1 From the Departments of Medical Physics (K.P., J.D., N.T.), Orthopaedics and Traumatology (G.P., A.H.), and Radiology (N.G.), University Hospital of Iraklion, University of Crete, Faculty of Medicine, PO Box 2208, 71003 Iraklion, Crete, Greece. Received September 2, 2003; revision requested November 20; revision received December 10; accepted January 13, 2004. Address correspondence to K.P. (e-mail: perisina@med.uoc.gr).
PURPOSE: To derive normalized data for the estimation of effective, gonadal, and peak skin doses to patients undergoing vertebroplasty or kyphoplasty and to investigate the potential for cancer induction, genetic effects, and radiation-induced skin injury after such procedures.
MATERIALS AND METHODS: Dose values normalized over dose-area product were determined for all radiosensitive organs and tissues by using a humanoid phantom and thermoluminescence dosimetry separately for anteroposterior and lateral projections. Measurements were obtained for treatments of the fifth, eighth, and 11th thoracic vertebrae and the first, third, and fifth lumbar vertebrae. Total fluoroscopy time and resultant dose-area product from each fluoroscopic exposure were monitored in 11 consecutive patients (seven women and four men) undergoing kyphoplasty. The age range of these patients was 4178 years, and the mean age was 58 years.
RESULTS: Mean total fluoroscopy time for kyphoplasty was 10.1 minutes ± 2.2 (standard deviation). Mean effective dose to patients from kyphoplasty was 8.512.7 mSv, and mean gonadal dose was 0.0416.4 mGy, depending on the level of the treated vertebra. Skin injuries after kyphoplasty are improbable if source-to-skin distance is 35 cm or more; however, such injuries may occur if the total fluoroscopy time per projection is extended and/or the source-to-skin distance is less than 35 cm during the procedure.
CONCLUSION: Patient radiation exposure and associated risks from vertebroplasty or kyphoplasty may be considerable. Data obtained in the current study may be used to establish patient effective dose, gonadal dose, and entrance skin exposure, as well as associated risks, from these fluoroscopically guided surgical treatments of spinal disorders.
© RSNA, 2004
Index terms: Dosimetry Experimental study Fluoroscopy, 32.1267 Radiations, injurious effects Phantoms Spine, surgery
This article has been cited by other articles:
![]() |
A.O. Ortiz, V. Natarajan, D.R. Gregorius, and S. Pollack Significantly Reduced Radiation Exposure to Operators during Kyphoplasty and Vertebroplasty Procedures: Methods and Techniques. AJNR Am. J. Neuroradiol., May 1, 2006; 27(5): 989 - 994. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. G. Hadjipavlou, M. N. Tzermiadianos, P. G. Katonis, and M. Szpalski Percutaneous vertebroplasty and balloon kyphoplasty for the treatment of osteoporotic vertebral compression fractures and osteolytic tumours J Bone Joint Surg Br, December 1, 2005; 87-B(12): 1595 - 1604. [Full Text] [PDF] |
||||
![]() |
D. L. Miller Patient Radiation Dose from Vertebroplasty and Kyphoplasty Radiology, March 1, 2005; 234(3): 970 - 971. [Full Text] [PDF] |
||||
![]() |
J. A. Seibert Vertebroplasty and Kyphoplasty: Do Fluoroscopy Operators Know about Radiation Dose, and Should They Want to Know? Radiology, September 1, 2004; 232(3): 633 - 634. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |