Experimental Studies |
1 From the Department of Radiology, University of Cologne, Joseph-Stelzmann-Str 9, D-50931 Cologne, Germany (F.S.); the Departments of Neuroradiology (F.S., M.W., A.B., M.S.), Neurosurgery (H.G.B.), and Pathology (E.W.), University of Freiburg, Germany; and the Department of Radiology, University of Bonn, Germany (A.B.). Received January 7, 1998; revision requested April 2; revision received August 11; accepted October 14. Address reprint requests to F.S.
| Abstract |
|---|
|
|
|---|
MATERIALS AND METHODS: Nineteen bilateral carotid arterytoexternal jugular vein AVFs were created in 10 adult dogs. Fourteen polyethylene terephthalatecoated and five polyurethane-coated self-expandable nitinol stents were placed with a transfemoral approach. Angiography was performed immediately after placement, after 1 and 3 weeks, and at months 3, 6, 9, and 12. All grafts were examined histopathologically.
RESULTS: The AVF was successfully occluded with all but one of the polyethylene terephthalatecoated stents. Graft patency was seen with 12 polyethylene terephthalatecoated stents, while acute vessel occlusion was evident with two polyethylene terephthalatecoated stents following percutaneous transluminal angioplasty immediately after implantation. In the group with polyurethane-coated stents, three graft dislocations and two delayed vessel occlusions were observed. In both groups, histologic examination disclosed a mild foreign body reaction with a few macrophages. No inflammatory reactions were seen. An unorganized fibrin layer was found adjacent to the polyurethane coating.
CONCLUSION: Polyethylene terephthalatecoated nitinol stents showed good biocompatibility and a high rate of occlusion of experimental AVFs. The mechanical properties of the polyethylene terephthalate coating were similar to those of the nitinol stents, which facilitated handling. Polyurethane-coated stents showed frequent dislocation.
Index terms: Animals, 276.245 Arteriovenous malformations, 276.245 Carotid arteries, interventional procedure, 172.1269 Fistula, arteriovenous, 276.245 Stents and prostheses, 276.245
| Introduction |
|---|
|
|
|---|
Polyethylene terephthalatecoated stents successfully obliterate human abdominal aortic aneurysms (3). However, perivascular and systemic inflammatory reactions have been reported (8,9), which seem to be responsible for poor short-term patency rates in experimental studies (9). Not yet clear is whether these reactions are directed against the specific polyester itself or manufacturing-related components (10,11). Polyurethane proved to be successful in the preparation of biomaterials with low bacterial adherence and improved biocompatibility (12). Contrary results were reported after placement of polyurethane-coated stents in porcine coronary arteries (13,14).
The current study was designed to investigate the use of self-expanding nitinol stents with a tubular coating of polyethylene terephthalate, in comparison with a coating of medical-grade polyurethane, for the treatment of high-flow experimental AVFs. The canine model was chosen because experiments with different porous and impervious vascular prostheses have shown that the rate of healing of implanted prostheses in dogs approximates that in humans, which is much slower than that in other animals, such as the pig, calf, and baboon (15).
| MATERIALS AND METHODS |
|---|
|
|
|---|
Fourteen self-expanding nitinol stents (heat-treated nickel-titanium alloy, 12 loops per circumference, 62.4 pores per square centimeter, 5-mm inner diameter, 40-mm length; Strecker stent, Boston Scientific, Watertown, Mass) were coated with a woven tube of polyethylene terephthalate (35-mm length, 5-mm inner diameter, 0.4-mm thickness [0.08-mm thickness under compression, by our own measurements]; Sulzer Orthopedics, Winterthur, Switzerland) (Fig 1a). The polyethylene terephthalate was thoroughly washed to remove manufacturing-related components. Because of a patent pending at the time of the investigation, the detailed technique of the washing procedure was unknown to the investigators.
|
|
All coated stents were mounted onto a 5-F angioplasty catheter (5-mm ballon diameter, 40-mm ballon length; Schneider, Lausanne, Switzerland) by a simple crochet technique with a thin nylon line (0.18-mm diameter, 3.0-kg strength) (19). Angioplasty catheters were chosen because of the soft surface of the deflated balloon, which kept the mounted stent in place. Sterilization was achieved under an ethylene oxide atmosphere at a maximum temperature of 45°C and an atmospheric humidity of 50%70%.
Endovascular treatment was performed 2129 days (mean, 23 days) following surgery. A 10-F Teflon introducer sheath (Cook, Bloomington, Ind) was placed in the right femoral artery, and a pressurized saline infusion was connected to the side port of the sheath. Before stent placement, diagnostic angiography (4-F catheter; Cook) was performed, and heparin (100 IU per kilogram of body weight) was administered. The angioplasty catheter was placed by using a 0.038-inch guide wire (260-cm length; Amplatz Super Stiff EX; Boston Scientific). A 60-mm piece of an 8-F introducer sheath served as a cover to avoid damage to the coated stent during the passage through the hemostatic valve. All angiographic procedures were performed with digital subtraction angiography (DAS, Picker International, Cleveland, Ohio; or Digitron, Siemens, Erlangen, Germany).
A road-map technique was used for precise positioning of the stents. After gradual retraction of the nylon line, the coated stents expanded, starting from their proximal end, to the luminal diameter of the parent vessel. Immediately after stent placement, percutaneous transluminal angioplasty (5-F angioplasty catheter; Schneider) of two of the aneurysms was performed with polyethylene terephthalatecoated stents because of hemodynamically significant vessel stenoses caused by incomplete expansion of the coated stents. For 3 weeks after implantation, the animals received 80 mg of aspirin (Bayer, Leverkusen, Germany) per day orally (20).
Control angiograms (5-F introducer sheath, 4-F diagnostic catheter; Cook) were obtained percutaneously at 4 frames per second immediately after stent placement, after 1 and 3 weeks, and at 3, 6, 9, and 12 months (Table). The arterial lumen before stent placement was assumed to be 100%; at each follow-up angiographic examination, the maximum percentage of stenosis of the vessel segment with the stent was measured with angiograms in two different projections (posteroanterior, lateral) by two independent investigators (F.S., M.W.). Occlusion of the arterial vessel was judged as acute (occlusion within 1 week), subacute (occlusion between weeks 1 and 3), or delayed (occlusion after week 3). Because the internal carotid artery is of minor importance to the cerebral blood supply in dogs, no angiography of the brain was performed to exclude thromboembolism.
|
This protocol was approved by the animal experimentation committee of the University of Freiburg (Germany) and was carried out according to the Regulations of Animal Protection.
| RESULTS |
|---|
|
|
|---|
Angiographic Results
All except one of the polyethylene terephthalatecoated stents were properly placed, covering the orifice of the AVF with the coated stent segment. The angiographic size of the AVF had no influence on the occlusion rate following stent placement.
Immediately after placement of polyethylene terephthalatecoated stents, contrast agent ran directly through the threads of the coating material in six of the 14 polyethylene terephthalatecoated stents; the leakage stopped in two stents within 1 hour. During further follow-up, a gradual diminution of these leakages was observed, with an interruption in one stent each after 1 week and 3 months, and in two stents after 6 months (Fig 2). Persistence of the AVF was evident with one stent because of incomplete covering of the orifice of the AVF. No recanalization of an occluded AVF was observed (Fig 3).
|
|
After placement, polyethylene terephthalatecoated stents showed a mean graft stenosis of 45% (n = 14), which was followed by a mean graft stenosis of 30% after 1 week (n = 12). A mean graft stenosis of 39% was seen after 3 weeks (n = 12) and after 3 months (n = 12). Further follow-up disclosed a mean graft stenosis of 30% after 6 months (n = 8), 31% stenosis after 9 months (n = 6), and 40% stenosis after 12 months (n = 4). No migration of a polyethylene terephthalatecoated stent was observed.
Proper stent placement and occlusion of the AVF were achieved in all five polyurethane-coated stents, which extended at least 8 mm distal to the orifice of the AVF. After 1 week, only two polyurethane-coated stents remained in place covering the AVFs. The distal portion of one polyurethane-coated stent was dislocated into the external jugular vein, with reestablishment of the AVF. Displacement of two stents was seen. These stents were not discovered in the carotid arteries or in the portion of the external jugular vein proximal to the AVF. At fluoroscopy, however, the displaced stents were faintly visualized over the heart. Polyurethane-coated stents that remained in the common carotid artery (n = 3) showed graft patency with one stent and delayed vessel occlusion after 3 months (n = 2). After placement of polyurethane-coated stents, mean graft stenosis was 35%, with a gradual increase to 83% during further follow-up of up to 9 months (n = 1).
Histopathologic Results
Histopathologic examination confirmed the angiographic stenoses in both groups but showed irregular intimal fibrosis that was not apparent at angiography. Uncoated nitinol filaments showed complete endothelialization with a smooth neointimal layer with a mean thickness of 153.9 µm ± 78.4 (range, 25.3381.3 µm); this layer merged to the original arterial intima. Compression and minor atrophy of the medial layer were evident subjacent to the nitinol filaments. The adventitia and the perivascular tissue were inconspicuous. No signs of acute inflammation or marked foreign body reactions were evident.
Within the polyethylene terephthalate coating, a few macrophages, giant cell formations, and newly sprouted capillaries were seen. Subjacent to the polyethylene terephthalate coating, a similar scar tissue with few cellular components and neocapillaries was found, representing the former arterial intimal layer. Coated stent segments showed complete endothelialization with a neointimal layer with a mean thickness of 167.6 µm ± 96.8 (range, 20.8452.6 µm) (Fig 4). The mean thickness of the polyethylene terephthalate coating, including the stent filaments, was 478.4 µm ± 122.7 (range, 290.6580.2 µm).
|
|
Between the polyurethane coating, which was wrinkled, and the medial layer, a scar tissue with very few macrophages and neocapillaries was seen. The nitinol filaments were embedded in an unorganized fibrinoid layer, which was covered by a thin, fibrosed neointimal layer. This complex measured 369.4 µm ± 337.1 (data obtained from one nonoccluded polyurethane-coated stent) (Fig 5). In the stent with graft occlusion, partly organized thrombus was observed without evidence of a neointimal layer. Toward the silicone glue, no acute inflammatory reactions were discovered.
|
|
| DISCUSSION |
|---|
|
|
|---|
Since the first successful occlusion of abdominal aortic aneurysms with transluminally placed endovascular prosthetic grafts, polyethylene terephthalate has been commonly used in interventional radiology (3,4). Recently, perivascular and systemic inflammatory reactions were reported following implantation of stents coated with polyethylene terephthalate (Dacron) (8,9). Link et al (9) reported granulocyte infiltration next to the polyethylene terephthalate (Dacron) and acute graft occlusion in six of seven stents in swine. Blum et al (8) observed an elevation of the C-reactive protein and leukocytosis in humans following endovascular treatment of abdominal aortic aneurysm with transluminally placed endovascular prosthetic grafts coated with polyethylene terephthalate (Dacron). No evidence of bacteremia or graft infection was seen.
Whether these reactions are directed toward the specific polyester itself or toward manufacturing-related components has been discussed (10,11). The industrial manufacture of polyethylene terephthalate involves numerous additives, such as salts of heavy metals (antimony, manganese, calcium), which accelerate reactions or minimize undesirable side reactions. Nitrogen-containing aromatic organic compounds and titanium dioxide, for example, affect the properties of the polymer (11). The polyethylene terephthalate used in our study was thoroughly washed after manufacture. The detailed technique of the washing procedure was unknown to the investigators. According to the manufacturer (Sulzer Orthopedics), this technique markedly reduces the extractable substances in the polyethylene terephthalate textile, compared with the use of customary Dacron.
Histopathologic examinations, which were performed after 3, 6, 9, or 12 months, revealed mild foreign body reactions, without acute inflammation. Although no laboratory examinations were performed, we did not observe clinical symptoms of inflammation such as fever, malaise, or weight loss. This, and the absence of frequent acute vessel occlusions as reported by Link et al (9), might be further hints of the importance of manufacturing-related components.
The textile texture of the polyethylene terephthalate coating used has elastic properties similar to those of the nitinol stent. With the exception of 23 mm left uncoated on both ends for better graft fixation, the nitinol stent was completely coated, even in a stretched profile, when mounted onto an angioplasty catheter. This facilitated proper placement of the self-expandable stent.
Polyethylene terephthalatecoated stents showed a mean graft stenosis that was twice as high as that of bare nitinol stents (mean, 15% graft stenosis) (2,17), which is mainly attributable to the polyethylene terephthalate coating (mean thickness in histologic specimens, 478.4 µm). No marked difference was seen between the thickness of the neointimal layer with coated and uncoated nitinol stents (167 µm vs 150 µm).
During the angiographic follow-up, the vessel stenosis of the arterial segments with stents decreased. Theoretically, the diameter of the self-expanding nitinol stents (5 mm) should have caused an overdilation of the common carotid artery of approximately 16%. It seems likely that the expansile forces of the self-expanding stents are responsible for decreased vessel stenoses in nonarteriosclerotic arteries.
Two stents showed a hemodynamically significant vessel stenosis following placement. These stenoses were due to incomplete expansion of the proximal end of both self-expanding stents for unknown reasons.
The thickness of the polyethylene terephthalate coating (0.4 mm) was markedly diminished under radial force (0.08 mm). This, however, reduces the density of the textile structure and may cause persistent leakages, as seen with six of the 14 polyethylene terephthalatecoated stents. Hemodynamic changes similar to those that cause occlusion of side-wall aneurysms with the use of plain stents (channeling effect) (2) and the deposition of fibrin and platelets (11) are likely to be responsible for early interruption of leakages up to week 1. During further follow-up, invasion of histiocytes and the progression of the neointimal cover may close remaining leakages (18). In dogs, healing of the inner vascular wall requires 812 weeks (15). This corresponds to the observation of interrupted leakages after 3 months in two stents.
