Experimental Studies |
1 From the Department of Radiology, Duke University Medical Center, Box 3808, Durham, NC 27710 (L.M.H., R.E.R., T.T.Y., P.C.G., C.L.); and Radiation Safety Division, Duke University Health System, Durham, NC (R.E.R., T.T.Y., G.T., G.N.). Received December 1, 2006; revision requested January 26, 2007; revision received February 8; accepted March 22; final version accepted May 24. Supported in part by GE Healthcare, Milwaukee, Wis. Address correspondence to L.M.H. (e-mail: hurwi001{at}mc.duke.edu).
Purpose: To measure prospectively and directly both organ dose and effective dose (ED) for adult cardiac and pulmonary computed tomographic (CT) angiography by using current clinical protocols for 64-detector CT in an anthropomorphic female phantom and to estimate lifetime attributable risk of breast and lung cancer incidence on the basis of measured ED and organ dose.
Materials and Methods: Cardiac and pulmonary 64-detector CT angiography was performed by using current clinical protocols to evaluate the pulmonary veins (electrocardiographically [ECG] gated, 64 sections at 0.625-mm collimation, 120 kVp, 300 mA, 0.35-second tube rotation), native coronary arteries (ECG gated; 64 sections at 0.625 mm; 120 kVp; maximum current, 500–750 mA; minimum, 100–350 mA; 0.35-second tube rotation) and pulmonary embolus (64 sections at 1.25 mm, 140 kVp, 645 mA, 0.5-second tube rotation). Absorbed organ doses were measured by using an anthropomorphic female phantom and metal oxide semiconductor field effect transistor detectors. ED was calculated from measured organ doses and the dose-length product.
Results: ED for current adult cardiac and pulmonary 64-detector CT angiography protocols were 12.4–31.8 mSv. Overall, skin, breast, and esophagus and heart had the highest recorded absorbed organ doses. Relative risk for breast cancer incidence for girls and women was 1.004–1.042 for a single examination. Relative risk for lung cancer incidence for men and women was 1.005–1.076 from a single examination.
Conclusion: EDs and organ doses from 64-detector CT are higher than those previously reported for adult cardiac and pulmonary CT angiography protocols. Risk for breast and lung cancer induction from these studies is greatest for the younger patient population.
© RSNA, 2007
This article has been cited by other articles:
![]() |
S. McDermott, S. C. Barry, E. E. Judge, S. Collins, P. A. de Jong, H. A. W. M. Tiddens, E. F. McKone, C. C. Gallagher, and J. D. Dodd Tracheomalacia in Adults with Cystic Fibrosis: Determination of Prevalence and Severity with Dynamic Cine CT Radiology, June 9, 2009; (2009) 2522081956. [Abstract] [Full Text] |
||||
![]() |
W. P. Shuman, K. R. Branch, J. M. May, L. M. Mitsumori, J. N. Strote, B. H. Warren, T. J. Dubinsky, D. W. Lockhart, and J. H. Caldwell Whole-Chest 64-MDCT of Emergency Department Patients with Nonspecific Chest Pain: Radiation Dose and Coronary Artery Image Quality with Prospective ECG Triggering Versus Retrospective ECG Gating Am. J. Roentgenol., June 1, 2009; 192(6): 1662 - 1667. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. K. Pahade, D. Litmanovich, I. Pedrosa, J. Romero, A. A. Bankier, and P. M. Boiselle Quality Initiatives: Imaging Pregnant Patients with Suspected Pulmonary Embolism: What the Radiologist Needs to Know RadioGraphics, May 1, 2009; 29(3): 639 - 654. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Mayo and J. A. Leipsic Radiation Dose in Cardiac CT Am. J. Roentgenol., March 1, 2009; 192(3): 646 - 653. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. R. Goodman, H. D. Sostman, P. D. Stein, and P. K. Woodard CT Venography: A Necessary Adjunct to CT Pulmonary Angiography or a Waste of Time, Money, and Radiation? Radiology, February 1, 2009; 250(2): 327 - 330. [Full Text] [PDF] |
||||
![]() |
P. C. Shrimpton, B. F. Wall, T. T. Yoshizumi, L. M. Hurwitz, and P. C. Goodman Effective Dose and Dose-Length Product in CT Radiology, February 1, 2009; 250(2): 604 - 605. [Full Text] [PDF] |
||||
![]() |
L. M. Hurwitz, T. T. Yoshizumi, P. C. Goodman, R. C. Nelson, G. Toncheva, G. B. Nguyen, C. Lowry, and C. Anderson-Evans Radiation Dose Savings for Adult Pulmonary Embolus 64-MDCT Using Bismuth Breast Shields, Lower Peak Kilovoltage, and Automatic Tube Current Modulation Am. J. Roentgenol., January 1, 2009; 192(1): 244 - 253. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. Mettler Jr, W. Huda, T. T. Yoshizumi, and M. Mahesh Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog Radiology, July 1, 2008; 248(1): 254 - 263. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Huda, T. T. Yoshizumi, L. M. Hurwitz, and P. C. Goodman Computing Effective Doses from Dose-Length Product in CT Radiology, July 1, 2008; 248(1): 321 - 322. [Full Text] [PDF] |
||||
![]() |
M. D. Siegel Evaluation of Suspected Pulmonary Embolism During Pregnancy JAMA, April 9, 2008; 299(14): 1665 - 1665. [Full Text] [PDF] |
||||