Musculoskeletal Imaging |
1 From the Department of Radiology, University of California, San Francisco, 400 Parnassus Ave, A 367, Box 0628, San Francisco, CA 94143-0628 (C.M.P., J.S.B., T.C.D., D.N., S.M., T.M.L.); Institute for Anatomy, University of Munich, Munich, Germany (M.M.); 1st Gynaecology Hospital, LMU, Munich, Germany (E.M.L.); and Institute for Anatomy, Paracelsus Medical Private University, Salzburg, Austria (F.E.). Received April 7, 2005; revision requested June 6; revision received June 30; accepted August 2. Supported by the Société Française de Radiologie, Philippe Foundation, Guerbet, and the German Research Foundation grant (Lo 730/3-1 and 3-2). Address correspondence to T.M.L. (e-mail: tmlink{at}radiology.ucsf.edu).
Purpose: To investigate in vitro the calcaneal trabecular bone structure in elderly human donors with high spatial resolution magnetic resonance (MR) imaging at 3.0 T and 1.5 T, to quantitatively compare MR measures of bone microarchitecture with those from microcomputed tomography (CT), and to compare the performance of 3.0-T MR imaging with that of 1.5-T MR imaging in differentiating donors with spinal fractures from those without spinal fractures.
Materials and Methods: The study was performed in line with institutional and legislative requirements; all donors had dedicated their body for educational and research purposes prior to death. Sagittal MR images of 49 human calcaneus cadaveric specimens were obtained (mean age of donors, 79.5 years ± 11 [standard deviation]; 26 male donors, 23 female donors). After the spatial coregistering of images acquired at 3.0-T and 1.5-T MR imaging, the signal-to-noise-ratios and structural parameters obtained at each magnetic field strength were compared in corresponding sections. Micro-CT was performed on calcaneus cores obtained from corresponding regions in 40 cadaveric specimens. Vertebral deformities of the thoracic and lumbar spine were radiographically classified by using the spinal fracture index. Diagnostic performance of the structural parameters in differentiating donors with vertebral fractures from those without was assessed by using receiver operator characteristic (ROC) analysis, including area under the ROC curve (Az).
Results: Correlations between structural parameters at 3.0-T MR imaging and those at micro-CT were significantly higher (P < .05) than correlations between structural parameters at 1.5-T MR imaging and those at micro-CT (trabecular thickness, r = 0.76 at 3.0 T vs r = 0.57 at 1.5 T). Trabecular dimensions were amplified at 3.0 T because of increasing susceptibility artifacts. Also, higher ROC values were found for structural parameters at 3.0 T than at 1.5 T, but differences were not significant (trabecular thickness, Az = 0.75 at 3.0 T vs Az = 0.66 at 1.5 T, P > .05).
Conclusion: MR imaging at 3.0 T provided a better measure of the trabecular bone structure than did MR imaging at 1.5 T. There was a trend for better differentiation of donors with from those without osteoporotic vertebral fractures at 3.0 T than at 1.5 T.
© RSNA, 2006
This article has been cited by other articles:
![]() |
M. A. Bredella, M. Misra, K. K. Miller, I. Madisch, A. Sarwar, A. Cheung, A. Klibanski, and R. Gupta Distal Radius in Adolescent Girls with Anorexia Nervosa: Trabecular Structure Analysis with High-Resolution Flat-Panel Volume CT Radiology, December 1, 2008; 249(3): 938 - 946. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. C. Strohm, D. Kubosch, T. A. Bley, C. M. Sprecher, N. P. Sudkamp, and S. Milz Detection of Bone Graft Failure in Lumbar Spondylodesis: Spatial Resolution with High-Resolution Peripheral Quantitative CT Am. J. Roentgenol., May 1, 2008; 190(5): 1255 - 1259. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.G. Hatipoglu, A. Selvi, D. Ciliz, and E. Yuksel Quantitative and Diffusion MR Imaging as a New Method To Assess Osteoporosis AJNR Am. J. Neuroradiol., November 1, 2007; 28(10): 1934 - 1937. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Bauer, S. J. Krause, C. J. Ross, R. Krug, J. Carballido-Gamio, E. Ozhinsky, S. Majumdar, and T. M. Link Volumetric Cartilage Measurements of Porcine Knee at 1.5-T and 3.0-T MR Imaging: Evaluation of Precision and Accuracy. Radiology, November 1, 2006; 241(2): 399 - 406. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |