Cardiac Imaging |
1 From the Institute of Diagnostic Radiology (S.L., S.W., T.B., L.D., B.M., H.A.), the Clinic for Cardiovascular Surgery (A.P.), and the Cardiovascular Center (P.K., T.S., P.A.K.), University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland; and the Center for Integrative Human Physiology, University of Zurich (P.A.K.). Received August 17, 2005; revision requested October 20; revision received December 29; final version accepted February 1, 2006. Supported by the National Center of Competence in Research, Computer Aided and Image Guided Medical Interventions of the Swiss National Science Foundation and by the Georg und Bertha Schwyzer-Winiker-Stiftung, Zurich, Switzerland. Address correspondence to H.A. (e-mail: hatem.alkadhi{at}usz.ch).
Purpose: To evaluate prospectively the effect of average heart rate and heart rate variability on image quality at 64-section computed tomographic (CT) coronary angiography.
Materials and Methods: The study protocol had local ethics committee approval; written informed consent was obtained. There were 125 patients (45 women, 80 men; mean age, 59.9 years ± 12.9 [standard deviation]; 79 receiving ß-blockers) who underwent 64-section CT coronary angiography with retrospective electrocardiographic gating. Data sets were reconstructed in 5% steps from 20% to 80% of R-R interval. Heart rate variability was calculated as 1 standard deviation from mean rate during scanning. Two observers rated image quality of each coronary segment at least 1.5-mm diameter (1 = no motion artifacts, 5 = not evaluative). Repeated analysis of variance measurements were performed to evaluate quantitative parameters. Pearson correlation analysis was performed to compare image quality in each patient with average heart rate and heart rate variability.
Results: Average heart rate was 63.3 beats per minute ± 13.1, with variability of 3.2 beats per minute ± 2.1. Diagnostic image quality (score
3) was attained in 1821 of 1836 segments at the best reconstruction interval. There was no correlation between mean heart rate and image quality for all segments of the right coronary and left anterior descending arteries, but there was a significant correlation for left circumflex artery (r = 0.33, P < .05). Heart rate variability was correlated with image quality overall (r = 0.75, P < .001) and for each coronary artery. Heart rate was less variable and image quality was better (P < .05) in patients receiving ß-blockers. Best image quality was obtained in diastole with heart rate less than 80 beats per minute and in systole with faster heart rate.
Conclusion: Coronary angiography with 64-section CT provides diagnostic image quality within a wide range of heart rates. Reducing average heart rate and heart rate variability is beneficial for reducing artifacts.
© RSNA, 2006
This article has been cited by other articles:
![]() |
S. Baumuller, S. Leschka, L. Desbiolles, P. Stolzmann, H. Scheffel, B. Seifert, B. Marincek, and H. Alkadhi Dual-Source versus 64-Section CT Coronary Angiography at Lower Heart Rates: Comparison of Accuracy and Radiation Dose Radiology, July 8, 2009; (2009) 2531090065. [Abstract] [Full Text] |
||||
![]() |
W T ROBERTS, A R WRIGHT, J B TIMMIS, and A D TIMMIS Safety and efficacy of a rate control protocol for cardiac CT Br. J. Radiol., April 1, 2009; 82(976): 267 - 271. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, B. A. Herzog, O. Gaemperli, F. Tatsugami, N. Burkhard, I. Valenta, P. Veit-Haibach, C. A. Wyss, U. Landmesser, and P. A. Kaufmann Diagnostic accuracy of computed tomography coronary angiography and evaluation of stress-only single-photon emission computed tomography/computed tomography hybrid imaging: comparison of prospective electrocardiogram-triggering vs. retrospective gating Eur. Heart J., March 1, 2009; 30(5): 600 - 607. [Abstract] [Full Text] [PDF] |
||||
![]() |
