Cardiac Imaging |
1 From the Departments of Clinical Physiology (H.A., O.P.), Radiology (C.H., S.L., F.S.), Cardiology (U.T.), and Pediatric Medicine (K.H., G.B.), Lund University Hospital, S-221 85 Lund, Sweden. Received June 2, 1998; revision requested July 22; revision received September 18; accepted November 19. Supported in part by the Swedish Medical Research Council, Stockholm, Sweden; Hellmuth Herz Foundation, Lund, Sweden; Swedish Royal Physiographic Society, Lund, Sweden; and Swedish Heart Lung Foundation, Stockholm, Sweden. Address reprint requests to H.A.
| Abstract |
|---|
|
|
|---|
MATERIALS AND METHODS: The accuracy and precision of MR velocity mapping were studied in 12 control subjects (six men, six women) and in a phantom. MR velocity mapping and radionuclide angiography were performed on the same day in 24 patients (16 adults, two adolescents, six children; five male patients, 19 female patients).
RESULTS: The mean error in QP/QS at MR velocity mapping in phantom experiments was -1% ± 1 (mean ± SD). In control subjects, QP/QS at MR velocity mapping was 1.03 ± 0.03, and the cardiac index was 3.1 L/min/m2 ± 0.2 and 3.2 L/min/m2 ± 0.3 for women and men, respectively. In patients, QP/QS at radionuclide angiography was 14% ± 13, higher than at MR velocity mapping. Interobserver variability was four times higher for radionuclide angiography compared with MR velocity mapping, 0% ± 16 versus 0% ± 4 (n = 12). The difference between repeated MR flow measurements in the same vessel was -1% ± 5 (n = 36).
CONCLUSION: The data suggest that MR velocity mapping is accurate and precise for measurements of shunt size over the whole range of possible QP/QS values.
Index terms: Atrial septal defect, 514.141 Heart, flow dynamics, 51.12144 Heart, MR, 51.121411, 51.121412, 51.12144 Heart, radionuclide studies, 51.12175, 51.12176 Magnetic resonance (MR), vascular studies, 51.121411, 51.121412, 51.12144 Pulmonary veins, 565.158 Ventricular septal defect, 515.142
| Introduction |
|---|
|
|
|---|
By using magnetic resonance (MR) velocity mapping, blood flow can be measured noninvasively with a high degree of accuracy (35), and consequently QP/QS can be determined from measurements of blood flow in the pulmonary trunk and the proximal aorta (6). In adults, QP/QS measured this way has been compared with QP/QS measured with oximetry (7,8), first-pass radionuclide angiography (9), and ventricular volumetric data (10). In children, MR velocity mapping has been compared with cardiac catheterization and ventricular volumetric data in four cases of shunt (11). However, to our knowledge, a numeric analysis of the agreement between MR velocity mapping and first-pass radionuclide angiography has not been performed, and the agreement of these methods in children remains to be established.
The objectives of the current study were therefore (a) to investigate the precision and accuracy of MR velocity mapping to determine QP/QS and (b) to evaluate how closely radionuclide angiography and MR velocity mapping agree in measurements of QP/QS in adults, adolescents, and children with left-to-right cardiac shunts.
| MATERIALS AND METHODS |
|---|
|
|
|---|
The control group for MR velocity mapping consisted of healthy volunteers (six men, six women; age range, 2053 years) with no history of cardiovascular disease and no direct or indirect signs of cardiac shunt as determined by means of cross-sectional and Doppler echocardiography.
The patient group (n = 24; five male patients, 19 female patients) (Table 1) consisted of 16 adults (age range, 2068 years), two adolescents (13 and 17 years), and six children (age range, 212 years). Among the adults, there were 11 with atrial septal defect, two with partially anomalous pulmonary venous return, and three with ventricular septal defect. One adolescent had scimitar syndrome, and the other had ventricular septal defect. Among the six children, there were two with atrial septal defect, one with partially anomalous pulmonary venous return, two with atrial septal defect and partially anomalous pulmonary venous return, and one with ventricular septal defect. Nonsinus rhythm was an exclusion criterion.
|
MR Imaging
Throughout the study, a 1.5-T system (Magnetom Vision; Siemens, Erlangen, Germany) with a 25-mT/m gradient strength and a 600-µsec gradient ramp time was used. Gradient-echo velocity mapping sequences provided by the manufacturer were used for the determination of blood flow.
In the phantom study, a standard head coil was used, and the flow rate of water doped with 0.13 mmol/L MnCl2 (Merck, Darmstadt, Germany) to the approximate relaxation time of blood was measured in an artificial homemade shunt system by means of MR velocity mapping. This flow rate was compared with the flow rate measured by means of beaker and timer. The artificial shunt system consisted of connected tubes with an inner diameter of 10 mm. The relation between the flow in one tube and the total flow was investigated in the range of a ratio of 15.
In control subject and patient studies, a phased-array body coil was used. To localize the ascending aorta, we used a set of coronal spin-echo images (635/30 [repetition time msec/echo time msec]) obtained with electrocardiographic triggering, which resulted in an effective repetition time of one cardiac cycle, a section thickness of 6 mm, a field of view typically of 400 mm, and a matrix typically of 172 x 256 (Fig 1). To localize the pulmonary trunk, after the coronal spin-echo images were obtained, an oblique sagittal turbo spin-echo breath-hold sequence (1,500/85) performed with electrocardiographic triggering was used, which resulted in an effective repetition time of two cardiac cycles, an echo train length of 23, a section thickness of 8 mm, a field of view typically of 400 mm, and a matrix typically of 256 x 512 (Fig 1).
|
|
|
MR velocity maps were obtained by means of automatic pixel-by-pixel subtraction of the two phase images, and each study resulted in 2540 subtracted velocity maps and corresponding modulus information. Measurements were made in vessel 1, vessel 2, and vessel 1 again, where the aorta and the pulmonary trunk were randomly assigned to represent vessel 1 and vessel 2, to avoid effects of physiologic drift in cardiac output (see MR Image Analysis). Subjects with a change in mean heart rate between measurements in the pulmonary trunk and the aorta of more than 10% were reexamined (n = 1).
MR Image Analysis
All data were transferred to a workstation (Sparc 10; Sun Microsystems, Mountain View, Calif) and evaluated by using a specially designed program (RADGOP; Context Vision, Linköping, Sweden). In each velocity map, a region of interest was drawn manually to completely cover the vessel of interest. The region of interest could be changed in size and shape between images, and it was possible to use the vessel contours from modulus images to delineate the vessel in the corresponding velocity maps (Fig 2).
|
|
|
Interobserver variability of QP/QS for both MR velocity mapping (H.A., C.H.) and radionuclide angiography (H.A., O.P.) was determined in the same group of 12 randomly selected patients with left-to-right shunts.
Statistical Analysis
Values are expressed as mean ± SD. Agreement between the two methods, repeatability of MR flow measurements, and interobserver variability were analyzed according to the method of Bland and Altman (16) and are expressed as percentages because QP/QS is a ratio. In the analysis of agreement or difference, the mean QP/QS at radionuclide angiography and at MR velocity mapping for each patient was considered the best estimate of the true value. The proportional difference in QP/QS was calculated as 100(radionuclide angiography QP/QS - MR velocity mapping QP/QS)/mean of (radionuclide angiography QP/QS + MR velocity mapping QP/QS). Error (bias) was calculated as 100(measurement - true value)/true value. Accuracy and precision were calculated as 100% - error (as a percentage) and 100% - SD (as a percentage), respectively. Repeatability was calculated as 100(measurement 1 - measurement 2)/mean of (measurement 1 + measurement 2).
Because the scatter of the differences between radionuclide angiography and MR velocity mapping in patients increased as QP/QS increased, a paired Student t test of differences between logarithmic measurements was used to test for significance (16). A P value less than .05 was considered to indicate a statistically significant difference.
| RESULTS |
|---|
|
|
|---|
|
Agreement between Radionuclide Angiography and MR Velocity Mapping
Radionuclide angiography yielded a higher QP/QS than MR velocity mapping in most patients (Fig 5a). In one of these patients, this may have been because radionuclide angiography was limited to an upper value of 3. The mean difference between the methods was 14% ± 13 (P < .001, n = 24) and was essentially proportional to shunt size (Fig 5b). Interobserver variability of QP/QS was four times higher for radionuclide angiography than for MR velocity mapping, 0% ± 16 versus 0% ± 4 (n = 12) (Table 2).
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
The error of the QP/QS measurement in the phantom, about 1%, is in agreement with expected theoretic errors in volume flow for vessels of the size examined (17). In control subjects, the mean QP/QS at MR velocity mapping was 1.03 ± 0.03, which is slightly higher than 1. This difference may be a result of the coronary circulation because flow measurements in the aorta were performed distal to the coronary arteries, because of drainage of the thebesian and bronchial veins into the systemic circulation, or because of systematic differences or variability in the measurements of pulmonary and aortic flow rates.
It seems reasonable to assume that the difference in QP/QS between radionuclide angiography and MR velocity mapping may be due to the lower accuracy and precision of the radionuclide angiographic method. The main reasons for this are that the interobserver variability was considerably higher for the radionuclide angiographic method than for the MR velocity mapping method, which is consistent with the findings of earlier studies (7,18), while results from the flow phantom experiment, from repeated flow measurements in vivo, and from the control subjects failed to demonstrate any appreciable bias or imprecision of the MR velocity mapping method (Figs 4, 6; Table 2).
Radionuclide angiography relies on a priori assumptions, becomes mathematically more sensitive when the degree of shunt increases, and cannot be used to accurately measure QP/QS greater than 3 (1315), which is not the case for MR velocity mapping. Radionuclide angiographic measurements of QP/QS are hampered by valvular regurgitation, slow circulation, or complex anatomy. This is not a major concern with MR velocity mapping, because it measures both forward and backward blood flow in specified vessels, and regurgitation is thus implicitly accounted for. Radionuclide angiography should be expected to show a relatively constant bias (overestimation of QP/QS) because the bronchial arterial circulation gives rise to an early recirculation in the analyzed region of interest (15). This may be one of the explanations for the differences in QP/QS obtained by using radionuclide angiography and by using MR velocity mapping in this study.
Respiratory gating was not used because it might introduce a systematic error in QP/QS measurements. Cardiac output of the left and right ventricles changes slightly in opposite directions during inspiration and expiration (19). QP/QS may thus be underestimated because acquisition of flow data is triggered during the expiration phase when right ventricle output decreases and left ventricle output increases. By not using respiratory gating, the flow data are averaged over the entire respiratory cycle, and the acquisition time is shortened. Acquisition of the radionuclide angiographic data requires 30 seconds while the patient is breathing, which also leads to averaging over a couple of respiratory cycles.
MR flow data were sampled over the entire R-R interval to account for regurgitation and anterograde blood flow during diastole. In this study, however, we found that it was sufficient to sample data during 75%80% of the R-R interval and extrapolate the last sample value through the true R-R interval because flow in late diastole is close to zero. QP/QS obtained in this way changed by only 1% ± 4 (n = 36) (Table 2). The practical implication of this is that the acquisition time for the flow data can be shortened by 50% without introducing major errors. The acquisition of MR velocity data for QP/QS measurement after anatomic MR mapping would thus require about 3 minutes each for acquisition of pulmonary and systemic flow, provided mean heart rate variation is less than 10% between measurements in the different vessels.
While acquisition of data is a relatively simple and short procedure with radionuclide angiography, it requires skill and takes longer with MR velocity mapping. On the other hand, while evaluation of data requires considerable experience with radionuclide angiography, it is easier with MR velocity mapping.
Study Limitations
MR velocity mapping was validated by using a flow phantom while radionuclide angiography was not. It is, however, an inherent difficulty of some methods, like radionuclide angiography or oximetry, that they cannot be validated by using the "true" values of QP/QS in a phantom. In this context, it may be tempting to regard MR velocity mapping as the standard method. Another limitation is that all subjects examined had sinus rhythm. One of the objectives of the study, however, was to evaluate the accuracy and precision of MR velocity mapping under defined conditions, which is why patients without sinus rhythm were not eligible for the study. This is not the case in the clinical setting. A highly irregular cardiac rhythm, especially with aberrant heartbeats, is likely to decrease the precision and accuracy of the MR velocity mapping method.
MR velocity mapping is an accurate and precise method for measurement of shunt size over the whole range of possible QP/QS values in adults, adolescents, and children with sinus rhythm. If the proportional difference between radionuclide angiography and MR velocity mapping is known, these noninvasive measurements of QP/QS are comparable. MR flow measurements may be the method of choice if MR anatomy mapping is performed anyway or if high accuracy and precision of QP/QS measurements are desirable.
| Acknowledgments |
|---|
| Footnotes |
|---|
Author contributions: Guarantor of integrity of entire study, H.A.; study concepts and design, H.A., C.H., F.S.; definition of intellectual content, H.A., C.H., F.S.; literature research, H.A., C.H., U.T., O.P., S.L., F.S.; clinical studies, H.A., C.H., U.T., K.H., G.B., O.P.; experimental studies, H.A., C.H., F.S.; data acquisition and analysis, H.A., C.H., O.P., F.S.; statistical analysis, H.A., C.H., F.S.; manuscript preparation, H.A., C.H., F.S.; manuscript editing, H.A., C.H., O.P., F.S.; manuscript review, H.A., C.H., U.T., K.H., G.B., O.P., S.L., F.S.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Petzina, M. Ugander, L. Gustafsson, H. Engblom, J. Sjogren, R. Hetzer, R. Ingemansson, H. Arheden, and M. Malmsjo Hemodynamic effects of vacuum-assisted closure therapy in cardiac surgery: Assessment using magnetic resonance imaging J. Thorac. Cardiovasc. Surg., May 1, 2007; 133(5): 1154 - 1162. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. I. Norton, C. Tong, R. B. J. Glass, and J. C. Nielsen Cardiac MR Imaging Assessment Following Tetralogy of Fallot Repair RadioGraphics, January 1, 2006; 26(1): 197 - 211. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Carlsson, P. Cain, C. Holmqvist, F. Stahlberg, S. Lundback, and H. Arheden Total heart volume variation throughout the cardiac cycle in humans Am J Physiol Heart Circ Physiol, July 1, 2004; 287(1): H243 - H250. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Beerbaum, H. Korperich, P. Barth, H. Esdorn, J. Gieseke, and H. Meyer Noninvasive Quantification of Left-to-Right Shunt in Pediatric Patients : Phase-Contrast Cine Magnetic Resonance Imaging Compared With Invasive Oximetry Circulation, May 22, 2001; 103(20): 2476 - 2482. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||