1 From the Department of Radiology, Division of Abdominal Imaging and Intervention, Massachusetts General Hospital and Harvard Medical School, White 270-E, 55 Fruit St, Boston, MA 02114 (M.K.K., M.M.M., R.S.K., E.F.H., S.S.); GE Yokogawa Medical Systems, Tokyo, Japan (T.H.); and GE Medical Systems, Waukesha, Wis (T.L.T.). From the 2003 RSNA scientific assembly. Received September 17, 2003; revision requested December 2; revision received January 6, 2004; accepted February 2. Address correspondence to M.K.K. (e-mail: mannudeep_k_kalra@yahoo.com).
PURPOSE: To retrospectively determine the optimal noise indexes required to obtain diagnostically acceptable computed tomographic (CT) images of the abdomen and pelvis with z-axis modulation.
MATERIALS AND METHODS: Ninety-five patients underwent 16-section multidetector row CT of the abdomen and pelvis with z-axis modulation at noise indexes of 10.5, 11.0, 11.5, and 12.0 HU with 10380 mA. Subsequently, 58 patients were scanned at noise indexes of 12.5 and 15.0 HU with 75380 mA. The weights of all subjects were recorded, and transverse and anteroposterior diameters were measured. The CT images were evaluated for abnormalities and graded for image quality in terms of noise and diagnostic acceptability by using a five-point scale. Objective noise in the liver parenchyma was measured, and the tube current was recorded at each section in all 153 patients. Statistical analyses were performed to determine the appropriate noise index and to assess the effect of patient weight and abdominal diameters on image noise and diagnostic acceptability at different noise indexes. Tube currenttime products (in milliampere seconds) at various noise indexes were compared with those at CT previously performed without z-axis modulation.
RESULTS: No significant difference in subjective image noise or diagnostic acceptability was found at noise indexes of 10.515.0 HU (P = .14), and objective noise was significantly inferior only at a noise index of 15.0 HU (P = .009). Compared with CT scanning at a 10.5-HU noise index, CT scanning at 12.5- and 15.0-HU noise indexes yielded, respectively, 10.0% and 41.3% reductions in radiation exposure. Patient weight and abdominal diameters affected subjective image quality.
CONCLUSION: Use of a 15.0-HU noise index at 75380 mA results in acceptable subjective image noise and diagnostic acceptability but significantly greater objective image noise at routine abdominal-pelvic CT. For greater image quality demands, a noise index of 12.5 HU results in acceptable image quality and a 19.6% reduction in radiation exposure.
© RSNA, 2004
Index terms: Abdomen, CT, 70.12111 Computed tomography (CT), image quality Computed tomography (CT), radiation exposure Pelvis, CT, 813.12111
This article has been cited by other articles:
![]() |
T. Miyoshi, M. Kanematsu, H. Kondo, S. Goshima, Y. Tsuge, A. Hatcho, Y. Shiratori, M. Onozuka, N. Moriyama, and K. T. Bae Abdomen: Angiography with 16-Detector CT--Comparison of Image Quality and Radiation Dose between Studies with 0.625-mm and those with 1.25-mm Collimation Radiology, October 1, 2008; 249(1): 142 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Roggenland, S. A. Peters, S. P. Lemburg, T. Holland-Letz, V. Nicolas, and C. M. Heyer CT Angiography in Suspected Pulmonary Embolism: Impact of Patient Characteristics and Different Venous Lines on Vessel Enhancement and Image Quality Am. J. Roentgenol., June 1, 2008; 190(6): W351 - W359. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Herzog, D. M. Mulvihill, S. A. Nguyen, G. Savino, B. Schmidt, P. Costello, T. J. Vogl, and U. J. Schoepf Pediatric Cardiovascular CT Angiography: Radiation Dose Reduction Using Automatic Anatomic Tube Current Modulation Am. J. Roentgenol., May 1, 2008; 190(5): 1232 - 1240. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Jaffe, T. T. Yoshizumi, G. I. Toncheva, G. Nguyen, L. M. Hurwitz, and R. C. Nelson Early First-Trimester Fetal Radiation Dose Estimation in 16-MDCT Without and With Automated Tube Current Modulation Am. J. Roentgenol., April 1, 2008; 190(4): 860 - 864. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. T. Schindera, R. C. Nelson, T. L. Toth, G. T. Nguyen, G. I. Toncheva, D. M. DeLong, and T. T. Yoshizumi Effect of Patient Size on Radiation Dose for Abdominal MDCT with Automatic Tube Current Modulation: Phantom Study Am. J. Roentgenol., February 1, 2008; 190(2): W100 - W105. