Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


DOI: 10.1148/radiol.2361040618
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lembcke, A.
Right arrow Articles by Enzweiler, C. N. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lembcke, A.
Right arrow Articles by Enzweiler, C. N. H.
(Radiology 2005;236:47-55.)
© RSNA, 2005


Cardiac Imaging

Assessment of Mitral Valve Regurgitation at Electron-Beam CT: Comparison with Doppler Echocardiography1

Alexander Lembcke, MD, Adrian C. Borges, MD, Simon Dushe, MD, Pascal M. Dohmen, MD, Till H. Wiese, MD, Patrik Rogalla, MD, Kay-Geert A. Hermann, MD, Bernd Hamm, MD and Christian N. H. Enzweiler, MD

1 From the Departments of Radiology (A.L., T.H.W., P.R., K.G.A.H., B.H., C.N.H.E.), Internal Medicine I–Cardiology, Angiology, and Pneumology Section (A.C.B.), and Cardiovascular Surgery (S.D., P.M.D.), Charité Medical School, Humboldt University, Berlin, Germany. Received April 4, 2004; revision requested June 17; revision received August 2; accepted September 23. Address correspondence to A.L., Department of Radiology, Massachusetts General Hospital, Harvard Medical School, 55 Fruit St, White Bldg 270, Boston, MA 02114 (e-mail: alexander.lembcke{at}gmx.de).

PURPOSE: To prospectively compare mitral valve regurgitation fractions calculated at electron-beam computed tomography (CT) (Doppler echocardiography as reference standard) and to evaluate accuracy of electron-beam CT volume and flow measurements compared with magnetic resonance (MR) imaging results.

MATERIALS AND METHODS: Institutional review board approval and informed consent were obtained. Volume and flow measurements were performed at electron-beam CT in 219 patients (197 men, 22 women; mean age, 61.5 years ± 10.4 [standard deviation]), of whom 157 had known isolated mitral valve regurgitation. Regurgitation volume was calculated as the difference between left ventricular total and forward stroke volumes. Regurgitation fractions were compared with corresponding echocardiographic grades (grades 0–IV) by using Spearman rank correlation and a weighted {kappa} test. In 22 patients, CT volume and flow measurements were compared with MR results by using intraclass correlation.

RESULTS: Regurgitation fractions at CT correlated well with echocardiographic grading (rank correlation coefficient, rS = 0.82; P < .05). Mean regurgitation fractions for echocardiographic grades 0, I, II, III, and IV were 3.1% ± 6.2, 12.7% ± 9.9, 25.3% ± 12.3, 40.4% ± 11.5, and 55.9% ± 13.7, respectively. The most suitable thresholds for differentiating echocardiographic grades were calculated regurgitation fractions of 6%, 20%, 30%, and 44%; with these thresholds, individual echocardiographic grades were differentiated (grades 0 vs I–IV, 0–I vs II–IV, 0–II vs III–IV, and 0–III vs IV, respectively) with sensitivities of 89%, 87%, 86%, and 93% and specificities of 81%, 87%, 92%, and 91%, respectively. There was perfect agreement in classification of mitral valve insufficiency between electron-beam CT and echocardiography in 134 (61%) patients and a mismatch by one grade in 72 (33%) and by two grades in 13 (6%) ({kappa} = 0.84). Intraclass correlation coefficients between CT and MR imaging for total and forward stroke volumes and regurgitation volume and fraction were 0.88, 0.79, 0.93, and 0.89, respectively.

CONCLUSION: Electron-beam CT provides quantitative information on severity of mitral valve regurgitation, but semiquantitative classification of regurgitation showed mismatch between electron-beam CT and Doppler echocardiography by at least one grade in more than one-third of all patients.

© RSNA, 2005




This article has been cited by other articles:


Home page
RadiologyHome page
Y.-K. Guo, Z.-G. Yang, G. Ning, L. Rao, L. Dong, Y. Pen, T.-M. Zhang, Y. Wu, X.-C. Zhang, and Q.-L. Wang
Isolated Mitral Regurgitation: Quantitative Assessment with 64-Section Multidetector CT--Comparison with MR Imaging and Echocardiography
Radiology, May 18, 2009; (2009) 2522081714.
[Abstract] [Full Text]


Home page
Ann. Thorac. Surg.Home page
P. M. Dohmen, A. Lembcke, S. Holinski, D. Kivelitz, J. P. Braun, A. Pruss, and W. Konertz
Mid-Term Clinical Results Using a Tissue-Engineered Pulmonary Valve to Reconstruct the Right Ventricular Outflow Tract During the Ross Procedure
Ann. Thorac. Surg., September 1, 2007; 84(3): 729 - 736.
[Abstract] [Full Text] [PDF]