiadis, D. Schmidt, C. D. Claussen, F. Schick, and P. L. Pereira Thermal Damage of the Genitofemoral Nerve Due to Radiofrequency Ablation of Renal Cell Carcinoma: A Potentially Avoidable Complication Am. J. Roentgenol., December 1, 2005; 185(6): 1627 - 1631. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Lee, J. K. Han, S. H. Kim, S. H. Choi, S. K. An, C. J. Han, and B. I. Choi Bipolar Radiofrequency Ablation Using Wet-Cooled Electrodes: An In Vitro Experimental Study in Bovine Liver Am. J. Roentgenol., February 1, 2005; 184(2): 391 - 397. [Abstract] [Full Text] [PDF] |
||||