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Matsubara, K. Koshida, M. Suzuki, H. Tsujii, T. Yamamoto, and O. Matsui Comparison between 3-D and z-axis automatic tube current modulation technique in multidetector-row CT Radiat Prot Dosimetry, January 1, 2008; 128(1): 106 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Coursey, D. P. Frush, T. Yoshizumi, G. Toncheva, G. Nguyen, and S. B. Greenberg Pediatric Chest MDCT Using Tube Current Modulation: Effect on Radiation Dose with Breast Shielding Am. J. Roentgenol., January 1, 2008; 190(1): W54 - W61. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. T. Schindera, R. C. Nelson, S. Mukundan Jr, E. K. Paulson, T. A. Jaffe, C. M. Miller, D. M. DeLong, K. Kawaji, T. T. Yoshizumi, and E. Samei Hypervascular Liver Tumors: Low Tube Voltage, High Tube Current Multi Detector Row CT for Enhanced Detection Phantom Study Radiology, December 1, 2007; 246(1): 125 - 132. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. D. Cody and M. Mahesh AAPM/RSNA Physics Tutorials AAPM/RSNA Physics Tutorial for Residents: Technologic Advances in Multidetector CT with a Focus on Cardiac Imaging RadioGraphics, November 1, 2007; 27(6): 1829 - 1837. [Abstract] [Full Text] [PDF] |
||||
![]() |
A J Van Der Molen, W J H Veldkamp, and J Geleijns 16-slice CT: achievable effective doses of common protocols in comparison with recent CT dose surveys Br. J. Radiol., April 1, 2007; 80(952): 248 - 255. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Li, U. K. Udayasankar, T. L. Toth, J. Seamans, W. C. Small, and M. K. Kalra Automatic Patient Centering for MDCT: Effect on Radiation Dose Am. J. Roentgenol., February 1, 2007; 188(2): 547 - 552. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. C. Dalrymple, S. R. Prasad, F. M. El-Merhi, and K. N. Chintapalli Price of Isotropy in Multidetector CT RadioGraphics, January 1, 2007; 27(1): 49 - 62. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Namasivayam, M.K. Kalra, K.M. Pottala, S.M. Waldrop, and P.A. Hudgins Optimization of Z-Axis Automatic Exposure Control for Multidetector Row CT Evaluation of Neck and Comparison with Fixed Tube Current Technique for Image Quality and Radiation Dose AJNR Am. J. Neuroradiol., November 1, 2006; 27(10): 2221 - 2225. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Graser, B. J. Wintersperger, C. Suess, M. F. Reiser, and C. R. Becker Dose reduction and image quality in MDCT colonography using tube current modulation. Am. J. Roentgenol., September 1, 2006; 187(3): 695 - 701. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Rizzo, M. Kalra, B. Schmidt, T. Dalal, C. Suess, T. Flohr, M. Blake, and S. Saini Comparison of Angular and Combined Automatic Tube Current Modulation Techniques with Constant Tube Current CT of the Abdomen and Pelvis. Am. J. Roentgenol., March 1, 2006; 186(3): 673 - 679. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Kalra, S. Rizzo, M. M. Maher, E. F. Halpern, T. L. Toth, J.-A. O. Shepard, and S. L. Aquino Chest CT Performed with Z-Axis Modulation: Scanning Protocol and Radiation Dose Radiology, October 1, 2005; 237(1): 303 - 308. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. M. Chapman, M. Kalra, E. Halpern, B. Grottkau, M. Albright, and D. Jaramillo 16-MDCT of the Posttraumatic Pediatric Elbow: Optimum Parameters and Associated Radiation Dose Am. J. Roentgenol., August 1, 2005; 185(2): 516 - 521. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Kalra, M. M. Maher, R. V. D'Souza, S. Rizzo, E. F. Halpern, M. A. Blake, and S. Saini Detection of Urinary Tract Stones at Low-Radiation-Dose CT with Z-Axis Automatic Tube Current Modulation: Phantom and Clinical Studies Radiology, May 1, 2005; 235(2): 523 - 529. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Vrtiska, J. G. Fletcher, and C. H. McCollough State-of-the-Art Imaging With 64-Channel Multidetector CT Angiography Perspectives in Vascular Surgery and Endovascular Therapy, March 1, 2005; 17(1): 3 - 8. [Abstract] [PDF] |
||||
![]() |
S. M. R. Rizzo, M. K. Kalra, M. M. Maher, M. A. Blake, T. L. Toth, and S. Saini Do Metallic Endoprostheses Increase Radiation Dose Associated with Automatic Tube-Current Modulation in Abdominal-Pelvic MDCT? A Phantom and Patient Study Am. J. Roentgenol., February 1, 2005; 184(2): 491 - 496. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Kalra, M. M. Maher, T. L. Toth, B. Schmidt, B. L. Westerman, H. T. Morgan, and S. Saini Techniques and Applications of Automatic Tube Current Modulation for CT Radiology, December 1, 2004; 233(3): 649 - 657. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |