Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Published online before print July 15, 2008, 10.1148/radiol.2482071800
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
2482071800v1
248/3/945    most recent
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Vano, E.
Right arrow Articles by Haskal, Z. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Vano, E.
Right arrow Articles by Haskal, Z. J.
(Radiology 2008;248:945-953.)
© RSNA, 2008


Medical Physics

Eye Lens Exposure to Radiation in Interventional Suites: Caution Is Warranted1

Eliseo Vano, PhD, Luciano Gonzalez, PhD, Jose M. Fernández, MSc, and Ziv J. Haskal, MD

1 From the Department of Radiology, Complutense University, 28040 Madrid, Spain (E.V., L.G., J.M.F.); Medical Physics Service, San Carlos University Hospital, Madrid, Spain (E.V., J.M.F.); and Department of Vascular and Interventional Radiology, New York-Presbyterian Hospital/Columbia University Medical Center, New York, NY (Z.J.H.). Received October 14, 2007; revision requested November 29; revision received January 18, 2008; accepted February 28; final version accepted March 5. Address correspondence to E.V. (e-mail: eliseo{at}med.ucm.es).

Purpose: To report estimated radiation doses to the eye lens of the interventionalist from procedures performed with and without use of radiation protection measures.

Materials and Methods: Scattered radiation doses for seven interventional radiology fluoroscopic systems were measured by using phantoms simulating patients 16–28 cm in thickness undergoing low-, medium-, and high-mode fluoroscopy, cine cardiac imaging, and digital subtraction angiography (DSA). The radiation doses to the eye lens in low- and high-dose scenarios were estimated. Beam angulation, biplanar equipment, working distance, procedure complexity, imaging collimation, and use of eyeglasses and/or protective suspended screens were taken into account. The doses to the lens in several procedures were assessed.

Results: Mean scattered radiation doses to the lens during fluoroscopy were 6.0 and 34.5 µSv/min in the low- and high-dose scenarios, respectively. For DSA, typical doses to the lens ranged from 0.77 to 3.33 µSv per image. Operation modes involving increasing or decreasing radiation doses were quantified. For hepatic chemoembolization, iliac angioplasty, pelvic embolization, and transjugular intrahepatic portosystemic shunt creation, lens doses ranged from 0.25 to 3.72 mSv per procedure when protection was not used. Lens doses in the neuroembolization procedures could exceed 10 mSv per procedure.

Conclusion: With typical reported workloads, radiation doses to eye lenses may exceed the threshold for deterministic effects (ie, lens opacities or cataracts) after several years of work if radiation protection tools are not used.