One persistent AVF following misplacement of a polyethylene terephthalatecoated stent was not affected by such phenomena. Histopathologic examinations revealed discontinuities of the polyethylene terephthalate textile structure up to 3.7 mm in four specimens. These discontinuities, which were covered by a neointimal layer, might correspond to damages caused by retraction of the nylon line during stent delivery. The polyethylene terephthalate coating was not sutured to the stents. Only expansile forces of the nitinol stent kept the coating in place between the arterial wall and the nitinol filaments. In large-orifice AVFs, this causes the risk of bulking of the coating toward the external jugular vein, as seen in one stent with a thrombosed pseudoaneurysm. This was not apparent at angiography.
A similar mechanism led to acute thrombosis with two stents following percutaneous transluminal angioplasty. The elastic recoil of the nitinol stents was not accompanied by the polyethylene terephthalate coating, which subsequently protruded into the lumen of the arterial vessel. However, the angiographic size of the experimental AVF did not influence the occlusion rate following placement of coated stents.
Polyurethane, a biostable and biocompatible polymer, has been used extensively for vascular applications and, to our knowledge, is the most widely used polymer for construction of permanent pacemaker leads. Polyurethane is also known for its low bacterial adherence (12). However, in porcine coronary arteries, contrary results have been reported. De Scheerder et al (13) described a neointimal thickening 6 weeks after placement of polyurethane-coated stainless steel stents, which was comparable to the neointimal thickening induced by bare metallic stents. Holmes et al (14) used polyurethane-embedded balloon-expandable tantalum stents. In contrast to De Scheerder et al (13), who did not observe a marked inflammatory reaction, Holmes et al (14) found a foreign body inflammatory response with multinucleate giant cells and a high rate of thromboembolic occlusions within the first 48 hours. Neither group of authors reported migration of the stents.
In our study, three of five polyurethane-coated stents dislocated within 1 week. Two stents that showed displacement were not discovered in the carotid arteries or in the external jugular vein but were faintly visualized over the heart. Despite complete covering of the AVF with the polyurethane coating, dislocation of the distal portion of the stent into the external jugular vein was evident with a third stent.
The design of polyurethane-coated stents left uncoated stent segments of 23 mm on both ends of the nitinol stents for better graft fixation. Immediately after placement, the polyurethane-coated stents straddled the lesion and extended at least 8 mm distal to the orifice of the AVF. Comparable to the polyethylene terephthalate coating, the polyurethane film was not sutured or glued to the nitinol filament. It is possible that during stent placement, the polyurethane coating slipped, and no uncoated stent segments remained for proper distal anchorage. Because only segments of the explanted stents underwent histopathologic examination, this theory could not be proved.
The impermeability of the polyurethane film prevented the invasion of histiocytes into the coating. Instead, an almost noncellular scar tissue and an unorganized fibrin layer were found around the polyurethane coating. On theoretic grounds, it seems possible that the smooth and impermeable polyurethane film, which, in contrast to the polyethylene terephthalate coating, was not invaded by histiocytes, glided on the subjacent fibrinoid layer. However, the phenomenon of the frequent migration of polyurethane-coated stents remains unclear.
Histopathologic examination revealed lots of wrinkles of the polyurethane coating, which might be caused by retraction of the nylon line during stent delivery. These wrinkles and the thickness of the "neointimal complex" of the unorganized fibrinoid layer and the fibrosed neointimal layer (mean, 369 µm) probably caused delayed vessel occlusion with two stents. Because of the impermeability of the polyurethane film, the intraluminal thrombus was only partly organized. In addition, the elastic mismatch of the nonelastic polyurethane coating and the self-expandable nitinol stent made proper stent placement more difficult compared with the placement of the polyethylene terephthalatecoated stent. Further attempts at using this particular polyurethane coating were stopped.
Practical application: Both polyethylene terephthalate and polyurethane showed good biocompatibility. The impermeability of polyurethane caused less favorable angiographic and histologic results. From our point of view, impermeable coating materials are not suitable for endovascular applications. The polyethylene terephthalate was easy to handle and was used successfully in the occlusion of high-flow experimental AVFs. The use of polyethylene terephthalate for the treatment of AVFs or fusiform aneurysms seems possible. However, the textile structure should be changed to achieve a higher textile density, combined with a reduced thickness of the coating. Integration of the nitinol filaments into the textile would be desirable.
| Acknowledgments |
|---|
| Footnotes |
|---|
Author contributions: Guarantors of integrity of entire study, F.S., M.W., M.S.; study concepts and design, F.S., M.W., M.S.; definition of intellectual content, F.S., M.S.; literature research, F.S., M.W.; experimental studies, F.S., M.W., A.B., H.G.B.; data acquisition, F.S., M.W., A.B., E.W.; data analysis, F.S., M.W., A.B., E.W., M.S.; manuscript preparation, F.S., M.W., M.S.; manuscript editing, F.S.; manuscript review, M.W., A.B., E.W., M.S.