Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Published online before print May 16, 2007, 10.1148/radiol.2441060756
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
2441060756v1
244/1/104    most recent
Right arrow Submit a response
Right arrow View responses
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yang, S. K.
Right arrow Articles by Im, J.-G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yang, S. K.
Right arrow Articles by Im, J.-G.
(Radiology 2007;244:104-111.)
© RSNA, 2007


Breast Imaging

Screening Mammography–detected Cancers: Sensitivity of a Computer-aided Detection System Applied to Full-Field Digital Mammograms1

Sang Kyu Yang, MD2, Woo Kyung Moon, MD, Nariya Cho, MD, Jeong Seon Park, MD, Joo Hee Cha, MD, Sun Mi Kim, MD, Seung Ja Kim, MD, and Jung-Gi Im, MD

1 From the Department of Radiology and Clinical Research Institute, Seoul National University Hospital and the Institute of Radiation Medicine, Seoul National University Medical Research Center, 28 Yongon-dong, Chongno-gu, Seoul 100-744, Korea (S.K.Y., W.K.M., N.C., J.S.P., S.J.K., J.G.I.); Department of Radiology, Boramae Municipal Hospital, Seoul, Korea (J.H.C.); and Department of Radiology, Bundang Seoul National University Hospital, Bundang, Korea (S.M.K.). Received April 30, 2006; revision requested June 26; revision received July 20; accepted August 23; final version accepted November 1. Supported by KISTEP, Ministry of Science and Technology, Korea. Address correspondence to W.K.M. (e-mail: moonwk{at}radcom.snu.ac.kr).

Purpose: To retrospectively evaluate the sensitivity of the performance of a computer-aided detection (CAD) system applied to full-field digital mammograms for detection of breast cancers in a screening group, with histologic findings as the reference standard.

Materials and Methods: This study had institutional review board approval, and patient informed consent was waived. A commercially available CAD system was applied to the digital mammograms of 103 women (mean age, 51 years; range, 35–69 years) with 103 breast cancers detected with screening. Sensitivity values of the CAD system according to mammographic appearance, breast composition, and histologic findings were analyzed. Normal mammograms from 100 women (mean age, 54 years; age range, 35–75 years) with no mammographic and clinical abnormality during 2-year follow-up were used to determine false-positive CAD system marks. Differences between the cancer detection rates in fatty and dense breasts for the CAD system were compared by using the {chi}2 test.

Results: The CAD system correctly marked 99 (96.1%) of 103 breast cancers. The CAD system marked all 44 breast cancers that manifested as microcalcifications only, all 23 breast cancers that manifested as a mass with microcalcifications, and 32 (89%) of 36 lesions that appeared as a mass only. The sensitivity of the CAD system in the fatty breast group was 95% (59 of 62) and in the dense breast group was 98% (40 of 41) (P = .537). The CAD system correctly marked all 31 lesions of ductal carcinoma in situ (DCIS), all 22 lesions of invasive ductal carcinoma with DCIS, the single invasive lobular carcinoma lesion, and 45 (92%) of 49 lesions of invasive ductal carcinoma. On normal mammograms, the mean number of false-positive marks per patient was 1.80 (range, 0–10 marks; median, 1 mark).

Conclusion: The CAD system can correctly mark most (96.1%) asymptomatic breast cancers detected with digital mammographic screening, with acceptable false-positive marks (1.80 per patient).

© RSNA, 2007