Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Published online before print November 7, 2006, 10.1148/radiol.2421051350
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
2421051350v1
242/1/208    most recent
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by van der Molen, A. J.
Right arrow Articles by Geleijns, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van der Molen, A. J.
Right arrow Articles by Geleijns, J.
(Radiology 2007;242:208-216.)
© RSNA, 2006


Medical Physics

Overranging in Multisection CT: Quantification and Relative Contribution to Dose—Comparison of Four 16-Section CT Scanners1

Aart J. van der Molen, MD and Jacob Geleijns, PhD

1 From the Department of Radiology C-2S, Leiden University Medical Center, Albinusdreef 2, NL-2333 ZA Leiden, the Netherlands. Received August 12, 2005; revision requested October 18; revision received January 4, 2006; accepted February 1; final version accepted April 25. Address correspondence to A.J.v.d.M. (e-mail: molen{at}lumc.nl).

Purpose: To quantify the number of overrange rotations and to assess their relative contribution to organ and effective doses at 16-section body computed tomography (CT).

Materials and Methods: Overranging was quantified for four 16-section scanners by means of free-in-air dose measurements at different scan lengths. Overrange rotations and lengths at a certain section width were derived for all collimations and clinically used pitches by extrapolation. The effect of reconstructed section width on overranging was analyzed separately. Results were applied to clinical protocols for the chest and abdomen. Thyroid and testicular dose and effective dose were established, and relative dose contributions from overranging were calculated. Statistical analysis was performed by using Pearson correlation and paired t tests. P < .05 indicated a significant difference.

Results: The number of overrange rotations showed considerable differences between scanners, with a range of 1.99–4.04 at the lowest and 0.93–2.59 at the highest pitch. Number of rotations correlated negatively with pitch, while overrange length correlated positively with collimation and pitch. The effect of section width was variable. In the protocols, overrange length ranged from 3.2 to 5.8 cm for chest and from 3.2 to 5.2 cm for abdominal CT. When the contribution of overranging was not taken into account, significantly lower values for thyroid (P = .012) and testicular (P = .025) doses and effective doses for chest (P = .005) and abdominal (P = .011) CT resulted.

Conclusion: Overranging is reconstruction-algorithm specific, and its length generally increases with collimation and pitch, while the effect of section width is variable. Overranging may lead to substantial but unnoticed exposure to radiosensitive organs.

© RSNA, 2006