PURPOSE: To prospectively compare dose reduction and image quality achieved with an automatic exposure control system that is based on both angular (x-y axis) and z-axis tube current modulation with dose reduction and image quality achieved with an angular modulation system for multidetector row computed tomography (CT).
MATERIALS AND METHODS: The study protocol was approved by the institutional review board, and oral informed consent was obtained. In two groups of 200 patients, five anatomic regions (ie, the thorax, abdomen-pelvis, abdomen-liver, lumbar spine, and cervical spine) were examined with this modulation system and a six-section multidetector row CT scanner. Data from these patients were compared with data from 200 patients who were examined with an angular modulation system. Dose reduction by means of reduction of the mean effective tube current in 600 examinations, image noise in 200 examinations performed with each modulation system, and subjective image quality scores in 100 examinations per-formed with each modulation system were compared with Wilcoxon signed rank tests.
RESULTS: Mean dose reduction for the angular and z-axis tube current modulation system and for the angular modulation system was as follows: thorax, 20% and 14%, respectively; abdomen-liver, 38% and 18%, respectively; abdomen-pelvis, 32% and 26%, respectively; lumbar spine, 37% and 10%, respectively; and cervical spine, 68% and 16%, respectively. These differences were statistically significant (P < .05). There was no significant difference in image noise and mean image quality scores between modulation systems, with the exception of cervical spinal examinations (P < .001 for both), where the examinations with angular modulation resulted in better scores. There is good correlation between the mean effective tube current level and the body mass index of patients with the new modulation system. Correlation was as follows: thorax, 0.77; abdomen-pelvis, 0.83; abdomen-liver, 0.84; lumbar spine, 0.8; and cervical spine, 0.6. This correlation was not observed with the angular modulation system.
CONCLUSION: An automatic exposure control mechanism that is based on real-time anatomy-dependent tube current modulation delivers good image quality with a significantly reduced radiation dose.
© RSNA, 2005
This article has been cited by other articles:
![]() |
S. Matsuoka, A. R. Hunsaker, R. R. Gill, I. B. Oliva, B. Trotman-Dickenson, F. L. Jacobson, and H. Hatabu Vascular Enhancement and Image Quality of MDCT Pulmonary Angiography in 400 Cases: Comparison of Standard and Low Kilovoltage Settings Am. J. Roentgenol., June 1, 2009; 192(6): 1651 - 1656. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Matsubara, K. Koshida, K. Ichikawa, M. Suzuki, T. Takata, T. Yamamoto, and O. Matsui Misoperation of CT Automatic Tube Current Modulation Systems with Inappropriate Patient Centering: Phantom Studies Am. J. Roentgenol., April 1, 2009; 192(4): 862 - 865. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.S.P. Tan, K.-L. Tan, J.C.L. Lee, C.-M. Wan, J.-L. Leong, and L.-L. Chan Comparison of Eye Lens Dose on Neuroimaging Protocols between 16- and 64-Section Multidetector CT: Achieving the Lowest Possible Dose AJNR Am. J. Neuroradiol., February 1, 2009; 30(2): 373 - 377. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Leswick, M. M. Hunt, S. T. Webster, and D. A. Fladeland Thyroid Shields versus z-Axis Automatic Tube Current Modulation for Dose Reduction at Neck CT Radiology, November 1, 2008; 249(2): 572 - 580. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Phillips, D. L. Stella, W. M. Rozen, M. Ashton, and G. I. Taylor Abdominal Wall CT Angiography: A Detailed Account of a Newly Established Preoperative Imaging Technique Radiology, October 1, 2008; 249(1): 32 - 44. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. H. Lee, J. M. Goo, H. J. Ye, S.-J. Ye, C. M. Park, E. J. Chun, and J.-G. Im Radiation Dose Modulation Techniques in the Multidetector CT Era: From Basics to Practice RadioGraphics, September 1, 2008; 28(5): 1451 - 1459. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Israel, S. Herlihy, A. N. Rubinowitz, D. Cornfeld, and J. Brink Does a Combination of Dose Modulation with Fast Gantry Rotation Time Limit CT Image Quality? Am. J. Roentgenol., July 1, 2008; 191(1): 140 - 144. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. B. Smith, W. P. Dillon, B. C. Lau, R. Gould, F. R. Verdun, E. B. Lopez, and M. Wintermark Radiation Dose Reduction Strategy for CT Protocols: Successful Implementation in Neuroradiology Section Radiology, May 1, 2008; 247(2): 499 - 506. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.T. Russell, J.R. Fink, F. Rebeles, K. Kanal, M. Ramos, and Y. Anzai Balancing Radiation Dose and Image Quality: Clinical Applications of Neck Volume CT AJNR Am. J. Neuroradiol., April 1, 2008; 29(4): 727 - 731. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Jaffe, T. T. Yoshizumi, G. I. Toncheva, G. Nguyen, L. M. Hurwitz, and R. C. Nelson Early First-Trimester Fetal Radiation Dose Estimation in 16-MDCT Without and With Automated Tube Current Modulation Am. J. Roentgenol., April 1, 2008; 190(4): 860 - 864. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kubo, P.-J. P. Lin, W. Stiller, M. Takahashi, H.-U. Kauczor, Y. Ohno, and H. Hatabu Radiation Dose Reduction in Chest CT: A Review Am. J. Roentgenol., February 1, 2008; 190(2): 335 - 343. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. T. Schindera, R. C. Nelson, T. L. Toth, G. T. Nguyen, G. I. Toncheva, D. M. DeLong, and T. T. Yoshizumi Effect of Patient Size on Radiation Dose for Abdominal MDCT with Automatic Tube Current Modulation: Phantom Study Am. J. Roentgenol., February 1, 2008; 190(2): W100 - W105. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, I. Valenta, O. Gaemperli, O. Adda, V. Treyer, C. A. Wyss, P. Veit-Haibach, F. Tatsugami, G. K. von Schulthess, and P. A. Kaufmann Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating Eur. Heart J., January 2, 2008; 29(2): 191 - 197. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Matsubara, K. Koshida, M. Suzuki, H. Tsujii, T. Yamamoto, and O. Matsui Comparison between 3-D and z-axis automatic tube current modulation technique in multidetector-row CT Radiat Prot Dosimetry, January 1, 2008; 128(1): 106 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Coursey, D. P. Frush, T. Yoshizumi, G. Toncheva, G. Nguyen, and S. B. Greenberg Pediatric Chest MDCT Using Tube Current Modulation: Effect on Radiation Dose with Breast Shielding Am. J. Roentgenol., January 1, 2008; 190(1): W54 - W61. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.B. Smith, W.P. Dillon, R. Gould, and M. Wintermark Radiation Dose-Reduction Strategies for Neuroradiology CT Protocols AJNR Am. J. Neuroradiol., October 1, 2007; 28(9): 1628 - 1632. [Abstract] [Full Text] [PDF] |
||||
![]() |
T.H. Mulkens, P. Marchal, S. Daineffe, R. Salgado, P. Bellinck, B. t. Rijdt, B. Kegelaers, and J.-L. Termote Comparison of Low-Dose With Standard-Dose Multidetector CT in Cervical Spine Trauma AJNR Am. J. Neuroradiol., September 1, 2007; 28(8): 1444 - 1450. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Deetjen, S. Mollmann, G. Conradi, A. Rolf, A. Schmermund, C. W Hamm, and T. Dill Use of automatic exposure control in multislice computed tomography of the coronaries: comparison of 16-slice and 64-slice scanner data with conventional coronary angiography Heart, September 1, 2007; 93(9): 1040 - 1043. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Fujii, T Aoyama, S Koyama, and C Kawaura Comparative evaluation of organ and effective doses for paediatric patients with those for adults in chest and abdominal CT examinations Br. J. Radiol., August 1, 2007; 80(956): 657 - 667. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Mulkens, S. Daineffe, R. De Wijngaert, P. Bellinck, A. Leonard, G. Smet, and J.-L. Termote Urinary Stone Disease: Comparison of Standard-Dose and Low-Dose with 4D MDCT Tube Current Modulation Am. J. Roentgenol., February 1, 2007; 188(2): 553 - 562. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Namasivayam, M.K. Kalra, K.M. Pottala, S.M. Waldrop, and P.A. Hudgins Optimization of Z-Axis Automatic Exposure Control for Multidetector Row CT Evaluation of Neck and Comparison with Fixed Tube Current Technique for Image Quality and Radiation Dose AJNR Am. J. Neuroradiol., November 1, 2006; 27(10): 2221 - 2225. [Abstract] [Full Text] [PDF] |
||||