L ROMANO, L GRAZIOLI, L BONOMO, J-R XU, K-M CHEN, R DORE, A VANZULLI, and C CATALANO Enhancement and safety of iomeprol-400 and iodixanol-320 in patients undergoing abdominal multidetector CT Br. J. Radiol., March 1, 2009; 82(975): 204 - 211. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Herzog, L. Husmann, N. Burkhard, O. Gaemperli, I. Valenta, F. Tatsugami, C. A. Wyss, U. Landmesser, and P. A. Kaufmann Accuracy of low-dose computed tomography coronary angiography using prospective electrocardiogram-triggering: first clinical experience Eur. Heart J., December 2, 2008; 29(24): 3037 - 3042. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Matsumoto, T. Kondo, S. Watanabe, R. Kikumoto, T. Shimada, Y. Hiraoka, and K. Ueda ECG-Edited Middiastolic Phase Reconstruction Improves Image Quality at 64-MDCT Coronary Angiography of Patients with Atrial Fibrillation Am. J. Roentgenol., December 1, 2008; 191(6): 1659 - 1666. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, H. Scheffel, S. Leschka, A. Plass, S. Baumuller, B. Marincek, and H. Alkadhi Influence of Calcifications on Diagnostic Accuracy of Coronary CT Angiography Using Prospective ECG Triggering Am. J. Roentgenol., December 1, 2008; 191(6): 1684 - 1689. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, S. Leschka, H. Scheffel, T. Krauss, L. Desbiolles, A. Plass, M. Genoni, T. G. Flohr, S. Wildermuth, B. Marincek, et al. Dual-Source CT in Step-and-Shoot Mode: Noninvasive Coronary Angiography with Low Radiation Dose Radiology, October 1, 2008; 249(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Leschka, H Scheffel, L Desbiolles, A Plass, O Gaemperli, P Stolzmann, M Genoni, T Luescher, B Marincek, P Kaufmann, et al. Combining dual-source computed tomography coronary angiography and calcium scoring: added value for the assessment of coronary artery disease Heart, September 1, 2008; 94(9): 1154 - 1161. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Bastarrika, C. N. De Cecco, M. Arraiza, S. Mastrobuoni, J. C. Pueyo, M. Ubilla, and G. Rabago Dual-Source CT for Visualization of the Coronary Arteries in Heart Transplant Patients with High Heart Rates Am. J. Roentgenol., August 1, 2008; 191(2): 448 - 454. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Gaemperli, T. Schepis, I. Valenta, P. Koepfli, L. Husmann, H. Scheffel, S. Leschka, F. R. Eberli, T. F. Luscher, H. Alkadhi, et al. Functionally Relevant Coronary Artery Disease: Comparison of 64-Section CT Angiography with Myocardial Perfusion SPECT Radiology, August 1, 2008; 248(2): 414 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Hirai, J. Horiguchi, C. Fujioka, M. Kiguchi, H. Yamamoto, N. Matsuura, T. Kitagawa, H. Teragawa, N. Kohno, and K. Ito Prospective versus Retrospective ECG-gated 64-Detector Coronary CT Angiography: Assessment of Image Quality, Stenosis, and Radiation Dose Radiology, August 1, 2008; 248(2): 424 - 430. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Cheng, A. Gutstein, A. Wolak, Y. Suzuki, D. Dey, H. Gransar, L. E.J. Thomson, S. W. Hayes, J. D. Friedman, and D. S. Berman Moving beyond binary grading of coronary arterial stenoses on coronary computed tomographic angiography insights for the imager and referring clinician. J. Am. Coll. Cardiol. Img., July 1, 2008; 1(4): 460 - 471. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Oncel, G. Oncel, A. Tastan, and B. Tamci Evaluation of Coronary Stent Patency and In-Stent Restenosis with Dual-Source CT Coronary Angiography Without Heart Rate Control Am. J. Roentgenol., July 1, 2008; 191(1): 56 - 63. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Matsuura, J. Horiguchi, H. Yamamoto, N. Hirai, T. Tonda, N. Kohno, and K. Ito Optimal Cardiac Phase for Coronary Artery Calcium Scoring on Single-Source 64-MDCT Scanner: Least Interscan Variability and Least Motion Artifacts Am. J. Roentgenol., June 1, 2008; 190(6): 1561 - 1568. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, H. Scheffel, L. Husmann, O. Gamperli, B. Marincek, P. A. Kaufmann, and H. Alkadhi Effect of Decrease in Heart Rate Variability on the Diagnostic Accuracy of 64-MDCT Coronary Angiography Am. J. Roentgenol., June 1, 2008; 190(6): 1583 - 1590. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Brodoefel, C. Burgstahler, I. Tsiflikas, A. Reimann, S. Schroeder, C. D. Claussen, M. Heuschmid, and A. F. Kopp Dual-Source CT: Effect of Heart Rate, Heart Rate Variability, and Calcification on Image Quality and Diagnostic Accuracy Radiology, May 1, 2008; 247(2): 346 - 355. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Alkadhi, H. Scheffel, L. Desbiolles, O. Gaemperli, P. Stolzmann, A. Plass, G. W. Goerres, T. F. Luescher, M. Genoni, B. Marincek, et al. Dual-source computed tomography coronary angiography: influence of obesity, calcium load, and heart rate on diagnostic accuracy Eur. Heart J., March 2, 2008; 29(6): 766 - 776. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach, U. Ropers, A. Kuettner, K. Anders, T. Pflederer, S. Komatsu, W. Bautz, W. G. Daniel, and D. Ropers Randomized comparison of 64-slice single- and dual-source computed tomography coronary angiography for the detection of coronary artery disease. J. Am. Coll. Cardiol. Img., March 1, 2008; 1(2): 177 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. T. Johnson, J. Eng, H. K. Pannu, and E. K. Fishman 64-MDCT Angiography of the Coronary Arteries: Nationwide Survey of Patient Preparation Practice Am. J. Roentgenol., March 1, 2008; 190(3): 743 - 747. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Earls, E. L. Berman, B. A. Urban, C. A. Curry, J. L. Lane, R. S. Jennings, C. C. McCulloch, J. Hsieh, and J. H. Londt Prospectively Gated Transverse Coronary CT Angiography versus Retrospectively Gated Helical Technique: Improved Image Quality and Reduced Radiation Dose Radiology, March 1, 2008; 246(3): 742 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schroeder, S. Achenbach, F. Bengel, C. Burgstahler, F. Cademartiri, P. de Feyter, R. George, P. Kaufmann, A. F. Kopp, J. Knuuti, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: Report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology Eur. Heart J., February 2, 2008; 29(4): 531 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, I. Valenta, O. Gaemperli, O. Adda, V. Treyer, C. A. Wyss, P. Veit-Haibach, F. Tatsugami, G. K. von Schulthess, and P. A. Kaufmann Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating Eur. Heart J., January 2, 2008; 29(2): 191 - 197. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Oncel, G. Oncel, and A. Tastan Effectiveness of Dual-Source CT Coronary Angiography for the Evaluation of Coronary Artery Disease in Patients with Atrial Fibrillation: Initial Experience Radiology, December 1, 2007; 245(3): 703 - 711. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Seifarth, S. Wienbeck, M. Pusken, K.-U. Juergens, D. Maintz, C. Vahlhaus, W. Heindel, and R. Fischbach Optimal Systolic and Diastolic Reconstruction Windows for Coronary CT Angiography Using Dual-Source CT Am. J. Roentgenol., December 1, 2007; 189(6): 1317 - 1323. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, S. Leschka, L. Desbiolles, T. Schepis, O. Gaemperli, B. Seifert, P. Cattin, T. Frauenfelder, T. G. Flohr, B. Marincek, et al. Coronary Artery Motion and Cardiac Phases: Dependency on Heart Rate Implications for CT Image Reconstruction Radiology, November 1, 2007; 245(2): 567 - 576. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Matt, H. Scheffel, S. Leschka, T. G. Flohr, B. Marincek, P. A. Kaufmann, and H. Alkadhi Dual-Source CT Coronary Angiography: Image Quality, Mean Heart Rate, and Heart Rate Variability Am. J. Roentgenol., September 1, 2007; 189(3): 567 - 573. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. M. Kroft, A. de Roos, and J. Geleijns Artifacts in ECG-Synchronized MDCT Coronary Angiography Am. J. Roentgenol., September 1, 2007; 189(3): 581 - 591. [Abstract] [Full Text] [PDF] |
||||