Breast Imaging |
1 From the Departments of Radiology (C.K.K., P.J., N.M., H.H.S., J.G.) and Gynecology (O.Z.), University of Bonn, Sigmund-Freud-Str 25, 53105 Bonn, Germany, and Philips Medical Systems, Best, the Netherlands (J.G.). From the 2004 RSNA Annual Meeting. Received March 27, 2005; revision requested May 24; revision received May 26; accepted June 21; final version accepted July 26. Address correspondence to C.K.K. (e-mail: kuhl{at}uni-bonn.de).
Purpose: To establish a pulse sequence for dynamic contrast materialenhanced magnetic resonance (MR) imaging of the breast at 3.0 T and to prospectively compare MR imaging at 3.0 T with MR imaging at 1.5 T in the same patients.
Materials and Methods: A prospective intraindividual internal review boardapproved study was performed in 37 women with 53 lesions (25 breast cancers, 28 benign focal lesions) who underwent contrast-enhanced dynamic bilateral subtraction MR imaging twice, once at 1.5 T with a standard technique (voxel size, 1.44 mm3) and once at 3.0 T (voxel size, 0.450.72 mm3) with variable repetition time and flip angle settings. Written informed consent was obtained. Sagittal single breast high-spatial-resolution MR imaging was performed with active fat suppression. Image quality, number and features of enhancing lesions, and Breast Imaging Reporting and Data System categories were compared by using the Wilcoxon matched-pairs signed rank test and Student t test for matched pairs. Diagnostic confidence was compared by using a receiver operating characteristic (ROC) analysis.
Results: With repetition time prolonged to account for longer T1 relaxation times at 3.0 T and a flip angle of 60°, enhancement rates at 3.0 T were substantially below those at 1.5 T. In two patients with benign lesions, enhancement was rated as insufficient to establish diagnosis. When parameter settings were kept equivalent, equivalent enhancement rates were observed with both systems. With these settings, 3.0-T MR imaging yielded homogeneous signal intensity over the entire field of view. No dielectric resonance effects were observed. Overall image quality scores for the dynamic series were slightly higher at 3.0 T (P < .01). A total of 49 lesions were prospectively identified with both systems. Owing to substantial patient motion at 1.5 T, two malignant lesions in one patient were visualized at 3.0 T only. At 3.0 T, differential diagnosis of enhancing lesions was possible with higher diagnostic confidence, as reflected by a larger area under the ROC curve (P < .05).
Conclusion: Initial experiences indicate that contrast-enhanced MR imaging at 3.0 T is nearing readiness for clinical use.
© RSNA, 2006
This article has been cited by other articles:
![]() |
A M BLAMIRE The technology of MRI -- the next 10 years? Br. J. Radiol., August 1, 2008; 81(968): 601 - 617. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. E. Mountford, P. Stanwell, S. Ramadan, C. K. Kuhl, H. Kooijman, J. Gieseke, and H. H. Schild Breast MR Imaging at 3.0 T Radiology, July 1, 2008; 248(1): 319 - 320. [Full Text] [PDF] |
||||
![]() |
C. K. Kuhl, F. Traber, J. Gieseke, W. Drahanowsky, N. Morakkabati-Spitz, W. Willinek, M. von Falkenhausen, C. Manka, and H. H. Schild Whole-Body High-Field-Strength (3.0-T) MR Imaging in Clinical Practice * Part II. Technical Considerations and Clinical Applications Radiology, April 1, 2008; 247(1): 16 - 35. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. K. Kuhl, F. Traber, and H. H. Schild Whole-Body High-Field-Strength (3.0-T) MR Imaging in Clinical Practice * Part I. Technical Considerations and Clinical Applications Radiology, March 1, 2008; 246(3): 675 - 696. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. S. Lee, E. M. Hecht, B. Taouli, Q. Chen, K. Prince, and N. Oesingmann Body and Cardiovascular MR Imaging at 3.0 T Radiology, September 1, 2007; 244(3): 692 - 705. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. K. Kuhl, H. Kooijman, J. Gieseke, and H. H. Schild Effect of B1 Inhomogeneity on Breast MR Imaging at 3.0 T Radiology, September 1, 2007; 244(3): 929 - 930. [Full Text] [PDF] |
||||
![]() |
T. G. ODLE Breast MR Radiol. Technol., September 1, 2006; 78(1): 45M - 66M. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |