Gastrointestinal Imaging |
1 From the Department of Diagnostic Radiology, Graduate School of Medical Sciences, Kumamoto University, 1-1-2 Honjyo, Kumamoto 860-8556, Japan. Received January 13, 2006; revision requested March 10; revision received April 25; accepted May 31; final version accepted August 1. Address correspondence to K.A.
Purpose: To evaluate prospectively bolus tracking to time hepatic arterial phase (HAP) imaging of hypervascular hepatocellular carcinomas (HCCs) with a 40-detector computed tomographic (CT) scanner.
Materials and Methods: This study received institutional review board approval; informed consent was obtained. The study included 192 patients (123 men, 69 women; mean age, 67.6 years) with known or suspected HCC who underwent dynamic CT, including HAP scanning; CT depicted 111 hypervascular HCCs in 72 patients. Scanning was performed with a 40-detector CT scanner, and bolus tracking was used to time the start of HAP imaging. Patients were randomly assigned to five protocols; HAP scanning was started at a specified interval after trigger threshold was reached: 9 seconds (protocol A), 12 seconds (protocol B), 15 seconds (protocol C), 18 seconds (protocol D), or 21 seconds (protocol E). Trigger threshold level was set at 100 HU above aortic baseline CT number. Enhancement values in the aorta and the tumor-liver contrast (TLC) were measured. Dunnett multiple comparisons were performed to compare enhancement values among the five protocols.
Results: Mean scanning time for the whole liver was 2.1 seconds. Mean enhancement value of the aorta in protocols A, B, C, D, and E were 284.3 HU ± 54.7, 293.8 HU ± 51.0, 308.7 HU ± 55.9, 291.5 HU ± 42.2, and 235.5 HU ± 51.2, respectively. Aortic enhancement was significantly lower in protocol E than in protocol A (P < .01); there was no significant difference between protocols A and B, A and C, and A and D. Mean TLCs in protocols A, B, C, D, and E were 23.4 HU ± 7.6, 35.5 HU ± 14.0, 36.2 HU ± 6.8, 47.2 HU ± 19.2, and 35.1 HU ± 15.8, respectively. A significant difference was found only between protocols A and D (P < .01).
Conclusion: Peak TLC during the HAP occurred 18 seconds after triggering.
© RSNA, 2007
This article has been cited by other articles:
![]() |
Y. Yanaga, K. Awai, T. Nakaura, T. Namimoto, S. Oda, Y. Funama, and Y. Yamashita Optimal Contrast Dose for Depiction of Hypervascular Hepatocellular Carcinoma at Dynamic CT Using 64-MDCT Am. J. Roentgenol., April 1, 2008; 190(4): 1003 - 1009. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kanematsu, S. Goshima, H. Kondo, K. T. Bae, K. Awai, and Y. Yamashita How Should We Optimize Bolus Tracking with Multidetector CT of the Abdomen? Radiology, February 1, 2008; 246(2): 643 - 645. [Full Text] [PDF] |
||||