Published online before print July 25, 2006, 10.1148/radiol.2403050519
JPEG2000 Compression of Thin-Section CT Images of the Lung: Effect of Compression Ratio on Image Quality1
Helmut Ringl, MD,
Ruediger E. Schernthaner, MD,
Alexander A. Bankier, MD,
Michael Weber, PhD,
Mathias Prokop, MD,
Christian J. Herold, MD and
Cornelia Schaefer-Prokop, MD
1 From the Department of Radiology, University Hospital Vienna, AKH Wien, Währinger Gürtel 18-20, A-1090, Vienna, Austria (H.R., R.E.S., A.A.B., M.W., C.J.H., C.S.); and Department of Radiology, Utrecht Medical Center, Utrecht, the Netherlands (M.P.). Received March 29, 2005; revision requested May 25; revision received September 2; accepted September 12; final version accepted November 23.
Address correspondence to H.R. (e-mail: helmut.ringl{at}meduniwien.ac.at).

View larger version (13K):
[in a new window]
|
Figure 1a: (a, b) Error bar charts. The center mark indicates the relative observed percentage of correct determinations of the original image in the forced-choice two-alternative setup. A value of 0.5 indicates no differentiation between the compressed and noncompressed images, and a value of 1.0 indicates that all original images could be differentiated from the compressed image copies. Error bars indicate 95% confidence intervals. (a) Single-detector CT data. (b) Multidetector CT data.
|
|

View larger version (12K):
[in a new window]
|
Figure 1b: (a, b) Error bar charts. The center mark indicates the relative observed percentage of correct determinations of the original image in the forced-choice two-alternative setup. A value of 0.5 indicates no differentiation between the compressed and noncompressed images, and a value of 1.0 indicates that all original images could be differentiated from the compressed image copies. Error bars indicate 95% confidence intervals. (a) Single-detector CT data. (b) Multidetector CT data.
|
|

View larger version (134K):
[in a new window]
|
Figure 2: Four image sections (four rows) from different patients. Sections in the two top rows were derived at single-detector CT, and sections in the two bottom rows were derived at multidetector CT. The compression ratio increases from left to right images. The original image is the first image in each row. The high-contrast areas (rows 1 and 3) are almost unchanged, whereas the low-contrast areas (rows 2 and 4) are susceptible to blurring artifacts. If viewed electronically, a monitor with high brightness and contrast settings should be used to visualize the artifacts clearly.
|
|
Copyright © 2006 by the Radiological Society of North America.