1 From the CT Clinical Innovation Center, Department of Radiology, Mayo Clinic College of Medicine, 200 First St SW, Rochester, MN 55905 (C.H.M., A.N.P.); Siemens Medical Solutions, Malvern, Pa (O.S.); Siemens Medical Solutions, Forchheim, Germany (H.B., K.S., R.R., C.S., B.S., B.M.O., T.G.F.); and Department of Diagnostic Radiology, Eberhard-Karls-Universität Tübingen, Tübingen, Germany (T.G.F.). From the 2005 RSNA Annual Meeting. Received July 6, 2006; revision requested August 31; revision received October 9; accepted November 2; final version accepted December 4. Address correspondence to C.H.M. (e-mail: mccollough.cynthia{at}mayo.edu).
Purpose: To prospectively compare the dose performance of a 64-channel multidetector row computed tomographic (CT) scanner and a 64-channel dual-source CT scanner from the same manufacturer.
Materials and Methods: To minimize dose in the cardiac (dual-source) mode, the evaluated dual-source CT system uses a cardiac beam-shaping filter, three-dimensional adaptive noise reduction, heart ratedependent pitch, and electrocardiographically based modulation of the tube current. Weighted CT dose index per 100 mAs was measured for the head, body, and cardiac beam-shaping filters. Kerma-length product was measured in the spiral cardiac mode at four pitch values and three electrocardiographic modulation temporal windows. Noise was measured in an anthropomorphic phantom. Data were compared with data from a 64-channel multidetector row CT scanner.
Results: For the multidetector row and dual-source CT systems, respectively, weighted CT dose index per 100 mAs was 14.2 and 12.2 mGy (head CT), 6.8 and 6.4 mGy (body CT), and 6.8 and 5.3 mGy (cardiac CT). In the spiral cardiac mode (no electrocardiographically based tube current modulation, 0.2 pitch), equivalent noise occurred at volume CT dose index values of 23.7 and 35.0 mGy (coronary artery calcium CT) and 58.9 and 61.2 mGy (coronary CT angiography) for multidetector row CT and dual-source CT, respectively. The use of heart ratedependent pitch values reduced volume CT dose index to 46.2 mGy (0.265 pitch), 34.0 mGy (0.36 pitch), and 26.6 mGy (0.46 pitch) compared with 61.2 mGy for 0.2 pitch. The use of electrocardiographically based tube currentmodulation and temporal windows of 110, 210, and 310 msec further reduced volume CT dose index to 9.125.1 mGy, dependent on the heart rate.
Conclusion: For electrocardiographically gated coronary CT angiography, image noise equivalent to that of multidetector row CT can be achieved with dual-source CT at doses comparable to or up to a factor of two lower than the doses at multidetector row CT, depending on heart rate of the patient.
© RSNA, 2007
This article has been cited by other articles:
![]() |
A. C. Weustink, N. R. Mollet, F. Pugliese, W. B. Meijboom, K. Nieman, M. H. Heijenbrok-Kal, T. G. Flohr, L. A. E. Neefjes, F. Cademartiri, P. J. de Feyter, et al. Optimal Electrocardiographic Pulsing Windows and Heart Rate: Effect on Image Quality and Radiation Exposure at Dual-Source Coronary CT Angiography Radiology, September 1, 2008; 248(3): 792 - 798. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. K. Courtney, N. R. Munce, K. J. Anderson, A. S. Thind, G. Leung, P. E. Radau, F. S. Foster, I. A. Vitkin, R. S. Schwartz, A. J. Dick, et al. Innovations in imaging for chronic total occlusions: a glimpse into the future of angiography's blind-spot Eur. Heart J., March 1, 2008; 29(5): 583 - 593. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Earls, E. L. Berman, B. A. Urban, C. A. Curry, J. L. Lane, R. S. Jennings, C. C. McCulloch, J. Hsieh, and J. H. Londt Prospectively Gated Transverse Coronary CT Angiography versus Retrospectively Gated Helical Technique: Improved Image Quality and Reduced Radiation Dose Radiology, March 1, 2008; 246(3): 742 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Schoenhagen Back to the future: coronary CT angiography using prospective ECG triggering Eur. Heart J., January 2, 2008; 29(2): 153 - 154. [Full Text] [PDF] |
||||
![]() |
H. Seifarth, S. Wienbeck, M. Pusken, K.-U. Juergens, D. Maintz, C. Vahlhaus, W. Heindel, and R. Fischbach Optimal Systolic and Diastolic Reconstruction Windows for Coronary CT Angiography Using Dual-Source CT Am. J. Roentgenol., December 1, 2007; 189(6): 1317 - 1323. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. D. Cody and M. Mahesh AAPM/RSNA Physics Tutorials AAPM/RSNA Physics Tutorial for Residents: Technologic Advances in Multidetector CT with a Focus on Cardiac Imaging RadioGraphics, November 1, 2007; 27(6): 1829 - 1837. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. J. Schoepf, P. Suranyi, C. Thilo, P. Costello, and P. L. Zwerner Replacing diagnostic catheterization with coronary CT angiography: the final frontier Eur. Heart J., October 1, 2007; 28(19): 2305 - 2306. [Full Text] [PDF] |
||||
![]() |
A. J. Einstein, K. W. Moser, R. C. Thompson, M. D. Cerqueira, and M. J. Henzlova Radiation Dose to Patients From Cardiac Diagnostic Imaging Circulation, September 11, 2007; 116(11): 1290 - 1305. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |