|
|
||||||||
Technical Developments |
1 From the Institute of Diagnostic Radiology, University Hospital Zurich, Ramistrasse 100, CH-8091 Zurich, Switzerland. Received May 27, 1998; revision requested August 5; revision received September 10; accepted December 16. Supported in part by German Research Foundation stipend Lu 687/1-1. Address reprint requests to J.F.D. (e-mail: debatin@drnr.usz.ch).
| Abstract |
|---|
|
|
|---|
Index terms: Colon, MR, 75.121412, 75.121419, 75.12143 Gadolinium Iron Magnetic resonance (MR), contrast agents, 75.121412, 75.121419, 75.12143 Magnetic resonance (MR), contrast enhancement, 75.12143 Manganese
| Introduction |
|---|
|
|
|---|
Presently, an enema with 10 mmol of gadopentetate dimeglumine (Magnevist; Berlex Laboratories, Wayne, NJ) per liter of water is used for MR colonography. Although it fulfills the SI requirements, the low viscosity induces backflow into the small bowel, which complicates subsequent analysis and reduces colonic distention. More important however, the enema has a volume of up to 2,000 mL and thus requires use of as much as 40 mL of a 0.5 mol/L paramagnetic contrast agent formulation. The associated cost is high, which negatively effects consideration of MR colonography as a screening alternative for colonic polyps.
The purpose of this study was to identify a less costly, more viscous enema with SI characteristics similar to those with the 10 mmol/L gadopentetate dimeglumine and water enema currently employed.
| Materials and Methods |
|---|
|
|
|---|
Assessment of New Contrast Media
Two agentsmanganese chloride and iron glycerophosphatewere assessed as possible replacements for gadopentetate dimeglumine. Water-based dilution series with each agent and with gadopentetate dimeglumine were imaged with the 3D spoiled GRE and 2D single-shot fast SE sequences. In addition, a water-dilution series with iron glycerophosphate and 600 mg of manganese chloride was imaged to determine possible synergistic effects with the two substances. On the basis of SI data acquired on 3D spoiled GRE images, the concentration with which the maximum SI was achieved was determined with a finely incremented dilution series. SI-concentration curves were obtained for each of the four evaluated contrast media.
The contrast media that provided the maximum contrast-to-noise ratio (CNR) on the 3D spoiled GRE images was used at the optimum concentration to fill the colon of three anesthetized pigs. The experiments were conducted in accordance with state regulations governing animal experiments. The colon was imaged with both the 3D spoiled GRE and 2D single-shot fast SE sequences. To determine CNR, SI characteristics of the contrast agentfilled colonic lumen were evaluated relative to the colonic wall and surrounding structures. Additionally, three colonic specimens exposed to the enema were obtained for histologic examination of iron uptake with Prussian blue staining.
Optimization of Enema Viscosity
Cellulose was added to enhance the mechanical properties of the enema by increasing its viscosity. In an initial step, the maximum additive of cellulose was determined that allowed passage of the mixture through the enema tip at 1.5 m of hydrostatic pressure. With this maximum cellulose concentration as a baseline, the enema viscosity was titrated for MR colonography by means of progressive dilution with water until complete colonic filling at 1.5-m hydrostatic pressure was possible.
Dilution experiments were performed in six patients (four men and two women; age range, 5372 years; mean age, 61 years) undergoing MR colonography in a larger study designed to determine the diagnostic performance of MR colonography in the detection of colonic masses. This study had been approved by the local ethics committee, and written informed consent was obtained from all patients. The patients had undergone colonic preparation for subsequent conventional colonoscopy. Colonic filling was monitored with a nonsection-selective 2D spoiled GRE acquisition collecting one image every second. The acquired images permitted assessment of colonic distention and reflux into the small bowel.
To determine the influence on SI characteristics of the cellulose additive in its previously optimized concentration, the gadopentetate dimeglumine and iron glycerophosphate dilution series were repeated with and without the cellulose additive, with the 3D spoiled GRE sequence. The ratio of the resultant SIs with and SIs without cellulose (SIwith/SIwithout) were plotted against the concentrations of gadopentetate dimeglumine and iron glycerophosphate. The paired Student t test was used to assess statistically significant differences in the collected data sets.
| Results |
|---|
|
|
|---|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Presently, the amount of gadopentetate dimeglumine required for optimum contrast on both T1-weighted 3D spoiled GRE and T2-weighted 2D single-shot fast SE images represents a cost factor that would preclude use of MR colonography for screening purposes. Since the contrast agent is administered rectally and not intravenously, there is a considerably wider latitude regarding tolerance and safety of agents. We evaluated two considerably less expensive agents with known T1- and T2-shortening effects as possible alternatives to gadopentetate dimeglumine: manganese chloride and iron glycerophosphate. Manganese was chosen as it is the substance responsible for the T1-shortening effects inherent in blueberry juice (4). Manganese chloride provided a maximum SI (SI, 640) that was 41% of that achieved with gadopentetate dimeglumine (SI, 1,560) at the 10 mmol/L concentration (Fig 1). This cannot be considered sufficient for MR colonography.
Subsequently, iron was tested in its most tolerable form as iron glycerophosphate (5). SI values achieved with iron glycerophosphate were far higher than those associated with manganese chloride and even exceeded those with gadopentetate dimeglumine (Fig 1). Maximum SI values were seen at a concentration of 45 mmol/L, which corresponds to an iron load of 28 g/L. In an attempt to lower the concentration of iron glycerophosphate, manganese chloride was added. The amount of manganese chloride added corresponded to the concentration associated with 87% (SI, 558) of the maximum SI increase in the manganese chloride dilution series (SI, 640). Indeed, the addition of manganese chloride to iron glycerophosphate resulted in higher SI values at lower concentrations. At higher concentrations, the combination reached a maximum SI value of 75% (SI, 1,560) of the maximum SI achieved with iron glycerophosphate alone (SI, 2,078) (Fig 1).
The in vivo imaging characteristics of iron glycerophosphate are favorable for MR colonography. At a concentration of 45 mmol/L, the enema was displayed as homogeneously bright on T1-weighted 3D spoiled GRE images and totally dark on T2-weighted 2D single-shot fast SE images (Fig 2). The visual impressions are corroborated by the high CNRs. In contrast to superparamagnetic iron oxide, which requires matching with diamagnetic barium sulfate particles to avoid susceptibility artifacts (6,7), images obtained with iron glycerophosphate depicted no artifacts up to the tested concentration of 45 mmol/L (Fig 2).
Iron glycerophosphate appears to be a safe agent. In fact, it has been used in high doses without relevant side effects in pediatric populations (8). It is poorly absorbed when administered orally (9,10). Although conclusive data are not available, it is not likely to cause any side effect if administered rectally. As was also confirmed in our study, there is no measurable resorption of this physiologic iron compound in the colon (11). It is conceivable, however, that iron could be resorbed if reflux occurs far into the proximal small bowel. Even if such an extensive reflux were to occur, the transient nature of the iron present in the small bowel would obviate any serious imbalance in iron metabolism. As an additional safeguard, iron could be embedded in macromolecules (iron-dextran, iron-cellulose) that are not resorbed. In this manner, the potential toxicity would be reduced as the viscosity increased.
Similar to double contrast barium enema examination (12,13), accurate assessment of the colonic lumen with MR colonography mandates adequate colonic distention (3). Beyond the administration of antiperistaltic, distention is dependent on the viscosity of the enema. Currently, the enema used at MR colonography consists mainly of water, which results in quick passage of enema fluid through the ileocecal valve into the small bowel in a vast majority of patients. Increases in viscosity must, however, be balanced against the need for easy and fast filling of the colon, particularly in view of limitations regarding usable hydrostatic pressure. Although results of in vitro experiments identified a 2% wt/wt cellulose additive as maximum, results of in vivo evaluations in six patients demonstrated a cellulose additive beyond 0.8% wt/wt to be impracticable. At this concentration, there was no reflux of contrast material into the small bowel and, consequently, distention was good (Fig 3). Results of in vitro experiments showed that the addition of cellulose to the enema did not have any measurable effect on SIs on 3D spoiled GRE images (Fig 4).
Iron glycerophosphate seems to be a feasible alternative to gadopentetate dimeglumine as a contrast enema in MR colonography. Before any clinical use, more safety data need to be collected (eg, by means of radiolabeling techniques with the radioactive isotope iron-59 instead of iron-56 [14]). A relative price difference of a factor of approximately 25 in conjunction with slightly better SI characteristics makes this a worthwhile endeavor. Even though MR colonography may become less expensive with the use of iron glycerophosphate, a cost analysis of MR colonography compared with other modalities, such as barium enema examination or colonoscopy, will need to be performed.
| Footnotes |
|---|
Author contributions: Guarantor of integrity of entire study, J.F.D.; study concepts, J.F.D., W.L., J.M.F.; study design, W.L.; definition of intellectual content, J.F.D., W.L., J.M.F.; literature research, W.L., N.S., F.L., D.W.; experimental studies, W.L., N.S., F.L.; data acquisition, W.L., N.S., D.W.; data analysis, D.W.; statistical analysis, D.W., N.S.; manuscript preparation and editing, W.L., J.F.D., D.W.; manuscript review, J.F.D., D.W., J.M.F.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
W Ajaj, G Pelster, U Treichel, F M Vogt, J F Debatin, S G Ruehm, and T C Lauenstein Dark lumen magnetic resonance colonography: comparison with conventional colonoscopy for the detection of colorectal pathology Gut, December 1, 2003; 52(12): 1738 - 1743. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Martin, M. Yang, D. Thomasson, and C. Acheson MR Colonography: Development of Optimized Method with ex Vivo and in Vivo Systems Radiology, November 1, 2002; 225(2): 597 - 602. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. T. Keogan and R. R. Edelman Technologic Advances in Abdominal MR Imaging Radiology, August 1, 2001; 220(2): 310 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Johnson and A. H. Dachman CT Colonography: The Next Colon Screening Examination? Radiology, August 1, 2000; 216(2): 331 - 341. [Abstract] [Full Text] |
||||
![]() |
W. Luboldt, P. Bauerfeind, S. Wildermuth, B. Marincek, M. Fried, and J. F. Debatin Colonic Masses: Detection with MR Colonography Radiology, August 1, 2000; 216(2): 383 - 388. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |